K973000 · Alimenterics, Inc. · MSQ · Feb 26, 1998 · Microbiology
Device Facts
Record ID
K973000
Device Name
ALIMENTERICS LARA (TM) BREATH TEST SYSTEM
Applicant
Alimenterics, Inc.
Product Code
MSQ · Microbiology
Decision Date
Feb 26, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3110
Device Class
Class 1
Indications for Use
The Pylori-Chek Test System is intended for use with the LARA Laser Assisted Ratio Analyzer for the qualitative detection of urease associated with Helicobacter pylori infection in the human stomach and as an aid in the diagnosis of H. pylori infection in symptomatic adult patients. The Pylori-Chek Test system consists of a Pylori-Chek test kit for the collection of breath samples and a LARA Laser Assisted Ratio Analyzer for the measurement of the ratio of 13CO2 to 12CO2 in the breath sample. For use by healthcare professionals. To be administered under a physician's supervision.
Device Story
The Alimenterics LARA Breath Test System is a non-invasive diagnostic tool for H. pylori infection. The patient ingests 100 mg of 13C-labeled urea; if H. pylori is present, the bacteria's urease enzyme hydrolyzes the urea into ammonia and 13CO2, which is exhaled. The system collects breath samples pre-ingestion (baseline) and at 30 and 60 minutes post-ingestion. Samples are processed by the LARA (Laser Assisted Ratio Analyzer), which uses the laser optogalvanic effect to measure the 13CO2 to 12CO2 ratio. The analyzer compares post-ingestion ratios against the baseline to determine a positive result. Used in clinical settings under physician supervision, the system provides a printed report to assist healthcare providers in diagnosing H. pylori, offering a non-invasive alternative to endoscopic biopsy-based tests like the CLOtest.
Clinical Evidence
Clinical evidence includes a pivotal study (n=875) and a supplemental 'cold trap' study (n=397). Studies compared the LARA System against five reference standards (including CLOtest and histopathology). Pivotal study sensitivity: 91.29%-94.15%; specificity: 90.38%-91.21%. Supplemental study (modified device) sensitivity: 90.09%-93.21%; specificity: 98.04%-98.11%. ROC analysis established a diagnostic cutoff of 6.18, yielding sensitivity 92.92%-96.05% and specificity 95.97%-96.71%. No device-related adverse events reported.
Technological Characteristics
System comprises Pylori-Chek breath collection kit and LARA Laser Assisted Ratio Analyzer. Detection principle: laser optogalvanic effect to measure 13CO2/12CO2 isotope ratios. Operates at ambient temperatures. Includes desiccant-based water vapor removal. Calibration maintained for 8 hours via control sample. Standalone analyzer unit.
Indications for Use
Indicated for symptomatic adult patients to aid in the diagnosis of Helicobacter pylori infection via qualitative detection of urease in the stomach.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Predicate Devices
CLOtest
Submission Summary (Full Text)
{0}------------------------------------------------
## FEB 2 8 1998
## K973000 510(k) SUMMARY FOR THE ALIMENTERICS LARA BREATH TEST SYSTEM
## I. SYSTEM SPONSOR
## A. Sponsor Name and Address
Alimenterics, Inc. 301 American Road Morris Plains, NJ 07950 (973) 285-3100
## B. Official Correspondent and Address
Janet George Murnick, Ph.D. President and CEO Alimenterics, Inc. 301 American Road Morris Plains, NJ 07950 (973) 285-3100
## SYSTEM IDENTIFICATION II.
## A. Trade/Proprietary Name of the System
The Alimenterics LARA Breath Test System is comprised of two components:
- (1) the Pylori-Chek Breath Test Kit; and
- (2) the LARATM (Laser Assisted Ratio Analyzer) System
## Common Name of the System B.
13C-Urea breath test for the presence of Helicobacter pylori
## PREDICATE DEVICE III.
The CLOtest® is a slide with a dispersed gel, which contains urea and a pH indicator. A biopsy acquired during endoscopy is immersed in the CLOtest® gel. If urease activity (associated with H.pylori) is present in the specimen, ammonia is generated, causing a pH change. This pH change is detected by the change in color of the pH indicator within 24 hours.
{1}------------------------------------------------
## IV. BACKGROUND
Peptic ulcer disease is a chronic inflammatory condition of the stomach and duodenum. Despite the fact that the disease has relatively low mortality, it results in substantial suffering of those affected.
A strong association between H. pylori, chronic superficial gastrointestinal disease has been well established. H.pylori is associated with type B gastritis, 12 duodenal ulcer, 34 gastric ulcer, 56 gastric cancer", and non-Hodgkins lymphoma.7
H. pylori was first cultured successfully from human gastric mucosa in 1982. The organisms, spiral gram negative bacteria, are found in the human stomach between the gastric epithelium and the mucosa. Isolates implicated in the above mentioned diseased states are distinguished by the production of copious amounts of endogenous urea amidohydrolase (urease).200 The enzyme catalyzes the breakdown of urea to carbon dioxide and ammonia, which are absorbed into the bloodstream.
Several methods are employed to determine the presence of H. pylori in the gastointestinal tract. Histologic staining of biopsy tissue using various stains has been shown to give adequate results with a specificity of over 90%.11 Mucosal biopsy samples can be cultured using selective and non-selective enriched media. However, due to the exacting needs of the organism, culture is the least sensitive (70-80%)" of all available techniques. Direct detection of urease activity of biopsy specimens is achieved by placing tissue in Christensen's urea agar and observing a color change. Biopsy and its associated analytic techniques are invasive and not particularly helpful in screening. Non-invasive tests consist of serum assays for IgG and IgA antibodies against H. pylori. However, these are not indicative of current infection. Urea breath tests using the radioactive isotope "4C or the stable isotope 13C can detect current infection with H. pylori, and have been shown to be highly sensitive methods. 12.13
### V. DEVICE DESCRIPTION
The Alimenterics LARA Breath Test System is based on the ability of the H.pylori organism to produce large amounts of the enzyme urease, which hydrolyses urea to NHL and CO3, the latter being exhaled. Using 30-labeled urea, an increase in the ratio of 130, to "CO, over time is an indication for the presence of H. pylori. This change can be detected by the LARA System based on a comparison of pre- and post-ingestion measurements of the patient's breath.
The Alimenterics breath test procedure involves a pre-ingestion analysis of a sample of the patient's breath to determine the baseline ratios of '30, to '200, Following ingestion of a test meal and 100 mg 13C-labeled urea, breath samples are collected at thirty and sixty minutes thereafter. The samples are then introduced into the LARA analyzer which, based on the laser
2
{2}------------------------------------------------
optogalvanic effect, measures the ratio of 13CO2 to 12CO2 The results of the three readings, with respect to a calibration standard, are determined concurrently by the analyzer. A specified change in the ratio of 13CO2 to 12CO2 in the 30 or 60 minute samples as compared to the baseline sample constitutes a positive test. Test values are generated and a report is printed for each patient.
## INTENDED USE VI.
The Alimenterics Pylori-Chek Breath Test kit is intended for use with the LARA Breath Test System for the qualitative detection of urease associated with Helicobacter pylori as an aid in the diagnosis of H.pylori infection.
## SUBSTANTIAL EQUIVALENCE COMPARISON VII.
The Alimenterics LARA Breath Test System and the CLOtest® have the same basic intended use: to detect the presence of H.pylori in human gastric mucosa.
The same chemical reaction (hydrolysis of urea catalyzed by H. pylori urease) forms the basis of both tests. When H. pylori is present, urea introduced either by ingestion in the case of the LARA Breath Test System or by contact in the case of the CLOtest, hydrolyses the urease produced by the H. pylori, which in turn produces both CO2 and NH. As explained above, with the CLOtest®, this chemical reaction takes place in vitro using an endoscopic biopsy, if H. pylori is present, the generation of NH, causes a pH change, which is reflected by a change in the color of the pH indicator. With the Alimenterics LARA Breath Test System, the chemical reaction takes place in vivo, and results in an increase in 13002 in the patient's excreted breath which is detected with the LARA instrument.
A table comparing the technological characteristics of both systems follows:
| | Organism | Reagent | Detection of<br>Urea<br>Degradation | Physical<br>Safety | Time | Temp. |
|-----------------------------|-----------------|----------|----------------------------------------------------------------------|-------------------------------------------------------------------------------|--------------|---------------------------------------------|
| CLOtest® | <i>H.pylori</i> | Urea | Visual detection of<br>urea degradation:<br>Urea→NH₄→Color<br>Change | Requires<br>invasive tissue<br>sampling | 3-24<br>hrs. | Requires<br>incubation<br>at 35 or<br>37° C |
| Alimenterics<br>Breath Test | <i>H.pylori</i> | ¹³C-Urea | LARA detection of<br>excess ¹³CO₂:<br>¹³C-Urea→¹³CO₂ | No tissue<br>sampling; no<br>adverse effects<br>from ingestion<br>of ¹³C-Urea | 1 hr. | LARA<br>operates at<br>ambient<br>temps. |
{3}------------------------------------------------
## PERFORMANCE DATA VIII.
## Clinical Tests A.
Clinical studies were conducted using the LARA Breath Test System. The objectives and results of the Pivotal and Supplemental studies are summarized below:
## Pivotal Study ﺎ .
The objectives of the study were to evaluate the ability of the LARA Breath Test System to detect the presence of H. pylori infections in the gastrointestinal system and to evaluate the sensitivity and specificity of the System as compared to reference methods. One thousand forty eight subjects were enrolled at 10 clinical sites. Of the 1048 subjects, 875 were included in the analysis. All patients who ingested the 3C-urea solution were included in the safety analysis. Only four patients reported adverse events. None of the adverse events was considered to be device related. Based on the results in this study, a diagnostic cutoff of 7.88 was calculated with an indeterminate zone between 7.018 and 8.600 (which excluded 4.1% of the patients).
Data were analyzed using five different reference standards: the IDE protocol standard, the CDER standard, the CDRH standard, the CLOtest, and central histopathology. Overall study sensitivity ranged between 91.29% and 94.15%, specificity ranged from 90.38% to 91.21%, PPV ranged from 90.50% to 91.51%, and NPV ranged from 90.93% to 94.12%. PPVs above 91% and NPVs above 94% were obtained for the breath test using all five reference standards.
During this study, some breath samples did not contain enough CO, to permit LARA System analysis. The Pylori-Chek breath collector used a desiccant to remove water vapor from the breath samples. The dessicant was found to absorb both water vapor and CO2, The device was modified to improve water vapor removal without affecting CO, content. A supplemental study, the "cold trap" study was conducted to evaluate the modified device.
## 2. Supplemental ("Cold Trap") Study
The objectives of the study were to verify that the modified breath test system performed comparably to the device used in the Pivotal Study and to provide an independent test (challenge) of the sensitivity and specificity calculated using the diagnostic cutoff and indeterminate zone determined in the Pivotal Study. Four hundred thirty-two patients were enrolled in this study. Four hundred thirty-one patients attempted to take the Pylori-Chek Breath Test and, of those, 397 were included in the analyses.
{4}------------------------------------------------
Of the 431 patients, six patients reported adverse events. The adverse events were considered to be related to either the endoscopy procedure or to ingestion of the test meal which in some instances followed the procedure by as little as one hour.
Results using the modified device were compared to results obtained in the Pivotal Study using the previous device. In the Pivotal Study, 120 (3.9%) of the breath samples had low CO2 content, while in this study only 32 (2.5%) of the samples had low CO2. The device modification reduced the number of UAP samples due to low CO, content.
Using the diagnostic cutoff of 7.80 with an indeterminate zone of 7.01 to 8.600 (which excluded 3.2% of the patients), sensitivity ranged from 90.09% to 93.21% and specificity ranged from 98.04% to 98.11%, PPV ranged from 98.54% - 98.58%, and NPV ranged from 86.71% to 91.23% using the same five reference standards that were used in the Pivotal Study. These results were comparable to those obtained in the Pivotal Study where sensitivity ranged from 91.29% -94.15% and sensitivity ranged from 90.38% to 91.21%. The specificity increased by more than 7%. These results of an independent data set using any of the five reference standards show that the high sensitivity and specificity observed in the Pivotal Study could be replicated and even improved.
Based on the results in this study, an ROC analysis was performed to yield a cutoff to maximize total predictive value. Using these results, a cutoff of 6.18 was calculated with an indeterminate zone between 5.510 and 6.700 (which excluded 3% of the patients). Using the 6.18 cutoff, sensitivity ranged from 92.92% to 96.05%, specificity ranged from 95.97% to 96.71%, PPV ranged from 97.29% to 97.84%, and NPV ranged from 89.24% to 93.59%. The modified device will be used in the clinical product. Thus, results for the diagnostic cutoff, and indeterminate zone determination, obtained with this device should be used for clinical decisions.
## Nonclinical Tests B.
In addition to clinical tests. Alimenterics conducted several nonclinical tests that demonstrated the performance of the Alimenterics LARA Breath Test System. These tests included the following:
Storage of Breath Specimens. This study demonstrated that the breath collectors could retain a sufficient amount of CO2 from a breath sample for 30 days under typical storage conditions. The study established a 30 day shelf life.
Transport Study. This study demonstrated that the breath collectors could retain a sufficient amount of CO, from a gas sample after experiencing the extreme transport condition.
Control Stability. The first control stability study did not demonstrate that the breath collectors could retain a sufficient amount of CO, from a control sample for 60 days under
5
{5}------------------------------------------------
varied storage conditions. Because a control failure occurred during the first study, a second study is being conducted to demonstrate that breath collectors packaged in sealed pouches currently used in the final product could retain a sufficient amount of CO2 from a control sample for 60 days under varied storage conditions including hot, humid conditions and cold, low pressure conditions.
Carryover Study. The data showed that the batch carryover effect ranged from 0.67% to 2.13%, with an average of 1.67% and a standard deviation of .51%. This study demonstrated that samples with high 13C content do not affect the baseline result of the next patient specimen.
Verification of Calibration (Control Frequency) Study The study demonstrated that the LARA System can maintain calibration for an 8 hour period using one control sample.
Intersvstem Comparison. The study demonstrated that analysis of a control gas was reproducible using three different LARA Systems.
## C. Conclusions
The clinical studies demonstrate that the Alimenterics LARA Breath Test System performs comparably to other diagnostic methods for the presence of H.pylori in the human gastrointestinal tract, including the CLOtest® In the clinical studies of the Alimenterics LARA Breath Test System, there were no adverse events directly associated with the system. The breath test system is, thus, safer than those test methods, like the CLOtest®, that require a biopsy for evaluation. The nonclinical studies demonstrate that the Alimenterics LARA Breath Test System performs reliably under the anticipated conditions of usage.
## IX. BIBLIOGRAPHY
- Andersen LP, Holck S, Poulsen CO, Elsborg L, Justesen T. Campylobacter pyloridis in 1. peptic ulcer disease. Scand. J. Gastoenterol. 22: 219-224, 1987.
- Blaser MJ. Gastric Campylobacter-like organisms, gastritis and peptic ulcer disease. 2. Gastroenterol. 93: 971-83, 1987.
- Langenberg ML, Tytgat GNJ, Schipper MEI, Rietra PJGM, Zanen HC. Campylobacter-3. like organisms in the stomach of patients and healthy individuals. Lancet i: 1348, 1984.
- Marshall, B.J., Guerrant, R.L., Plankey, M.W., et al. Comparison of "C-urea breath test, 4. microbiology and histology for the diagnosis of Campylobacter pylori (Abstract). Gastroenterol. 94:A284, 1988.
{6}------------------------------------------------
- Graham DY, Klein PD, Opekun AR, Boutton TW. Effect of age on the frequency of ર્ડ . active Campylobacter pylori infection diagnosed by the [13C]urea breath test in normal subjects and patients with peptic ulcer disease. J. Infect. Dis. 157: 777-780, 1988.
- Marshall BJ, Warren JR. Unidentified curved bacilli in the stomach of patients with ર . gastritis and peptic ulceration. Lancet 1: 1311-15, 1984.
- Parsonnet J, Hansen S, Rodriguez L, Gelb AB, Warnke RA, Jellum E, Orentreich N, 7. Vogelman JH, Friedman GD. Helicobacter pylori infection and gastric lymphoma. The New England Journal of Medicine 330: 1267-1271, May 1994.
- Marshall BJ. Royce H. Annear D1. Goodwin CS, Pearman JW, Armstrong JA. Original 8. isolation of Campylobacter pyloridis from human gastric mucosa. Microbios. Lett. 25: 83-88, 1983.
- Hazell SL, Borody TJ, Gal A, Lee A. Campylobacter pyloridis gastritis I: detection of ರು. urease as a marker of bacterial colonization. Am. J. Gastroenterol. 82: 292-296, 1987.
- 10. Marshall BJ, Warren JR, Francis GJ, et al. Rapid urease test in the management of Campylobacter pyloridis-associated gastritis. Am. J. Gastroenterol. 82: 200-10, 1987.
- 1 1 . Helicobacter pylori in peptic ulcer disease. Consensus Development Conference Statement, National Institutes of Health: Feb. 7-9, 1994.
- 12. Rauws EAJ, Royen EAV, Langenberg W, Woensel JV, Vrij AA, Tytgat GNJ. "C-urea breath test in C. pylori gastritis. Gut 30: 798-803, 1989.
- 13. Eggers RH, Kulp A, Ludtke FE, Bauer FE. Characterization of the [13]urea breath test for the diagnosis of Campylobacter pylori infections. Stable Isotopes in Paedetriac Nutritional and Metabolic Research: 295-301, 1990.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 26 1998
Janet G. Murnick, Ph.D. Alimenterics Inc. 301 American Road Morris Plains, NJ, 07950
Re: K973000 Device: Alimenterics LARATM Breath Test System Regulatory Class: I Product Code: MSO (LYR) Dated: August 6, 1997 Received: August 12, 1997
Dear Dr. Murnick:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
In addition, we have determined that your product contains the following component subject to regulation as drugs: 13C-Urea (powder-100 mg).
Our substantially equivalent determination does not apply to the drug component (NDA 20-900) of your product. For-information-on applicable-Agency requirements regarding the " drug component, we suggest you contact:
> Mark Goldberger, M.D., M.P.H. Division Director Division of Special Pathogens and Immunologic Drug Products (HFD-590) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 827-2335
{8}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality Systems Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification only after the Center for Drug Evaluation and Research has approved the drug component of your product. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising, please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-302) at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
510(k) Number (if known): K973000
Device Name: Alimenterics LARA™ Breath Test System
Indications For Use:
The Pylori-Chek Test System is intended for use with the LARA Laser Assisted Ratio Analyzer for the qualitative detection of urease associated with Helicobacter pylori infection in the human stomach and as an aid in the diagnosis of H. pylori infection in symptomatic adult patients. The Pylori-Chek Test system consists of a Pylori-Chek test kit for the collection of breath samples and a LARA Laser Assisted Ratio Analyzer for the measurement of the ratio of 13CO2 to 12CO2 in the breath sample.
For use by healthcare professionals. To be administered under a physician's supervision.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Jtton
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K973000
Prescription Use
(Per 21 CFR 801.109)
Or
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.