HemaPrompt is a gualac-based in-vilro slide method for the qualitative detection of occult blood in feces and gastric aspirate or vomitus by medical professionals only. For fecal testing, it is a useful ald In the diagnosis of a number of gastrointestinal disorders, and is recommended for use in 1) routine physical examinations 2) routine hospital testing 3) screening for colorectal cancer or gastro-Intestinal bleeding from any other source. For the testing of gastric contents it is used for the early detection of occult blood in conditions such as gastric trauma, gastric or duodenal ulceration, gastric cancer, esophageal varices, situations of likely exogenous or endogenous gastrills, leukemia, and hereditary telangiectasia. These conditions may be encountered in the emergency room, recovery room or intensive care.
Device Story
HemaPrompt is a guaiac-based slide test for qualitative detection of occult blood in fecal and gastric samples. Input: fecal or gastric specimen applied to guaiac-impregnated paper. Operation: hemoglobin in sample reacts with guaiac; buffered developer (alcohol/hydrogen peroxide) applied via pull-tab triggers peroxidase-like reaction; blue color indicates positive result. Used in clinical settings (ER, recovery, ICU) by medical professionals. Output: visual color change (blue). Clinical utility: preliminary screening aid for GI bleeding/pathology; results guide further diagnostic procedures like endoscopy or X-ray. Benefits: rapid, reliable detection of hemoglobin concentrations >2 mg/g stool or >200 mcg/ml gastric juice.
Clinical Evidence
Bench testing only. Fecal study: evaluated sensitivity using homogenized blood-spiked stool samples; detected hemoglobin at concentrations ≥2 mg/g. Gastric study: evaluated 12 patient samples, 8 volunteer samples, and synthetic gastric juice (PBS) across pH 1.0-7.0; detected hemoglobin at concentrations ≥200 mcg/ml. Reproducibility confirmed over two weeks with samples stored at 5°C. Interference testing showed no impact from iron, ranitidine, cimetidine, or antacids at normal therapeutic concentrations.
Indicated for medical professionals to detect occult blood in feces and gastric aspirate/vomitus for patients undergoing routine physicals, hospital testing, or screening for colorectal cancer, GI bleeding, gastric trauma, ulcers, cancer, esophageal varices, gastritis, leukemia, or hereditary telangiectasia.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Submission Summary (Full Text)
{0}------------------------------------------------
HEMAPROMPT for use in detection of gastric and fecal occult blood. 510K SUMMARY OF
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
The HemaPrompt slide, used with a buffered developer, is a guaiac based test for the detection of occult blood in both gastric and fecal samples. The test is not affected by low pH. When used with gastric specimens, it is free from interferences by normal therapeutic concentrations of iron, ranitidine, cimetidine and antacids.
When a gastric or fecal specimen containing blood is applied to the HemaPrompt test paper, hemoglobin from the lysed blood cells in the sample comes in contact with the guaiac in the paper. The developer, a buffered solution of alcohol and hydrogen peroxide, is applied by pulling the tab on the slide and creates a guaiac / peroxidase-like reaction in the presence of hemoglobin which in turn causes a blue coloration in the test paper. This test will turn blue in the presence of more than 2 mg Hb/ G stool and more than 200 mcg Hb/ ml gastric juice
As with any occult blood test, results with HemaPrompt cannot be conclusive evidence of the presence or absence of upper or lower bowel bleeding or pathology. HemaPrompt is designed for use as a preliminary screening aid and is not intended to replace other diagnostic procedures such as endoscopy or Xray procedures.
## EXPECTED AND PERFORMANCE RESULTS
The use of guaiac impregnated paper for the detection of fecal occult blood has been extensively studied. These clinical studies indicate that guaiac impregnated slide tests yield a positive result 3-5% of the time in screening programs and the percent of false positive results lies in the range of 1-2% of persons on a controlled diet i.e. a diet excluding substances such as raw/ rare meat, and raw peroxidase-containing vegetables such as broccoli and cauliflower, with the normal daily fecal blood loss in an adult lying in the range of 1-2 mls. of blood per 100 gram of stool. Sensitivity (% of subjects with the condition being sought who test positive) is difficult to estimate, but in series of patients with known colorectal cancer, 50-87% have been reported to yield positive reactions. Estimates of positive reactions with adenomatous bleeding have varied widely, and appear dependant to a degree on the size of polyp, with polyps less than 2 cms yielding less than 5% positive reactions.
Feces Study: Stool from a healthy volunteer on a diet as described was used as a baseline specimen and assumed to contain 1 ml of blood / 100 G of stool which is about the normal daily blood loss of an adult. Blood (Hb 14G/dl by Coulter) was diluted in the stool by homogenization to provide concentrations of 0.1, 0.2, 0.4 and 0. 6 grams hemoglobin per 100 grams of stool. In an anemic person with a hemoglobin of 10G/dl these levels would be achieved with approximately 1, 2, 4 and 6 mls. of blood per 100 grams of stool. These dilutions were used to test the sensitivity of HemaPrompt as well as to compare the reactions to those obtained with another commercially available guaiac slide test concurrently with HemaPrompt on the same specimens four times on three occasions over a period of twelve days, with precautions taken to avoid specimen desiccation.
HemaPrompt reacted positively at total hemoglobin concentrations of 2mg/G of stool ( i.e. about twice the expected normal blood loss) or greater, in 60 seconds, with no reactions below this concentration. Hydrating stool (unless dried by adding distilled water may produce reactions below this level (2mg/G) by eluting blood out of the sample onto the test paper resulting in a possible false positive reaction.
It was concluded that with stool, HemaPrompt reacted reliably and definitely to all hemoglobin levels above 2mg/G of stool. It was also concluded that HemaPrompt with a buffered developer reacted in the same way as another commercially available guaiac test with buffering.
The significance of gastric occuit blood has been less extensively studied than fecal occult blood. One study of 153 gastric aspirates from 50 intubated healthy adults indicated all aspirates with more than 50 micrograms of hemoglobin/ ml were positive with a buffered guaiac impregnated test paper. There was an apparent overall false positive rate of 25.5% in this study of normal intubated individuals, but even using less than 25 micrograms of hemoglobin / ml. as the test cut-off. 11.8% showed a positive reaction. The positive rate will be affected by the method of collection. A traumatic
{1}------------------------------------------------
intubation can be expected to produce some degree of bleeding .
Gastric Juice Study: HemaPrompt was studied with
- a) 12 patient gastric samples obtained by gastroscopy.
...
- b) 8 healthy volunteer gastric samples obtained by intubation.
- c) samples of a synthetic gastric juice of Phosphate Buffered Saline (PBS) each titrated with HCI to give pH's ranging from 1.0 to 7.0 and hemoglobin concentrations of 50, 100, 200 and 500 mog/ml.
(a) It was found of the patient samples, none were positive initially, 37.5% (3/8) were positive with blood added to produce a concentration of 50 mcg / ml Hb, and 66% (8/12) were positive with 100 mcg / ml Hb and 100% of samples with blood added in concentrations of 200 mcg Hb / ml or greater reacted positively.
(b) Of the volunteer samples, 37.5% (3/8) were positive with no added blood. 60% (3/5) were positive with 20 mcg/ml Hb added, 87.5% (7/8) were positive with 50 mcg/ml Hb added to the specimen. and 100% were positive at concentrations of 100 mcg / ml and greater of added hemoglobin. All samples with blood added in concentrations of 200 mcg Hb / ml or greater reacted positively, and all reacted in less than 60 seconds. The positive rate will be affected by the method of collection. A traumatic intubation can be expected to produce some degree of bleeding. An initial negative result under these circumstances assumes added weight compared to a positive result ..
(c) Of the PBS samples, 50 mcg Hb/ mi produced a positive reaction 44% (7/16) of the time, with 100 mcg / ml 81.25% (13/16) of samples showed a positive reaction. At 200 mcg / ml and above all samples showed a positive reaction. Furthermore, concentrations of ranitidine, ferrous, suffate, and an antacid (Mylanta) to be expected in the stomach after a maximum recommended dosing did not alter the HemaPrompt results. This does not necessarily apply in overdose situations when excessive iron compounds and certain H2 blockers (e.g. cimetidine/Tagamet) could produce false positive reactions, and excess antacid or Zantac (ranitidine) could produce a false negative.
The samples from each person were repeated over a two week period. Results demonstrated excellent HemaPrompt reproducibility at levels above 200 mog/ ml gastric juice with all samples stored up to 10 days at 5℃ and showed excellent comparison to results obtained on the same samples with another commercially available test for this purpose
It was concluded that with gastric juice. HemaPrompt reacted reliably and definitely to hemoglobin levels above 200mcg / ml. gastric juice.
All monitors reacted in the expected manner (+ve turned blue ). Exposing the gualac paper to UV light for ten minutes inactivated the expected reaction.
{2}------------------------------------------------
OCT 2 0 1997
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with three intertwined lines, representing the three branches of government, and a stylized eagle's head.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Robert Schreiber, M.D. CEO Aerscher Diagnostics, Inc. 359 High Street Chestertown, Maryland 21620
Re : K972763 Trade Name: HemaPrompt for Fecal and Gastric Occult Blood Regulatory Class: II Product Code: KHE Dated: July 22, 1997 Received: July 24, 1997
Dear Dr. Schreiber:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and 594-4588. advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1 -
510(k) Number (if known):_ K972763
Device Name:__________________________________________________________________________________________________________________________________________________________________
## Indications For Use:
HemaPrompt is a gualac-based in-vilro slide method for the qualitative detection of occult blood in feces and gastric aspirate or vomitus by medical professionals only.
For fecal testing, it is a useful ald In the diagnosis of a number of gastrointestinal disorders, and is recommended for use in 1) routine physical examinations 2) routine hospital testing 3) screening for colorectal cancer or gastro-Intestinal bleeding from any other source.
For the testing of gastric contents it is used for the early detection of occult blood in conditions such as gastric trauma, gastric or duodenal ulceration, gastric cancer, esophageal varices, situations of likely exogenous or endogenous gastrills, leukemia, and hereditary telangiectasia. These conditions may be encountered in the emergency room, recovery room or intensive care.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | | Benz Book Book Book Barner Barner Barner Barner Barner Barner Bronner Bronner Bergen Berger | Caroline of Children of Children Comments of Children and | Specificant of Summer of Summer of Summers |
|---------------------------------------------------------|--|---------------------------------------------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------|
| Concurrence of CDRIX, Office of Dovice Evaluation (ODF) | | | | |
Cita E. Mayumi
(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number
**Prescription Use** √
(Per 21 CFR 801.109)
"
()[{
Over-The-Counter Use --
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.