K972880 · Vidco, Inc. · MSX · Oct 22, 1997 · Cardiovascular
Device Facts
Record ID
K972880
Device Name
MDP2000 MEDICAL DATA PROCESSOR
Applicant
Vidco, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Oct 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected. It accepts analog and/or digital physiological information from patient monitoring equipment and processes these signals to provide outputs to: - Color monitors for local and remote display of information - Chart recorders or printers for hardcopy records - Media for storage - Other devices to share information Examples of areas where the MDP2000 is used include ICU, CCU, PACU, Emergency, Telemetry, Stepdown and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values such as heart rate and systolic, mean, and diastolic pressure.
Device Story
The MDP2000 is a medical data processor that acts as an intermediary between patient monitoring equipment and output peripherals. It is not patient-connected. It accepts analog and digital physiological signals (e.g., ECG, blood pressure) from monitors and processes them for distribution. Outputs include video signals for local/remote color monitors, hardcopy data for printers/recorders, storage media, and data sharing with other devices. Used in clinical environments like ICU, CCU, and PACU; operated by healthcare staff to observe patient physiological status at one or more locations. The device facilitates centralized or remote monitoring, allowing clinicians to view waveforms and numeric values (heart rate, blood pressure) to support clinical decision-making and patient observation.
Clinical Evidence
Bench testing only. The device was validated against Engineering Instrument Specifications, including testing for compliance with ANSI/AAMI EC13 (1992) for heart rate meters and alarms. Electromagnetic compatibility was verified per EN60601-1-2. Safety testing was conducted per UL2601. Performance results were reported as equivalent to or better than predicate devices.
Technological Characteristics
Class I Type B device; processes analog/digital physiological signals. Features improved display resolution and user interface compared to predecessors. Conforms to UL2601 and EN60601-1-2 (EMC). Connectivity includes inputs for patient monitors and outputs for video, hardcopy, and data storage. No specific materials or software architecture details provided.
Indications for Use
Indicated for use in clinical settings including ICU, CCU, PACU, Emergency, Telemetry, and Stepdown units to display and record physiological data (ECG, blood pressure, heart rate) from patient monitoring equipment.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "VIDCO" in a bold, sans-serif font. The letters are outlined in black, giving them a hollow appearance. The letters are evenly spaced and aligned horizontally.
OCT 22 1997
K972880
Information Display & Recording Systems
July 31, 1997
## VIDCO MDP2000
### 510k SUMMARY
# 510k number:
Submitter: VIDCO, Inc. 6175 SW 112th Avenue Beaverton, OR 97008
Bradley A. Cohen Contact: Phone: (503) 641-1804 FAX (503) 641-1806
Device Name/Common Name: MDP2000 Medical Data Processor/Remote Display & Central Display
The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected. It accepts analog and/or digital physiological information from patient monitoring equipment and processes these signals to provide outputs to:
- Color monitors for display of information
- Chart recorders or printers for hardcopy records
- Media for storage
- Other devices to share information
The MDP2000 is substantially equivalent to current and previously marketed VIDCO products:
VIDCO model 580CD (K926162) VIDCO model 620CSR (K932894) VIDCO model 580CDR (K934093) VIDCO model 516YT (K926219) VIDCO model 520 (K926298)
The MDP2000 has no new questions of safety or effectiveness as it is performing essentially the same functions as the predicate devices. For the same rationale, no new hazards arise when comparing the MDP2000 to the predicate devices.
Technical characteristics of the MDP2000 are similar to the predicate devices. Primary differences include: Reduced physical size, improved display resolution capabilities (aesthetics), and improved user interface to simplify operation.
The MDP2000 has been tested to and conforms with UL2601, the domestic standard of IEC 601-1, as a Class I Type B device. Testing has been conducted by Intertek Testing Services (formerly Inchcape, Inc.), an NRTL (nationally recognized test laboratory), and Notified Body in the European Union. The MDP2000 is ETL listed indicating conformance to UL2601. The predicate devices, except 516YT and 520, were Underwriters Laboratory Listed under the now obsolete UL544.
The MDP2000 has been tested to and conforms to the requirements of EN60601-1-2 (IEC 601-1-2), Collateral Standard for Electromagnetic Compatibility. The testing was conducted by Northwest EMC which is an FCC, NIST, and NVLAP recognized testing facility in the United States and accepted and accredited internationally by agencies such as TÜV and NEMKO. The EMC performance of the MDP2000 exceeds the predicate devices.
Validation performed consists of bench testing of specifications as called out in the Engineering Instrument Specification that is an integral part of the Device Master Record. All tests were completed successfully and the MDP2000 meets it's specified performance including those related to ANSI/AAMI EC13 (1992) Heart Rate Meters & Alarms. The MDP2000 performance results are equivalent to or better than the predicate devices.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Image /page/1/Picture/3 description: The image shows the date October 22, 1997. The month is abbreviated to OCT. The day is 22 and the year is 1997. The text is in a simple, sans-serif font.
Mr. Bradley A. Cohen VIDCO, Inc. 6175 SW 112th Avenue 97008 Beaverton, Oregon
Re: K972880 MDP2000 Medical Data Processor Regulatory Class: II (two) 74 MSX Product Code: Dated: July 31, 1997 Received: August 5, 1997
Dear Mr. Cohen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to. 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{2}------------------------------------------------
Paqe 2 - Mr. Bradley A. Cohen
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### VIDCO MDP2000 510k
### INTENDED USE STATEMENT
**510k number:** K972880
Device Name: MDP2000 Medical Data Processor
The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected. It accepts analog and/or digital physiological information from patient monitoring equipment and processes these signals to provide outputs to:
- Color monitors for local and remote display of information
- Chart recorders or printers for hardcopy records
- · Media for storage
- Other devices to share information
Examples of areas where the MDP2000 is used include ICU, CCU, PACU, Emergency, Telemetry, Stepdown and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values such as heart rate and systolic, mean, and diastolic pressure. This ملكة الكم over (1) to sixteen (16) pedside moritir Block Diagram
Image /page/3/Figure/10 description: The image shows a block diagram of a system labeled "MDP2000". The diagram has three inputs on the left side labeled "ANALOG", "DIGITAL", and "CONTROL". On the right side, there are four outputs labeled "VIDEO", "HARDCOPY", "DATA", and "SOUND". The arrows indicate the direction of signal flow from the inputs to the outputs through the MDP2000 system.
Arth. A. Liarlowski
(Division Sign-Off) Division of Cardiovascular, and Neurological Device 510(k) Number
Prescription Use \$\frac{x}{(Per 21 CFR 801.109)}\$
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.