K973167 · Icu Medical, Inc. · FPA · Nov 17, 1997 · General Hospital
Device Facts
Record ID
K973167
Device Name
POSITIVE PRESSURE CONNECTOR
Applicant
Icu Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Nov 17, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The CLC 2000 is a single use, sterile, non-pyrogenic, swab-able, bi-directional valve intended for use as an accessory to Intravascular administration set. The CLC 2000 provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein or artery). The CLC 2000 also provides access for the withdrawal of fluids from a patient's vascular system.
Device Story
The CLC 2000 is a sterile, single-use, swab-able, bi-directional luer-activated valve used as an accessory to intravascular administration sets. It functions as a closed system; insertion of a luer taper physically displaces an internal poppet and O-rings to open the valve for fluid injection or aspiration. Upon removal of the luer, the poppet and O-rings automatically return to a sealed position, displacing fluid through the male luer to prevent blood reflux into the catheter lumen. The device is used in clinical settings by healthcare providers to facilitate vascular access. It benefits patients by maintaining a closed system during fluid management, reducing the risk of contamination and blood exposure.
Indicated for patients requiring intravenous or intra-arterial access for fluid administration or fluid withdrawal via an administration set.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
NOV 17 1997
K473/47
# 510(k) Summary of Safety and Effectiveness
# The following section is included as required by the Safe Medical Device Act (SMDA) of 1990.
ICU Medical, Inc. Name: Address: 951 Calle Amanecer San Clemente, CA 92673
Contact Person: Salvadore F. Palomares Phone Number: (714)366-2183 (714)366-8368 Fax Number:
## 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is belng submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K973167
#### Applicant Information:
| Date Prepared: | November 14, 1997 |
|----------------|----------------------------------------------|
| Name: | ICU Medical Inc. |
| Address: | 951 Calle Amanecer<br>San Clemente, CA 92673 |
Contact Person: Salvadore F. Palomares Phone Number: (714)366-2183 Fax Number: (714)366-8368
### Device Information:
Trade Name: CLC 2000 Common Name: Intravascular Administration Set Classification Name: Intravascular Administration Set
#### Equivalent Device:
B. Braun V2 Ultrasite Injection Site (K955585) ICU Medical CLAVE Connector (K970855)
#### Device Description:
The CLC 2000 is a single use, sterile, non-pyrogenic, swab-able, bi-directional valve intended for use as an accessory to Intravascular administration set.
#### Intended Use:
The CLC 2000 provides access for the administration of fluids from a container to a patient's vascular system through the administration sot's needle or catherer (which is inserted into a vein or artery). The CLC 2000 also provides access for the withdrawal of fluids from a patient's vascular system.
{1}------------------------------------------------
Comparison To Predicate Device:
| Product Labeling | Sterile, Non-Pyrogenic, Single<br>Use, Swab-able, IV Connector.<br>Directions for Use included | Sterile, Non-Pyrogenic, Single<br>Use, Directions for Use<br>included |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The CLC 2000 is a swab-able,<br>bi-directional valve used to<br>access any vein or artery or<br>aspiration of fluids from a<br>patient's vascular system. | The Ultrasite Injection Site Is<br>used for the injection<br>gravity/pump flow or aspiration<br>of fluids. |
| Design | Swab-able bl-directional luer<br>activated valve. Closed system<br>activated by a luer taper. The<br>luer taper physically moves the<br>poppet and O-rings, opening<br>the valve. With the valve in the<br>open position, fluids can be<br>injected or withdrawn. The<br>poppet and O-rings move to the<br>sealed position automatically<br>when the luer taper is removed.<br>As the poppet and O-rings<br>move to the seal position, fluid<br>is displaced through the male<br>luer preventing blood from<br>entering the lumen of the<br>catheter. | Swab-able cap-less two-way<br>valve. Closed system activated<br>by a luer taper. The luer taper<br>puts pressure on the plunger,<br>opening the valve. With the<br>valve in the open position,<br>fluids can be injected or<br>withdrawn. The valve closes<br>automatically when the luer<br>taper is removed. |
| Materials | Body - 30% glass filled<br>polyester<br>Poppet - Polycarbonate<br>O-rings - Silicone Rubber<br>Spring - Stainless Steel 302<br>Packaging-Medical packaging<br>grade fiber-free peelable paper<br>lidding and pouching material. | Body - Polyurethane<br>Plunger - latex free non-<br>compromised elastomeric<br>material<br>Spring - metal<br>Packaging-Medical packaging<br>grade fiber-free peelable paper<br>lidding and pouching material. |
Sffalman
Salvadore F. Palomares Manager of Regulatory Affairs iCU Medical, Inc.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES ---
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 7 1997
Mr. Salvadore F. Palomares Manager of Regulatory Affairs ICU Medical, Incorporated 951 Calle Amenecer San Clemente, California 92673
K973167 Re : CLC 2000 Swabbable IV Connector Trade Name: Requlatory Class: II Product Code: FPA Dated: Auqust 21, 1997 Received: August 25, 1997
Dear Mr. Palomares:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531
{3}------------------------------------------------
Page 2 - Mr. Palomares
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact ' the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timo A Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
. Enclosure
{4}------------------------------------------------
# 510(k) Number (if known): K973167
# Device Name: CLC 2000 SWahle V Connector
## Indications For Use:
The CLC 2000 is a single use, sterile, non-pyrogenic, swab-able, bi-directional valve intended for use as an accessory to Intravascular administration set. The CLC 2000 provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein or artery). The CLC 2000 also provides access for the withdrawal of fluids from a patient's vascular system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NBBDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Paulocesar Cavalcante
(Division Sign-Off) (1) wision of Dental, Infection Com a d General Hospital D E, Nik) Number .
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.