ONTRAK TESTSTIK FOR AMPHETAMINES, ONTRAK TESTSTIK FOR COCAINE, ONTRAK TESTSTIK FOR MORPHINE, ONTRAK TESTSTIK FOR PCP, ON
K973075 · Roche Diagnostic Systems, Inc. · DKZ · Oct 9, 1997 · Clinical Toxicology
Device Facts
Record ID
K973075
Device Name
ONTRAK TESTSTIK FOR AMPHETAMINES, ONTRAK TESTSTIK FOR COCAINE, ONTRAK TESTSTIK FOR MORPHINE, ONTRAK TESTSTIK FOR PCP, ON
Applicant
Roche Diagnostic Systems, Inc.
Product Code
DKZ · Clinical Toxicology
Decision Date
Oct 9, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Indications for Use
The 5 ONTRAK TESTSTIK Assays contained in this submission are in vitro diagnostic tests intended for professional use for the qualitative detection of drug or drug metabolite in urine. They are the ONTRAK TESTSTIK for Amphetamines (1000 ng/mL cutoff), the ONTRAK TESTSTIK for Cocaine (300 ng/mL cutoff), the ONTRAK TESTSTIK for THC (50 ng/mL cutoff), the ONTRAK TESTSTIK for Morphine (300 ng/mL cutoff) and the ONTRAK TESTSTIK for PCP (25 ng/mL cutoff).
Device Story
ONTRAK TESTSTIK assays are lateral flow, competitive microparticle capture inhibition tests for professional use. Device consists of a reagent strip with antibody-coated blue microparticles and immobilized drug conjugate on a membrane. Urine sample is absorbed into a pad and migrates via capillary action. In the absence of drug, antibody-coated microparticles bind to the immobilized drug conjugate, forming a blue band (negative result). If drug is present, it binds to the microparticles, inhibiting binding to the conjugate, resulting in no blue band (positive result). A 'TEST VALID' band confirms proper flow. Results are visually interpreted by a clinician to aid in drug screening. Benefits include rapid, qualitative detection of specific drugs of abuse in urine samples.
Clinical Evidence
Bench testing only. Performance evaluated by comparing TESTSTIK results to GC/MS reference methods. Precision reported as >95% confidence at 150% of cutoff. Accuracy studies showed high concordance with GC/MS across all five drug analytes.
Technological Characteristics
Lateral flow immunoassay using competitive microparticle capture inhibition. Reagents: blue dyed microparticles coated with mouse monoclonal or rabbit polyclonal antibodies, drug conjugates immobilized on a membrane, and mouse monoclonal anti-BSA antibody. Qualitative colorimetric endpoint read. Sample type: urine.
Indications for Use
Indicated for professional in vitro diagnostic qualitative detection of amphetamines (1000 ng/mL), cocaine metabolite (300 ng/mL), morphine (300 ng/mL), phencyclidine (25 ng/mL), or cannabinoids (50 ng/mL) in human urine.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "Roche" in bold, white letters against a black, hexagon-shaped background. The hexagon is oriented horizontally, with the word "Roche" centered within it. The font is sans-serif and appears to be a simple, clean design.
# Roche Diagnostic Systems
A Member of the Roche Group
Roche Diagnostic Systems, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
Direct Dial Fax
K973075
# 510(k) Summary
## ONTRAK TESTSTIK Assays
In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.
#### Identification of 510(k) Sponsor: I.
Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
510(k) Submission dated August 15, 1997
James W. Haynes Contact: Regulatory Affairs Associate Phone: (908) 253-7569 Fax: (908) 253-7547
{1}------------------------------------------------
#### II. Device Name:
The device name, including both the trade/proprietary name and the classification name are provided in the table below.
| Proprietary Name | Classification Name | Regulation Number |
|------------------------|---------------------------|-------------------|
| ONTRAK TESTSTIK for... | | |
| Amphetamines | Amphetamines test system | 862.3100 |
| Cocaine | Cocaine test system | 862.3250 |
| Morphine | Morphine test system | 862.3640 |
| PCP | Phencyclidine test system | NA |
| THC | Cannabinoid test system | 862.3870 |
Table 1
{2}------------------------------------------------
### Identification of the legally marketed device to which the 510(k) sponsor claims III. equivalence:
The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.
| Proprietary Name | Predicate Product Name | K number | Date of substantial<br>equivalence |
|------------------------|-----------------------------------|-----------|------------------------------------|
| ONTRAK TESTSTIK for... | | | |
| Amphetamines | Abuscreen ONTRAK for Amphetamines | K881817/A | 7/29/88 |
| Cocaine | Abuscreen ONTRAK for Cocaine | K881815 | 7/29/88 |
| Morphine | Abuscreen ONTRAK for Morphine | K882496 | 9/2/88 |
| PCP | Abuscreen ONTRAK for PCP | K895906 | 1/26/90 |
| THC | Abuscreen ONTRAK for THC | K904856 | 12/13/90 |
## Table 2
#### IV. Description of the Device/Statement of Intended Use:
The 5 ONTRAK TESTSTIK Assays contained in this submission are in vitro diagnostic tests intended for professional use for the qualitative detection of drug or drug metabolite in urine. They are the ONTRAK TESTSTIK for Amphetamines (1000 ng/mL cutoff), the ONTRAK TESTSTIK for Cocaine (300 ng/mL cutoff), the ONTRAK TESTSTIK for THC (50 ng/mL cutoff), the ONTRAK TESTSTIK for Morphine (300 ng/mL cutoff) and the ONTRAK TESTSTIK for PCP (25 ng/mL cutoff).
The ONTRAK TESTSTIK Assays are based on the principle of microparticle capture inhibition. These tests rely on the competition between the specific drug, which may be present in the urine being tested, and drug conjugate immobilized on a membrane test chamber.
When an ONTRAK TESTSTIK is immersed in the urine sample, some of the sample is absorbed into the TESTSTIK sample pad. The absorbed sample travels through a reagent strip contained in the device by capillary action. In the reagent strip, the sample rehydrates and mobilizes antibody-coated blue microparticles. The microparticle-urine suspension continues to migrate through the reagent strip and comes in contact with the immobilized drug conjugate. In the absence of drug in the urine, the antibody-coated microparticles bind to the drug conjugate and a blue band is formed at the result window ("negative" sign).
{3}------------------------------------------------
When drug is present in the specimen. it binds to the antibody-coated particles, the microparticles are inhibited from binding the drug conjugate and no blue band is formed at the result window. Therefore, a positive sample causes the membrane to remain white ("positive" sign).
An additional antibody/antigen reaction occurs at the "TEST VALID" area. The "TEST VALID" blue band forms when antibodies, which are embedded in the reagent membrane, bind to the antigen on the blue microparticles.
### Summary of the technological characteristics of the new device in comparison to V. those of the predicate.
Tables 3-7 outline the technological characteristics (methodologies) of the ONTRAK TESTSTIK Assays in comparison to those of legally marketed predicate products.
### VI. Brief discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence:
Tables 3-7 demonstrate the results of clinical and nonclinical studies performed using the ONTRAK TESTSTIK Assays. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this information concludes that the performance of these devices are essentially equivalent to other legally marketed devices of a similar kind.
{4}------------------------------------------------
-
| Table 3 - ONTRAK TESTSTIK for Amphetamines | | |
|--------------------------------------------|--|--|
|--------------------------------------------|--|--|
| | ONTRAK TESTSTIK<br>for Amphetamines | Abuscreen ONTRAK for<br>Amphetamines |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Methodology | Competitive microparticle capture<br>inhibition | Competitive latex agglutination<br>inhibition |
| Measurement | Qualitative | Qualitative |
| Sample type | urine | urine |
| Endpoint read | color | agglutination pattern |
| Cutoff(s) | 1000 ng/mL | 1000 ng/mL |
| Reagent (active<br>ingredients) | 1. Blue dyed microparticles coated<br>with mouse monoclonal anti-<br>amphetamine analogue<br>2. Drug conjugates immobilized on a<br>membrane<br>3. Mouse monoclonal anti-BSA<br>antibody immobilized on a<br>membrane | 1. Sheep anti-amphetamine<br>antibody in a buffered<br>solution<br>2. Reaction buffer<br>3. Latex-amphetamine<br>conjugate in a buffered<br>solution |
| Controls | ONTRAK TESTCUP Positive & Negative<br>Controls | Abuscreen ONTRAK for<br>Amphetamines Negative Control |
| Performance Characteristics: | | |
| Precision | > 95% confidence at 150% cutoff | > 99% confidence at 200% of<br>cutoff |
| Accuracy:<br>Positive Samples | TesTstik | GC/MS |
| + | 50 | 45 |
| - | 0 | 1 |
{5}------------------------------------------------
# Table 4 - ONTRAK TESTSTIK for Cocaine
| | ONTRAK TESTSTIK<br>for Cocaine | | Abuscreen ONTRAK for<br>Cocaine |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Methodology | Competitive microparticle capture<br>inhibition | | Competitive latex agglutination<br>inhibition |
| Measurement | Qualitative | | Qualitative |
| Sample type | urine | | urine |
| Endpoint read | color | | agglutination pattern |
| Cutoff(s) | 300 ng/mL | | 300 ng/mL |
| Reagent (active<br>ingredients) | 1. Blue dyed microparticles coated<br>with mouse monoclonal anti-<br>benzoylecgonine antibody<br>2. Drug conjugates immobilized on a<br>membrane<br>3. Mouse monoclonal anti-BSA<br>antibody immobilized on a<br>membrane | | 1. Mouse monoclonal anti-<br>benzoylecgonine antibody in<br>a buffered solution<br>2. Reaction buffer<br>3. Latex- benzoylecgonine<br>conjugate in a buffered<br>solution |
| Controls | ONTRAK TESTCUP Positive & Negative<br>Controls | | Abuscreen ONTRAK for Cocaine<br>Negative Control |
| Performance Characteristics: | | | |
| Precision | > 95% confidence at 150% cutoff | | > 99% confidence at 200% of<br>cutoff |
| Accuracy:<br>Positive Samples | | TESTSTIK<br>ONTRAK | GC/MS<br>GC/MS |
| | + | 50<br>50 | 49<br>0 |
| | - | 0<br>0 | 0<br>0 |
{6}------------------------------------------------
| | ONTRAK TESTSTIK<br>for Morphine | | Abuscreen ONTRAK for<br>Morphine |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Methodology | Competitive microparticle capture<br>inhibition | | Competitive latex agglutination<br>inhibition |
| Measurement | Qualitative | | Qualitative |
| Sample type | urine | | urine |
| Endpoint read | color | | agglutination pattern |
| Cutoff(s) | 300 ng/mL | | 300 ng/mL |
| Reagent (active<br>ingredients) | 1. Blue dyed microparticles coated<br>with mouse monoclonal anti-<br>morphine antibody<br>2. Drug conjugates immobilized on a<br>membrane<br>3. Mouse monoclonal anti-BSA<br>antibody immobilized on a<br>membrane | | 1. Mouse anti- morphine<br>antibody in a buffered<br>solution<br>2. Reaction buffer<br>3. Latex-morphine conjugate in<br>a buffered solution |
| Controls | ONTRAK TESTCUP Positive & Negative<br>Controls | | Abuscreen ONTRAK for<br>Morphine Negative Control |
| Performance Characteristics: | | | |
| Precision | > 95% confidence at 150% cutoff | | > 99% confidence at 200% of<br>cutoff |
| Accuracy:<br>Positive Samples | | TESTSTIK<br>49 | ONTRAK GC/MS<br>50 GC/MS<br>54 |
0
l
-
0
0
## Table 5 - ONTRAK TESTSTIK for Morphine
{7}------------------------------------------------
## Table 6 - ONTRAK TESTSTIK for PCP
| | ONTRAK TESTSTIK<br>for PCP | | Abuscreen ONTRAK for PCP | | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------|-------|
| Methodology | Competitive microparticle capture inhibition | | Competitive latex agglutination inhibition | | |
| Measurement | Qualitative | | Qualitative | | |
| Sample type | urine | | urine | | |
| Endpoint read | color | | agglutination pattern | | |
| Cutoff(s) | 25 ng/mL | | 25 ng/mL | | |
| Reagent (active ingredients) | 1. Blue dyed microparticles coated with rabbit polyclonal anti-phencyclidine antibody<br>2. Drug conjugates immobilized on a membrane<br>3. Mouse monoclonal anti-BSA antibody immobilized on a membrane | | 1. Rabbit anti- phencyclidine antibody in a buffered solution<br>2. Reaction buffer<br>3. Latex- phencyclidine conjugate in a buffered solution | | |
| Controls | ONTRAK TESTCUP Positive & Negative Controls | | Abuscreen ONTRAK for PCP Negative Control | | |
| Performance Characteristics: | | | | | |
| Precision | > 95% confidence at 150% cutoff | | > 99% confidence at 200% of cutoff | | |
| Accuracy:<br>Positive Samples | | TESTSTIK | ONTRAK | GC/MS | GC/MS |
| | + | 50 | 50 | 50 | 52 |
| | - | 0 | 0 | 0 | 1 |
{8}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| | ONTRAK TESTSTIK<br>for THC | Abuscreen ONTRAK for THC | | | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|-------|----|
| Methodology | Competitive microparticle capture<br>inhibition | Competitive latex agglutination<br>inhibition | | | |
| Measurement | Qualitative | Qualitative | | | |
| Sample type | urine | urine | | | |
| Endpoint read | color | agglutination pattern | | | |
| Cutoff(s) | 50 ng/mL | 50 ng/mL | | | |
| Reagent (active<br>ingredients) | 1. Blue dyed microparticles coated<br>with mouse monoclonal anti-<br>cannabinoid antibody<br>2. Drug conjugates immobilized on a<br>membrane<br>3. Mouse monoclonal anti-BSA<br>antibody immobilized on a<br>membrane | 1. Mouse monoclonal anti-<br>cannabinoid in a buffered<br>solution<br>2. Reaction buffer<br>3. Latex-tetrahydrocannabinoid<br>conjugate in a buffered<br>solution | | | |
| Controls | ONTRAK TESTCUP Positive & Negative<br>Controls | Abuscreen ONTRAK for THC<br>Negative Control | | | |
| Performance Characteristics: | | | | | |
| Precision | > 95% confidence at 150% cutoff | > 99% confidence at 200% of<br>cutoff | | | |
| Accuracy:<br>Positive Samples | TESTSTIK | ONTRAK | GC/MS | GC/MS | |
| | + | 45 | 45 | 45 | 50 |
| | - | 0 | 0 | 0 | 0 |
## Table 7 - ONTRAK TESTSTIK for THC
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. James W. Haynes Regulatory Affairs Associate Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
OCT - 9 1997
Re: K973075 Trade Name: OnTrak TesTstik Assays Requlatory Class: II Product Code: DKZ Dated: Auqust 15, 1997 Received: August 18, 1997
Dear Mr. Haynes:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{10}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
510(k) Number (if known) Jan 9730 73
ONTRAK TESTSTIK for Amphetamines Device Name: ONTRAK TESTSTIK for Cocaine ONTRAK TESTSTIK for Morphine ONTRAK TESTSTIK for PCP ONTRAK TESTSTIK for THC
Indications for Use:
- 1. The ONTRAK TESTSTIK for Amphetamines is an in vitro diagnostic test intended for professional use for the qualitative detection of amphetamines in urine at or above a cutoff concentration of 1000 ng/mL.
Page 1 of 1
:
- 2. The ONTRAK TESTSTIK for Cocaine is an in vitro diagnostic test intended for professional use for the qualitative detection of cocaine metabolite in urine at or above a cutoff concentration of 300 ng/mL.
- 3. The ONTRAK TESTSTIK for Morphine is an in vitro diagnostic test intended for professional use for the qualitative detection of morphine in urine at or above a cutoff concentration of 300 ng/mL
- 4. The ONTRAK TESTSTIK for PCP is an in vitro diagnostic test intended for professional use for the qualitative detection of phencyclidine in urine at or above a cutoff concentration of 25 ng/mL.
- 5. The ONTRAK TESTSTIK for THC is an in vitro diagnostic test intended for professional use for the qualitative detection of cannabinoids in urine at or above a cutoff concentration of 50 ng/mL.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | |
| Prescription Use<br>(Per 21 CFR 801,109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.