RAICHEM PREALBUMIN SPIA

K972812 · Hemagen Diagnostics, Inc. · DDS · Aug 28, 1997 · Immunology

Device Facts

Record IDK972812
Device NameRAICHEM PREALBUMIN SPIA
ApplicantHemagen Diagnostics, Inc.
Product CodeDDS · Immunology
Decision DateAug 28, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5060
Device ClassClass 1

Indications for Use

The Prealbumin SPIA™ is a quantitative turbidimetric assay intended for the detection of prealbumin levels in human serum and plasma. Measurement of prealbumin levels may aid in the assessment of an individual's nutritional status.

Device Story

Prealbumin SPIA is a quantitative turbidimetric assay for measuring prealbumin in human serum or plasma. The device utilizes a polymer diluent and polyclonal anti-prealbumin antibodies; these form a complex with prealbumin in the sample, creating turbidity. The change in optical density is measured and compared to a standard curve to determine prealbumin concentration. The assay is designed for manual use or adaptation to automated clinical chemistry analyzers (e.g., Roche COBAS-MIRA). It is intended for use by laboratory professionals in clinical settings. Results assist healthcare providers in assessing a patient's nutritional status. The device provides a diagnostic tool for evaluating protein levels, aiding clinical decision-making regarding nutritional intervention.

Clinical Evidence

Bench testing only. Precision studies (inter-assay and intra-assay) performed on 8 serum samples showed %CVs ranging from 1.4% to 6.7%. Method comparison (n=134) against the predicate device showed high linear correlation. Comparison between manual and automated methods (n=30) yielded r2 = 0.959. Assay sensitivity (limit of detection) is 0.9 mg/dL. Interference testing confirmed no significant effect from hemoglobin (≤ 500 mg/dL) or bilirubin (≤ 20 mg/dL); lipids > 200 mg/dL showed significant interference.

Technological Characteristics

Quantitative turbidimetric assay. Reagents: polymer diluent, polyclonal anti-human prealbumin antibodies, calibrators, and controls. Standardized to CAP Reference Preparation for Proteins in Human Serum (RM002). Dynamic range: 0 to 55 mg/dL. Designed for manual or automated (e.g., COBAS-MIRA) use. Detection principle: immunoprecipitin-induced turbidity measured via optical density change.

Indications for Use

Indicated for the quantitative measurement of prealbumin levels in human serum or plasma to aid in the assessment of nutritional status.

Regulatory Classification

Identification

A prealbumin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the prealbumin (a plasma protein) in serum and other body fluids. Measurement of prealbumin levels in serum may aid in the assessment of the patient's nutritional status.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 4 10012 # 000001 ### 510(k) Summary ## AUG 28 1997 ### Submitter's Name/Contact Person 1. Joseph M. Califano Manager, Regulatory Affairs ### Address Hemagen Diagnostics, Inc. 34-40 Bear Hill Road Waltham, MA, 02154 (617) 890-3766 Phone: (617) 890-3748 Fax: icalifano@hemagen.com email: ### Date Prepared 25 July 1997 ### Device Name 2. | Trade Name: | Prealbumin SPIATM | |----------------------|-----------------------------------------| | Common Name: | Prealbumin | | Classification Name: | System, Test, immunological, prealbumin | ### Predicate Device 3. Incstar Antibody Reagent Set II for Prealbumin {510 (k) Docket No. K 884297} ### За. Methods Described in Immunoturbidimetry of transthyretin (prealbumin) in Manual method: human sera. Clin. Chem: 33: 7, 1260 1987. Ledue T.B., Rifai N, Irish, G.R., Silverman, L.M. COBAS-MIRA Analyzer. {510 (k) Docket No. K 851172} Automated System: ## 100000 , 510(k) Summary Page 1 {1}------------------------------------------------ ### 4. Description of Device Raichem's Prealbumin SPIA ™ is a quantitative turbidimetric assay for the detection and measurement of prealbumin in human serum and plasma. The assay has been standardized to a CAP Reference Preparation. The assay reagents consist of a polymer diluent, a polyclonal antibody to human prealbumin, calibrators, and controls. In this assay, a complex forms between the prealbumin, and anti-prealbumin antibodies causing turbidity. The change in optical density is proportional to the amount of prealbumin present. A quantitative determination of the amount of prealbumin present in a serum/plasma sample is made by comparison to a standard curve. ### 5. Intended Use of Device The Prealbumin SPIA™ is a quantitative turbidimetric assay intended for the detection of prealbumin levels in human serum and plasma. Measurement of prealbumin levels may aid in the assessment of an individual's nutritional status. #### Technological Characteristics 6.(A) ### Proposed Device Raichem's Prealbumin SPIA ™ is a quantitative turbidimetric assay. This assay is performed manually following clinically accepted methodologies. The assay is designed to enable users to readily adapt it for use with automated systems such as the Roche COBAS MIRA Analyzer. ### Predicate Device Incstar's Antibody Reagent Set II for Prealbumin is also a quantitative turbidimetric assav that utilizes immunoprecipitin analysis for the determination of prealbumin levels. . {2}------------------------------------------------ ### Performance Data 6.(B) #### Precision . To evaluate precision, inter-assay and intra-assay studies were conducted with the Raichem's Prealbumin SPIA ™ on an automated system {COBAS-MIRA} ### A. Inter-assay reproducibility Eight different serum samples were assayed ten times over six different days. | SAMPLE | Mean mg/dL | Std. Dev | % CV | Mean Delta | Std. Dev | % CV | |--------|------------|----------|------|------------|----------|------| | 1 | 36.3 | 1.9 | 5.2 | 0.272 | 0.007 | 2.7 | | 2 | 17.1 | 1.1 | 6.6 | 0.188 | 0.007 | 3.5 | | 3 | 7.7 | 0.4 | 5.6 | 0.115 | 0.004 | 3.3 | | 4 | 14.1 | 0.9 | 6.4 | 0.168 | 0.006 | 3.7 | | 5 | 22.3 | 1.1 | 4.8 | 0.217 | 0.004 | 1.9 | | 6 | 26.3 | 1.2 | 4.7 | 0.237 | 0.004 | 1.9 | | 7 | 36.7 | 2.5 | 6.7 | 0.273 | 0.008 | 2.9 | | 8 | 50.4 | 2.5 | 4.9 | 0.305 | 0.007 | 2.2 | ### B. Intra-assay reproducibility The eight serum samples were also assayed 20 consecutive times in a single run. | SAMPLE | Mean mg/dL | Std. Dev | % CV | Mean Delta | Std. Dev | % CV | |--------|------------|----------|------|------------|----------|------| | 1 | 34.3 | 1.6 | 4.8 | 0.280 | 0.006 | 2.2 | | 2 | 17.4 | 0.7 | 4.1 | 0.195 | 0.003 | 1.7 | | 3 | 8.0 | 0.2 | 2.6 | 0.114 | 0.002 | 1.8 | | 4 | 13.7 | 0.8 | 5.7 | 0.159 | 0.006 | 3.7 | | 5 | 22.7 | 0.6 | 2.5 | 0.214 | 0.003 | 1.4 | | 6 | 28.6 | 1.0 | 3.6 | 0.240 | 0.004 | 1.7 | | 7 | 34.7 | 1.4 | 4.0 | 0.263 | 0.005 | 1.8 | | 8 | 50.7 | 2.7 | 5.3 | 0.304 | 0.006 | 1.8 | #### Standardization of the Calibrators 11. The set of 5 calibrators supplied with the assay have been standardized within the dynamic range { 0 to 55 mg/dL} using the College of American Pathologists Reference Preparation for Proteins in Human Serum, Catalog Number RM002. ### 111. Assay Sensitivity The detection limit of the assay was determined by running multiple replicates the 0 mg/dL Calibrator and computing the mean and standard deviation. The resultant detection limit was found to be 0.9 mg/dL. {Mean + 2SD} 510(k) Summary Page 3 -ﻳﺤﻴﺔ - {3}------------------------------------------------ #### Comparison Studies IV. - The Prealbumin SPIA ™ and the Incstar Antibody Reagent Set II for Prealbumin a. were used to assay serum specimens from individuals being screened for prealbumin levels and apparently healthy blood donors. The specimens were assayed concurrently with both the proposed and predicate devices, using an automated method {COBAS-MIRA} A total of 134 samples were evaluated. The results indicated a high degree of linear correlation between the proposed and predicate device. The resultant linear regression relationship is: YPROPOSED = 1.06 XPREDICATE = 1.15, #### Comparison of a manual method and an automated method{COBAS-MIRA} b. The Raichem Prealbumin SPIA ™ was used to assay serum specimens concurrently by a manual method and an automated method {COBAS-MIRA} A total of 30 samples were evaluated. The results indicated a high degree of linear correlation between the manual and automated methods. The resultant linear regression relationship is: VAUTOMATED = 1.10 XMANUAL + 1.07, r2 = 0.959 #### V Assay performance with Serum and Plasma Twenty five (25) matched serum and EDTA-plasma samples were compared. Half of the volume of each sample was converted to serum by recalcification using a standard Ca2"/thrombin methodology. Each of the plasma and converted serum samples were evaluated with the proposed device on an automated system {COBAS-MIRA}. The results of the evaluation with the proposed device indicate that it can provide accurate estimates of prealbumin levels in both human serum and EDTA-plasma. #### VI. Interfering Substances Lipemic, hemolytic, and icteric samples were evaluated with the assay. The results indicate that there is no significant effect (< 20 % variation) on the assay for samples with: | Hemoglobin concentration: | ≤ 500 mg/dL | |---------------------------|-------------| | Bilirubin concentration: | ≤ 20 mg/dL | Lipid concentrations of > 200 mg/dL showed significant decreases in prealbumin values. #### 7. Conclusion The results of the comparative studies support the claim that the Prealbumin SPIA ™ is substantially equivalent to the predicate device. {4}------------------------------------------------ ***_*****_**, and the list goes on. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 AUG 28 1997 Mr. Joseph M. Califano Manager, Regulatory Affairs Hemagen Diagnostics, Inc. 34-40 Bear Hill Road ... ...... Waltham, Massachusetts 02154 Re : K972812 Trade Name: Prealbumin SPIA™ Requlatory Class: I Product Code: DDS Dated: July 25, 1997 Received: July 28, 1997 Dear Mr. Califano: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic {QS} inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you miqht have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {5}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Device Name: Prealbumin SPIA ™ Indication(s) For Use The use of these reagents is indicated for the measurement of prealbumin levels in human serum or plasma. (PLEASE DO NOT WRITE BELOW THIS LINE) Concurrence of CDRH, Office of Device Evaluation (ODE) **Prescription Use** (Per 21 CFR 801.109) OR Over-The-Counter-Use_ Peter S. Makim (Division Sign-Off) Division of Clinical Laboratory Devices 610(k) Number ___
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...