DEPUY MOTECH PROFILE ANTERIOR THORACOLUMBAR PLATE SYSTEM

K973060 · Depuy, Inc. · KWQ · Nov 3, 1997 · Orthopedic

Device Facts

Record IDK973060
Device NameDEPUY MOTECH PROFILE ANTERIOR THORACOLUMBAR PLATE SYSTEM
ApplicantDepuy, Inc.
Product CodeKWQ · Orthopedic
Decision DateNov 3, 1997
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DePuy Motech Profile Anterior Thoracolumbar Plate System is intended for use in the treatment of thoracic and thoracolumbar (T1-L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

Device Story

System consists of metal plates, cancellous screws, bolts, and locking nuts for anterior/anterolateral spinal fixation. Plates (40-120mm) and screws (15-60mm) implanted by surgeons in clinical settings to stabilize thoracic/thoracolumbar spine (T1-L5). Provides mechanical support for spinal instability; graft screws allow fixation to strut grafts. Output is physical stabilization of vertebral segments. Benefits include correction of spinal deformities and stabilization of unstable segments.

Clinical Evidence

Bench testing only. Mechanical fatigue strength testing demonstrated performance at least equivalent to predicate MOSS Miami Titanium and 4.0mm systems.

Technological Characteristics

System components: Ti-6Al-4V alloy plates, cancellous screws, bolts, and locking nuts. Plate lengths 40-120mm; screw lengths 15-60mm. Mechanical fixation via bolt slots and threaded screw holes. Non-sterile, manual surgical instrumentation.

Indications for Use

Indicated for patients requiring anterior/anterolateral spinal stabilization in the thoracic and thoracolumbar region (T1-L5) due to instability from fracture, tumor, degenerative disc disease, scoliosis, kyphosis, lordosis, spinal stenosis, or failed prior surgery. Contraindicated for pedicular screw fixation/attachment.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### K973060 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS | NAME OF FIRM: | DePuy, Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46581-0988 | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | | NOV - 3 199 | | 510(K) CONTACT: | John D. Van Vleet<br>Manager, Clinical and Regulatory Applications<br>DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, IN 46581-0988 | | TRADE NAME: | DePuy Motech™ Profile™ Anterior Thoracolumbar<br>Plate System | | COMMON NAME: | Anterior Thoracic/Thoracolumbar Plate | | CLASSIFICATION: | 888.3060: Spinal intervertebral body fixation<br>orthosis | | DEVICE PRODUCT CODE: | 87 KWO | ### SUBSTANTIALLY EQUIVALENT DEVICES: DePuy Motech MOSS Miami Titanium Spinal System DePuy Motech MOSS Miami Spinal System -4.0mm Amset Anterior Locking Plate System ## DEVICE DESCRIPTION AND INTENDED USE: The DePuy Motech Profile Anterior Thoracolumbar Plate System consists of 10 thoracolumbar and 12 high thoracic plates of various lengths (40-120mm); 9 cancellous screws in varying lengths (20-60mm); 8 cancellous bolts in progressive lengths (25-60mm) with locking nuts and screws; and four bone graft screws of different lengths (15, 20, 25, 30mm). The anterior thoracolumbar plates are implanted using 2 cancellous bolts placed through the plate's bolt slots and fixed with locking nuts and screws, 2 cancellous screws placed through the plate's threaded screw holes, and an optional graft screw which can be placed through the plates's center slot to provide fixation between the plate and a strut graft if desired. The high thoracic plates are implanted using 4-6 cancellous screws, placed through the plate's screw holes. {2}------------------------------------------------ The DePuy Motech Profile Anterior Thoracolumbar Plate System is intended for use in the treatment of thoracic and thoracolumbar (T1-L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery. #### BASIS OF SUBSTANTIAL EQUIVALENCE: The DePuy Motech Profile Anterior Thoracolumbar (ATL) Plates are substantially equivalent to the DePuy Motech MOSS Miami Spinal System Titanium sub-system and 4.0mm sub-system in that all of these systems are intended for anterolateral fixation to the thoracic and thoracolumbar spine (T1-L5) for the same indications. Both the ATL plates and the MOSS Miami Titanium sub-system are manufactured from Ti-6A1-4V alloy. Mechanical testing shows that the fatigue strength of the ATL plates is, at least, equivalent to that of both the MOSS Miami Titanium sub-system and the MOSS Miami 4.0mm sub-system. The DePuy Motech Profile ATL Plates are also substantially equivalent to the Amset ALPS plates in that both systems are intended for anterior stabilization of the spine and the designs of the systems are similar. The major difference between these systems is that the Amset ALPS plates are manufactured from 316 LVM Stainless Steel. {3}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles a human figure embracing or protecting another, representing the department's mission of promoting health and well-being. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV - 3 1997 | Ms. Cheryl K. Hastings | | |---------------------------------|--| | Manager, Regulatory Submissions | | | DePuy®, Inc. | | | P.O. Box 988 | | | 700 Orthopaedic Drive | | | Warsaw, Indiana 46581-0988 | | K973060 Re: Profile™ Anterior Thoracolumbar Plate System Requlatory Class: II Product Code: KWQ Dated: October 17, 1997 Received: October 20, 1997 Dear Ms. Hastings: We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act You may, therefore, market your device system subject (Act) . to the general controls provisions of the Act and the limitations identified below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment would cause the device system to be adulterated under 501(f)(1) of the Act. FDA identifies that any device system, if intended for use in pedicular screw fixation/attachment, except for some limited indications, would be found not substantially equivalent and would be a class III device under Section 513(f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly: {4}------------------------------------------------ Page 2 - Ms. Cheryl K. Hastings - You may not label or in any way promote this device 1. system for pedicular screw attachment to, or fixation of the cervical, thoracic or lumbar vertebral column. Therefore, in order to prevent off-label promotion, the package insert must include the following statement, "WARNING: This device system is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine."; - All labeling for this device system, including the 2. package label, must state that there are labeling limitations. The package insert must prominently state that the device system is intended for the specific use(s) described in the enclosure only; and - Pedicular screw fixation/attachment to the cervical, 3. thoracic or lumbar vertebral column, except for limited indications, of any device system is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device system for pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಕ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical ( Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {5}------------------------------------------------ Page 3 - Ms. Cheryl K. Hastings FDA advises that the use of the subject device system and/or device components with any other device components but those identified in this 510(k) would require submission of a new 510 k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or other manufacturers', may also be required. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, As well as Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {6}------------------------------------------------ #### INDICATIONS # 510(k) Number (if known) Device Name: DePuy Motech™ Profile™ Anterior Thoracolumbar Plate System #### Indications for Use: The DePuy Motech Profile Anterior Thoracolumbar Plate System is intended for use in the treatment of thoracic and thoracolumbar (T1-L5) spine instability as a result of fracture, (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery. Concurrence of CDRH, Office of Device Evaluation Aceda Prescription Use X OR Over-The Counter Use (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...