← Product Code [DQA](/productcode/DQA) · K972829

# CRICKET 2000 RECORDING PULSE OXIMETER (K972829)

_Respironics, Inc. · DQA · Oct 28, 1997 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K972829

## Device Facts

- **Applicant:** Respironics, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Oct 28, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Cricket 2000 is a recording pulse oximeter which measures functional blood oxygen saturation, pulse rate/interval and finger movement. Because of SpO2 and pulse rate alarm limits that can be set by the operator, it can be used for continuous monitoring. It also interfaces with a commercially available computer. The pulse oximetry, raw infrared and red waveform data, as well as the pulse rate data, can be viewed real-time on the computer using custom Windows -based software that is provided with the pulse oximeter. The software also enables the user to do some data analysis, which can be used to draw meaningful implications during sleep apnea analysis studies. The Cricket 2000 is primarily indicated for use with patients suspected of suffering from, or with potential to suffer from, episodes of hypoxemia. The Cricket 2000 is intended for patients weighing more than 30 kg in both the hospital/institutional environment (supervised) and the home (unsupervised) environment.

## Device Story

Cricket 2000 is a recording pulse oximeter measuring functional blood oxygen saturation (SpO2), pulse rate/interval, and finger movement. Device inputs include infrared and red light signals; transforms these into real-time waveform and pulse rate data. Output is displayed on a connected computer via custom Windows-based software. Used in hospital or home settings by clinicians or patients for continuous monitoring and sleep apnea analysis. Software enables data analysis to assist clinical decision-making regarding hypoxemia episodes. Benefits include continuous patient monitoring and diagnostic support for sleep-related breathing disorders.

## Technological Characteristics

Recording pulse oximeter; measures SpO2, pulse rate, and finger movement via infrared/red light sensors. Interfaces with external computer. Software-based data processing and visualization. Intended for patients >30 kg.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Vitalog VX4 Recording Pulse Oximeter ([K935510](/device/K935510.md))
- HMS series pulse oximetry component ([K914620](/device/K914620.md)/A)
- HMS series pulse oximetry component ([K914085](/device/K914085.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K972829

Image /page/0/Picture/1 description: The image shows the logo for Respironics. The logo consists of a triangle shape with a stylized figure inside. Below the triangle is the word "RESPIRONICS" in bold, sans-serif font.

OCT 28 1997

100 | Murry Ridge Drive Murryvilla, Pannsylvania 1568-8550 USA Telephous: 1-112-735-0200 Far: 1-412-733-0299

#### 510(K) Summary . Cricket 2000 Pulse Oximeter 10/23/97

## Reason for 510(k):

Modification to existing device, primarily replacing the DOS-based software with the Windows-based User Interface Software.

## Device Description/ Intended Use/Indications for Use/Pationt Population:

The Cricket 2000 is a recording pulse oximeter which measures functional blood oxygen saturation, pulse rate/interval and finger movement. Because of Sp()2 and pulse rate alarm limits that can be set by the opcrator, it can be used for continuous monitoring. It also interfaces with a commercially available computer. The pulse oximerry, raw infrared and red waveform data, as well as the pulse rate data, can be viewed real-time on the computer using custom Windows -based software that is provided with the pulse oximeter. The software also enables the user to do some data analysis, which can be used to draw meaningful implications during sleep apnea analysis studies. The Cricket 2000 is primarily indicated for use with patients suspected of suffering from, or with potential to suffer from, episodes of hypoxemia. The Cricket 2000 is intended for patients weighing more than 30 kg in both the hospital/institutional environment (supervised) and the home (unsupervised) environment.

### Predicate Devices:

The predicate devices for the Cricket 2000 are the Vitalog VX4 Recording Pulse Oximeler (K935510) and the pulse oximetry component of the HMS series (K914620/A and K914085).

{1}------------------------------------------------

Image /page/1/Picture/10 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread. The eagle is facing to the left.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 28 1997

Mr. Francis X. Dobscha Respironics, Inc. 1001 Murry Ridge Drive Murrysville, Pennsylvania 15668-8550

Re : K972829 Cricket 2000 Pulse Oximeter Requlatory Class: II (two) Product Code: 74 DQA Dated: July 29, 1997 July 30, 1997 Received:

Dear Mr. Dobscha:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{2}------------------------------------------------

Page 2 - Mr. Francis X. Dobscha

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its = = toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html", . . . . .

Sincerely yours,
Thomas J. Callehan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page _1_ of

510(k) Number (if known): K_i5510 modification

Device Name: Cricket 2000 (Pulse Oximeter)

# Intended Use/Indications fur Use

Intended Use/Indications for Use/Patient Population: The Cricket 2000 is a recording pulse oximeter which measures functional blood oxygen saturation, pulse rate/interval and finger movement. Because of SpO2 and pulse rate alarm limits that can be sct by the operator, it can be used for continuous monitoring. It also interfaces with a commercially available computer. The pulse oximetry, raw infrared and red waveform data, as well as the pulse rate data, can be viewed real-time on the computer using custom Windows -based software that is provided with the pulse oximeter. The software also enables the user to do some data analysis, which can be used to draw meaningful implications during sleep apnca analysis studies. The Cricket 2000 is primarily indicated for use with patients suspected of suffering from, or with potential to suffer from, episodes of hypoxemia. The Cricket 2000 is intended for patients weighing more than 30 kg in both the hospital/institutional environment (supervised) and the home (unsupervised) environment.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE UN ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)            |                          |
|-------------------------------------------------------------------|--------------------------|
| Christy Foreman for                                               |                          |
| (Division Sign-Off)                                               |                          |
| Division of Cardiovascular, Respiratory, and Neurological Devices |                          |
| Prescription Use<br>(Per 21 CFR 801.109)                          | <span></span>            |
|                                                                   | OR                       |
| 510(k) Number                                                     | K972829                  |
| Over-The-Counter Use                                              |                          |
|                                                                   | (Optional Format 1-2-96) |

---

**Source:** [https://fda.innolitics.com/device/K972829](https://fda.innolitics.com/device/K972829)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
