QUICKVUE ONE-STEP HCG-URINE

K972742 · Quidel Corp. · JHI · Aug 29, 1997 · Clinical Chemistry

Device Facts

Record IDK972742
Device NameQUICKVUE ONE-STEP HCG-URINE
ApplicantQuidel Corp.
Product CodeJHI · Clinical Chemistry
Decision DateAug 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

The QuickVue® One-Step hCG-Urine is a one-step immunoassay intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. The QuickVue One-Step hCG-Urine is intended for use by health care professionals.

Device Story

QuickVue One-Step hCG-Urine is a lateral-flow immunoassay for qualitative hCG detection in urine. Device utilizes monoclonal and polyclonal antibodies to detect hCG, a placental hormone. Operation involves applying urine sample to test; blue procedural control line appears to confirm test validity; pink-to-purple test line appears if hCG is present. Results are read visually by healthcare professionals at three minutes. Used in physician office laboratories and clinical settings to confirm pregnancy. Provides rapid, qualitative diagnostic information to assist clinicians in pregnancy management.

Clinical Evidence

Bench testing and clinical comparison studies performed. Accuracy exceeded 99% in direct comparison to other FDA-cleared qualitative hCG devices. Intra- and inter-assay precision demonstrated. Physician's Office Laboratory studies at three U.S. sites showed 100% agreement with expected results across diverse personnel.

Technological Characteristics

Lateral-flow immunoassay; utilizes monoclonal and polyclonal antibodies. Qualitative visual readout. No external energy source required. Standalone test format.

Indications for Use

Indicated for the qualitative detection of hCG in urine for early pregnancy detection in patients. Intended for use by healthcare professionals.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 0(k) SUMMARY OF SAFETY AND EFFECTIVENESS K 472742 Product: uickVue® One-Step hCG-Urine AUG 2 9 1997 Manufacturer: OUIDEL Corporation 10165 McKellar Court San Diego, CA 92121 U.S.A. ## ce Classification: evice, OuickVue One-Step hCG-Urine, is similar to other FDA-cleared s used for the qualitative detection of human chorionic gonadotropin in urine. The test is used in the early detection of pregnancy and is d to measure hCG, a placental hormone, in urine (21 CFR 862.1155). DA has proposed that hCG test systems be classified as Class II. #### ended Use: e test is a rapid immunoassay for the qualitative detection of hCG in urine. his test is to be used for the early detection of pregnancy. #### Physiologic Basis for the Assay: Human chorionic gonadotropin (hCG) is a glycoprotein hormone secreted by the trophoblastic cells of the developing placenta as early as 7 to 8 days after ovulation. This hormone stimulates the production of progesterone and estradiol which are required to sustain pregnancy. In normal pregnancy, serum levels of hCG continue to rise during the first trimester to levels as high as 100,000 mIU/mL. Serum hCG is rapidly cleared in the urine and the corcentration of hCG in serum is approximately equal to the concentration in urine. HCG is an excellent indicator of pregnancy early in the gestational period. {1}------------------------------------------------ # ciple of the Test: est, a lateral-flow immunoassay, employs monoclonal and polyclonal bodies. Shortly after addition of the urine sample, a blue procedural ol Line will appear. If hCG is present in the sample, a pink-to-purple Line will also appear. The result is read at three minutes. If no blue cedural Control Line develops, the result is considered invalid. # Safety and Effectiveness: Jumerous studies were undertaken to document the performance haracteristics and the substantial equivalence of the test to other ommercially available products for the qualitative detection of hCG in urine. hese studies included the following: - · The test was shown to be similar to other commercially distributed in vitro tests in terms of features and intended use. - · The test was shown to have excellent intra- and inter-assay precision. - · Lot-to-lot consistency analyses showed the test to be reproducibly manufacturable. - · Common drugs, chemicals, and biologicals were shown not to interfere with the test's performance. - · Using urine samples obtained from women presenting for pregnancy testing, a direct comparison of the test to other FDA-cleared devices for qualitative hCG detection was conducted. An accuracy exceeding 99% was observed. - · Physician's Office Laboratory studies were conducted to show that doctor's office personnel with diverse educational backgrounds and work experience could perform the test accurately and reproducibly. Testing was performed at three geographically distinct sites in the United States. The results obtained at each site agreed 100% with the expected results. #### Conclusion: These studies demonstrated the substantial equivalence of the test to existing products already marketed. They further demonstrated the suitability of the product for use by health care professionals. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Aarrinistration 2098 Gaither Road Rockville MD 20850 AUG 2 9 1097 Robin Weiner Vice President, Clinical Develop. & Reg. Affairs QUIDEL Corporation 10165 Mc Kellar Court San Diego, California 92121 K972742 Re: QuickVue® One-Step hCG-Urine Requlatory Class: II Product Code: JHI July 22, 1997 Dated: Received: July 23, 1997 Dear Ms. Weiner: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | | | Page <u> 1 </u> of <u> 1 </u> | |--|--|-----------------------------------------------------------------------| |--|--|-----------------------------------------------------------------------| | 510(k) Number (if known): | K972742 | |---------------------------|------------------------------| | Device Name: | QuickVue® One-Step hCG-Urine | Indications for Use: The QuickVue® One-Step hCG-Urine is a one-step immunoassay intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. The QuickVue One-Step hCG-Urine is intended for use by health care professionals. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K972742 | | Prescription Use (Per 21 CFR 801.109) | OR | Over-The Counter Use | |---------------------------------------|----|----------------------| |---------------------------------------|----|----------------------| (Optional Format 1-2-96) ಷ i
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...