K973169 · Span-America Medical Systems, Inc. · FNM · Feb 20, 1998 · General Hospital
Device Facts
Record ID
K973169
Device Name
PRESSURE GUARD SITE SELECT (A)
Applicant
Span-America Medical Systems, Inc.
Product Code
FNM · General Hospital
Decision Date
Feb 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For the prevention and treatment of pressure ulcers.
Device Story
PressureGuard Site Select is an air flotation, alternating pressure mattress system. It consists of an air inflation system housed within a foam shell, featuring a foam topper and bolsters. Unlike the predicate, which uses lengthwise cylinders, this device utilizes side-to-side air cylinders organized into four zones: head, upper torso, lower torso, and heel. The system provides multiple operational modes, including Normal (alternating on/off), Heel Relief (alternating on/off), Contour (customized firmness per zone), and CPR (firm surface). The device is intended for use in clinical or home settings to manage patient pressure distribution. Healthcare providers or caregivers adjust air pressure set points to meet individual patient needs. By providing alternating pressure and customized zone firmness, the device aims to reduce interface pressure, thereby preventing or treating pressure ulcers.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Air flotation alternating pressure mattress. Components: foam shell with topper and bolsters, side-to-side air cylinders. Four-zone configuration (head, upper torso, lower torso, heel). Operational modes: Normal, Heel Relief, Contour, CPR. Connectivity: None specified. Power: Air inflation system.
Indications for Use
Indicated for the prevention and treatment of pressure ulcers in patients requiring pressure relief support.
Regulatory Classification
Identification
An alternating pressure air flotation mattress is a device intended for medical purposes that consists of a mattress with multiple air cells that can be filled and emptied in an alternating pattern by an associated control unit to provide regular, frequent, and automatic changes in the distribution of body pressure. The device is used to prevent and treat decubitus ulcers (bed sores).
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
{1}------------------------------------------------
SPAN-AMERICA | K973169 MEFICAL SYSTEMS; TNC.
博
K973169
FEB 20 1998
## PressureGuard Site Select 510K Summary
| 1. | Submitter's Name: | Span-America Medical Systems, Inc. |
|----|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Address: | 70 Commerce Center<br>Greenville, SC 29615 |
| | Telephone Number: | (864) 288-8877 |
| | Contact Person: | James D. Ferguson, President & CEO<br>Wanda Totton, Director of Quality |
| | Date Prepared: | October 21, 1997 |
| 2. | Trade Name: | PressureGuard Site Select, Model A |
| | Common Name: | Alternating Pressure Mattress |
| | Classification Name: | Alternating Pressure Air Flotation Mattress,<br>CFR 880.5550, Classification No. 80 FMN. |
| 3. | Predicate Device: | PressureGuard IV, 510K953503 |
| 4. | Description: | The PressureGuard Site Select is an air flotation,<br>alternating pressure support system which provides for<br>customized selection of surface firmness for a head, upper<br>torso, lower torso, and heel zones. |
| 5. | Indications for Use: | For the prevention and treatment of pressure ulcers. |
| 6. | Substantial Equivalence: | The product is similar in function and intended use to the<br>PressureGuard IV, 510K953503. Both systems consists of<br>an air inflation system within a foam shell. The foam shell<br>includes a foam topper with bolsters that frame the<br>mattress. For the PressureGuard IV, four air cylinders<br>run lengthwise within the mattress and operate as two-<br>pairs. For the PressureGuard Site Select, the air cylinders<br>run side-to-side and operate as four zones, defined as<br>head, upper torso, lower torso, and heel zones |
{2}------------------------------------------------
Span-America Medical Systems, Inc. PressureGuard Site Select 510K Summary Page 2 of 2
> For both systems, air in the air cylinders is adjusted to defined pressure set points for four defined modes of operation. For the PressureGuard IV, a Center Mode (supine surface), CPR Mode (firm surface for performing CPR with a crash board), Fixed Roll ( 30 degree left and right rolls with two-hour dwell times), and Custom Roll (provides for options for roll pattern, roll angles, and dwell times) are provided.
> For the PressureGuard Site Select, Normal Mode, Alternating Off; Normal Mode, Alternating On; Heel Relief Mode, Alternating Off; Heel Relief Mode, Alternating On; Contour, and CPR modes are provided. The Normal Mode, Alternating Off Mode is comparable to the Center Mode for PressureGuard IV. The Heel Relief Mode, Alternating Off is comparable to the Center Mode for PressureGuard IV with lowered internal air pressures for the heel zone. The CPR Mode is comparable to the CPR Mode for PressureGuard IV. In addition, alternating pressure modes are options for both the Normal Mode and Heel Relief Modes. The Contour Mode provides for customized adjustment of firmness settings for each of the four zones to meet individual patient needs.
> Both products are indicated for the prevention and treatment of pressure ulcers.
lho
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized eagle. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The eagle is depicted with its wings spread, symbolizing protection and service.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 20 1998
Ms. Wanda J. Totton Director of Quality Span-America Medical Systems, Inc. P.O. Box 5231 29606 Greenville, South Carolina
Re: K973169 PressureGuard Site Select Trade Name: Requlatory Class: II Product Code: _ ENM _ _ _ _ _ Dated: January 26, 1998 Received: January 27, 1998
Dear Ms. Totton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ದ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Wanda J. Totton
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page ol
510(k) Number (if known):
Device Name: Pressure Guard Site Alert
Indications For Use:
For the prevention and treatment of
Pressure ulcers.
L.C. Patterson
8-7-97
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
tion. (GLE)
(Division Sign-Off)
Division of General Restorative Devices K973169:
510(k) Number.
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use X
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.