COMPOSITE 168 RESTORATIVE SYSTEM is substantially equivalent in intended use to the predicate device, K941899 (SDI Composites). It is to be used with Prime & Bond™ 2.1 Universal Light Cured Dental Adhesive System and it: a. Is indicated for cavity Classes I, II, III, IV, V, and VI in anterior and posterior teeth. b. May be used as a direct restorative material and for fabrication of inlays and onlays
Device Story
Composite 168 Restorative System is a visible light-activated, radiopaque dental restorative material. Used in conjunction with Prime & Bond 2.1 Universal Light Cured Dental Adhesive System. Packaged in unit doses or syringes. Applied by dental professionals in clinical settings for anterior and posterior tooth restorations. Material functions as a direct restorative or for inlay/onlay fabrication. Provides structural restoration of tooth defects. Benefits include esthetic repair and functional replacement of missing tooth structure.
Clinical Evidence
Bench testing only. Physical properties verified against ISO 4049 standards. Biocompatibility supported by negative Ames testing on main components and prior use of components in legally marketed devices.
Technological Characteristics
Visible light-activated, radiopaque, one-component VLC composite restorative. Meets ISO 4049 physical property standards. Packaged in unit dose or syringes. Compositional similarity to predicate K863092.
Indications for Use
Indicated for cavity Classes I, II, III, IV, V, and VI in anterior and posterior teeth; may be used as a direct restorative material and for fabrication of inlays and onlays.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
FEB - 9 1998
# 510(k) SUMMARY
)
ﻣﺴﻴﺔ
NAME & ADDRESS:
K973221
#### DENTSPLY International
570 West College Avenue P.O. Box 872 York, PA 17405-0872 (717) 845-7511 ax 1777 854-2343
P. J. Lehn Telefax (717) 849-4343
P. Jeffery Lehn CONTACT:
August 20 . 1997 DATE PREPARED:
COMPOSITE 168 RESTORATIVE SYSTEM TRADE OR PROPRIETARY NAME:
Dental restorative material COMMON OR USUAL NAME:
CLASSIFICATION NAME: Tooth resin shade material
K863092 Dental Composite Restorative PREDICATE DEVICE: SDI Composites K941899
DEVICE DESCRIPTION: COMPOSITE 168 RESTORATIVE is a visible light activated, radiopaque restorative material for anterior and posterior restorations of primary and permanent teeth. It is to be used with Prime & Bond 2.1 Universal Light Cured Dental Adhesive System. This one-component. VLC composite restorative is packaged in unit dose or svringes.
The physical properties of COMPOSITE 168 RESTORATIVE meet ISO Standard 4049.
INTENDED USE: COMPOSITE 168 RESTORATIVE SYSTEM is substantially equivalent in intended use to the predicate device, K941899 (SDI Composites). It is to be used with Prime & Bond™ 2.1 Universal Light Cured Dental Adhesive System and it:
a. Is indicated for cavity Classes I, II, III, IV, V, and VI in anterior and posterior teeth.
- b. May be used as a direct restorative material and for fabrication of inlays and onlays
TECHNOLOGICAL CHARACTERISTICS: The new COMPOSITE 168 RESTORATIVE SYSTEM is substantially equivalent in composition to the predicate device, K863092 (dental composite restorative).
We believe that the prior use of the components found in COMPOSITE 168 RESTORATIVE SYSTEM in DENTSPLY legally marketed devices support our decision that additional biocompatibility studies with the final formulation are not necessary.
The results of Ames testing on the main components were negative.
We believe that the prior use of the components in the COMPOSITE 168 RESTORATIVE SYSTEM in predicate devices, the performance data provided, and the results of previous biocompatibility testing support the safety and effectiveness of COMPOSITE 168 RESTORATIVE SYSTEM for the intended uses.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### 8 1398 FEB
Mr. P. Jeffrey Lehn Director, Corporate Compliance and Regulatory Affairs DENTSPLY International 570 West College Avenue P.O. Box 872 York, Pennsylvania 17405-0872
Re: K973221 Composite 168 Restorative System Trade Name: Requlatory Class: II Product Code: EBF Dated: January 28, 1998 Received: January 30, 1998
Dear Mr. Lehn:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਖੇ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{2}------------------------------------------------
Page 2 - Mr. Lehn
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
the A. Tim Ulatowsk Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## PREMARKET NOTIFICATION
## INDICATIONS FOR USE STATEMENT
|< 973221 510(K) Number:
Device Name:
j
-
### COMPOSITE 168 RESTORATIVE SYSTEM
COMPOSITE 168 RESTORATIVE SYSTEM is to be used with Prime & Bond™ 2.1
Universal Light Cured Dental Adhesive System:
• Indicated for cavity Classes I, II, III, IV, V, and VI in anterior and posterior teeth.
•May be used as a direct restorative material
·May be used for fabrication of inlays and onlays
Susan Runser
n-Off Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sia Division of Dental, Infaction Contr and General Hospital D 5 1 O(k) Number OR Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use No
{4}------------------------------------------------
#### a 1998 133
BISCO, INC., 1100 W. Irving Park Road, Schaumburg, IL 60193
510(k) submission for NTL-FLOW™
### 510(k) SUMMARY
As Required by the Safe Medical Devices Act of 1990
#### IDENTIFICATION OF THE LEGALLY MARKETED PREDICATE DEVICE
PREDICATE DEVICE AELITEFLO
AELITEFLO is a low viscosity, light-cure hybrid composite restorative material. It is a moderately filled (59% w/w), radiopague, flowable composite designed primarily for restoration of shallow defects such as incipient Class V lesions. It functions well as an anterior esthetic material to reshape and esthetically restore stained or otherwise developmentally compromised anterior teeth. AELITEFLO bonds micromechanically and chemically to dental primers/bonding resin adhesives through co-polymerization of the former's air inhibited layer. AELITEFLO is designed to be marketed as a stand alone product but can be made available in kit form including semi-gel etchant (K945604) and COMPOSITE PRIMER.
#### DESCRIPTION OF APPLICANT DEVICE NTL-FLOW
NTL-FLOW is a dual-cure (light or heat and light) moderately filled (60%), radiopaque low modulus flowable composite. Its physical properties are similar to the predicate device and uses are identical. Like the predicate device, NTL-FLOW is a glass frit filled dimethacrylate composite. It hardens by a light or heat and light cure polymerization mechanism employing a light initiator, and a tertiary amine activator and peroxide initiator. Both devices are designed to be used with high quality dentin / enamel adhesive systems. The material can be used in a conventional fashion or it can be processed in a dedicated curing device that employs light, and a substantially oxygen free atmosphere created by flushing the device with an inert gas such as nitrogen. The heat is supplied by the lamp providing the light. The comparative low modulus facilitates the use of the material as a first increment liner that is intended to bond to the dental adhesive as well as the next increment composite.
#### INTENDED USES OF APPLICANT DEVICE NTL-FLOW
NTL-FLOW is indicated for: anterior esthetic restorative material for Class III and IV as well as conservative Class I restorations not involving opposing occlusal wear, direct esthetic veneer restorative for masking stains and developmental anomalies, pit and fissure sealant, core buildup material to replace missing tooth structure, dental cement/luting agent, composite repair, and a liner/base material.
{5}------------------------------------------------
BISCO, INC., 1100 W. Irving Park Road, Schaumburg, IL 60193 510(k) submission for NTL-FLOW"
# SCIENTIFIC CONCEPTS and SIGNIFICANT PERFORMANCE CHARACTERISTICS
AELITEFLO and Bisco NTL-FLOW are very similar with regard to chemical composition and AELT EF LO und Blood Nical properties. Significantly, AELITEFLO and NTL-FLOW, following curing, have been designed to be more flexible than conventional dental composites. Also, the flow characteristics or so-called syringability of these materials are superior to more thighly filled materials. The viscoelastic properties of conventional microfil composites are dependent on the interaction of submicron silica fillers and the design resin. Because of the enormous surface area of submicron silica, a much smaller amount of filler can be incorporated into a compared with a larger filler such as 0.7um barium glass as used in NTL-FLOW and AELITEFLO. The flowability of NTL-FLOW, therefore, is achieved by a careful balance of resin viscosities and filler loading. Submicron filler is also used in NTL-FLOW but to a much !esser extent than conventional microfil composites.
The flexibility of cured NTL-FLOW is achieved by an imaginative balance of polymerizable resins. These resins in concert with the fillers noted above achieve the properties described in this application, namely high service strength, flexibility, and ease of manipulation or svringability. Relative reduction of filler content typically reduces surface hardness of composites and that is evident with NTL-FLOW which has approximately 10% lower Barcol hardness than conventional composites.
The chemical composition of NTL-FLOW and AELITEFLO are quite similar. Both are silica and /or glass filled dimethacrylate light-cure composites. NTL-FLOW is based on the ethoxylated version of Bis-GMA whereas AELITEFLO is primarily a Bis-GMA triethyleneglycol dimethacrylate combination.
The non-clinical tests used for this submission are similar to those specified in ISO 4049 and American Dental Association Specification #27; both are for dental resin based filling materials. Diametral tensile testing (DTS) is an accepted method to characterize the tensile strength of brittle materials and the flexural modulus test addresses the strength in three point loading. DTS values are 41.3 and 39.0 MPa for NTL-FLOW and AELITEFLO respectively. Cytotoxicity of NTL-FLOW was performed by a commercial testing laboratory (NAMSA) and the product was found to be non-toxic.
Comparison of NTL-FLOW to the predicate device shows both are very similar with the major difference being the methods of final cure of the product. The proposed new product cures using visible light or heat (125°C) and light. The predicate device can cures using visible light exclusively.
Carme h. Sandik
James L. Sandrik, Ph.D. BISCO, Inc. Schaumburg, IL 60193 November 25, 1997
113.
{6}------------------------------------------------
BISCO, INC., 1100 W. Irving Park Road, Schaumburg, IL 60193 510(k) submission for NTL-FLOW™
### REFERENCES
- 1. Craig, R. G. (ed), Restorative Dental Materials, eighth edition, C. V. Mosby Co., St. Louis, 1989, p 277.
- 2. International Standards Organization, ISO 4049:1988/Cor. 1:1992 (E), Resin Based Filling Materials, 1992.
- 3. Quality Control Procedures Manual, QC0003, Bisco, Inc., 1995.
- 4. ibid. QC0004
- 5. ibid. QC0006
T
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.