The PSA Calibrator is an in vitro diagnostic product intended to be used to calibrate the Prostate Specific Antigen (PSA) method for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassy module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method.
Device Story
PSA Calibrator is a frozen, multi-level in vitro diagnostic product used to calibrate PSA assays on the Dimension® RxL clinical chemistry system. Level 1 consists of horse serum; Levels 2-5 contain human PSA-ACT in a bovine serum albumin base. The kit includes ten 1 mL vials. It is used in clinical laboratory settings by trained technicians to establish a calibration curve for the chemistry system. By ensuring accurate measurement of PSA-ACT levels across the assay range, the device supports the clinical utility of the PSA test, which is used by physicians to monitor prostate health. The calibrator does not perform analysis itself but provides the reference points necessary for the instrument to translate raw signals into clinically actionable PSA concentration values.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Frozen, multi-level calibrator kit. Level 1: horse serum; Levels 2-5: human PSA-ACT in bovine serum albumin base. 5 levels total. Designed for use with the Dimension® RxL clinical chemistry system heterogeneous immunoassay module.
Indications for Use
Indicated for use as an in vitro diagnostic calibrator for the Prostate Specific Antigen (PSA) method on the Dimension® RxL clinical chemistry system. No specific patient population, age, or gender restrictions are defined for the calibrator itself.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
aca® plus Prostate Specific Antigen Calibrator
Related Devices
K020980 — DIMENSION TPSA CALIBRATOR, MODEL #RC459 · Dade Behring, Inc. · Jun 21, 2002
K973099 — PROSTATE SPECIFIC ANTIGEN CONTROL · Dade Chemistry Systems, Inc. · Sep 29, 1997
K123127 — VALIDATE PSA CALIBRATION VERIFICATION/ LINEARITY TEST KIT MODEL 406 · Maine Standards Co. · Jun 14, 2013
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "DADE" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The letters are evenly spaced and aligned horizontally.
KA73100
SEP 2 9 1997
## DADE INTERNATIONAL
Chemistry Systems P.O. Box 6101
Newark, DE 19714
Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter's Name: | Rebecca S. Ayash<br>Dade International Inc.<br>Building 500, Mailbox 514<br>P.O. Box 6101<br>Newark, DE 19714-6101 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| Date of Preparation: | 8/15/97 |
| Device Name: | Prostate Specific Antigen (PSA) Calibrator |
| Classification Name: | Calibrator, Secondary |
| Predicate Device: | aca® plus Prostate Specific Antigen Calibrator |
Device Description: Prostate Specific Antigen (PSA) Calibrator for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module is a frozen product. The Level 1 calibrator is horse serum based with no detectable PSA. Levels 2 through 5 contain human PSA-ACT in a bovine serum albumin base. The kit consists of ten vials; two at each of five levels, containing 1 mL each.
Intended Use: PSA Calibrator is intended to be used to calibrate the PSA method for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module.
## Comparison to Predicate Device:
| | Dimension® RxL PSA Calibrator | aca® plus PSA Calibrator |
|--------------|-------------------------------|------------------------------|
| Intended Use | Calibrator | Calibrator |
| Analyte | PSA-ACT | PSA-ACT |
| Matrix | L-1 horse serum<br>L-2-5 BSA | L-1 horse serum<br>L-2-3 BSA |
| Form | frozen | liquid |
| Volume | 1.0 mL per vial | 3.0 mL per vial |
| Levels | 5 levels | 3 levels |
{1}------------------------------------------------
## Comments on Substantial
Equivalence: Both the PSA Calibrator for the Dimension® RxL system and the aca® plus PSA Calibrator are manufactured using the same matrices and contain PSA-ACT as the analyte source. Both products are intended to be used as calibrators for prostate specific antigen assays.
Conclusion: PSA Calibrator for the Dimension® RxL system is substantially equivalent to the aca® plus PSA Calibrator based on the comparison summarized above.
Rebecca Slayden
Rebecca S. Ayash Regulatory Affairs and Compliance Manager Date: 8/15/97
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure with three tail feathers.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Rebecca S. Ayash Requlatory Affairs and Compliance Manager Dade International Inc. Building 500, mailbox 514 P.O. Box 6101 Newark, DE 19714-6101
SEP 2 9 1997
Re: K973100 Trade Name: Prostate Specific Antigen (PSA) Calibrator Requlatory Class: II Product Code: JIT Dated: August 15, 1997 Received: August 19, 1997
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications Statement
Device Name: Prostate Specific Antigen (PSA) Calibrator
Indications for Use: The PSA Calibrator is an in vitro diagnostic product intended to be used to calibrate the Prostate Specific Antigen (PSA) method for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassy module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method.
Rebecca Scapph
Rebecca S. Ayas Regulatory Affairs and Compliance Manager Date: 9/25/97
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
נ
K 973100
(510(k) Number
Victor E. Macher
Division Sign-Off Office of Device Evaluation
Prescription Use ✓
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.