ENDOPATH ENDOSCOPIC TISSUE FASTENING SYSTEM (ETFS)
Applicant
Ethicon Endo-Surgery, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Sep 19, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) is intended for use in minimally invasive surgical applications where soft tissue is being approximated with interrupted stitches. ETHIBOND EXCEL Polyester Suture is intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
ENDOPATH Endoscopic Tissue Fastening System (ETFS) facilitates endoscopic tissue approximation and intracorporeal knot placement. System components include SEW E-Z Reload Unit (knot/grasper element with ETHIBOND EXCEL polyester suture and curved needle), SEW E-Z Suture Assistant, and SEW-RIGHT Self Righting Needle Holder. Instrument features control lever for grasper operation and knot deployment lever with ratchet mechanism for controlled knot placement. Used in minimally invasive surgery by physicians; instrument allows reloading during single procedure (max 8 knot deployments). Device assists in rapid needle orientation and one-handed knot deployment, improving surgical efficiency and tissue approximation.
Clinical Evidence
Bench testing and pre-clinical laboratory evaluations confirmed device functionality. Studies demonstrated that the ETFS system facilitated laparoscopic suturing, allowed for one-handed knot deployment with adequate security, and enabled rapid needle orientation/management.
Technological Characteristics
System includes ETHIBOND EXCEL polyester suture (polyethylene terephthalate) with polybutilate coating. Instrument features mechanical levers and ratchet mechanism for knot deployment. Designed for use with 5.4 mm or larger trocars. Sterile, single-patient use components.
Indications for Use
Indicated for patients undergoing minimally invasive surgical procedures requiring soft tissue approximation with interrupted stitches, or general soft tissue approximation/ligation including cardiovascular, ophthalmic, and neurological applications.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Submission Summary (Full Text)
{1}------------------------------------------------
SEP | 9 |997
್ಕೆ
K972679
# Appendix A: 510(k) Summary of Safety and Effectiveness
| Statement | Information supporting claims of substantial equivalence, as defined under the<br>Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is<br>summarized below.<br><br>For the convenience of the Reviewer, this summary is formatted in<br>accordance with the Agency's final rule "...510(k) Summaries and 510(k)<br>Statements..." (21 CFR §807) and can be used to provide a substantial<br>equivalence summary to anyone requesting it from the Agency. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>description | The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) consists of<br>the SEW E-Z Reload Unit, the SEW E-Z Suture Assistant, and the SEW-<br>RIGHT Self Righting Needle Holder (see Instructions for Use for the<br>ENDOPATH® SEW-RIGHT Self Righting Needle Holder).<br><br>The SEW E-Z Reload Unit is a sterile, single patient use device consisting of<br>a knot/grasper element comprised of an ETHIBOND EXCEL polyester suture<br>(in sizes 0 or 2-0) with a pre-tied knot and a curved needle attached. The<br>knot/grasper element is contained inside a carrier for loading the SEW E-Z<br>Suture Assistant.<br><br>The SEW E-Z Suture Assistant is a sterile, single patient use instrument<br>designed to assist endoscopically in approximating tissue and in placing a<br>suture and a secure knot intracorporeally. It is designed for use with a 5.4<br>mm trocar or a larger trocar with use of reducer cap.<br><br>The instrument has two levers: the control lever facilitates loading the<br>instrument and opening and closing the grasper; the knot deployment lever<br>contains a ratchet mechanism for controlled deployment of the pre-tied knot.<br><br>The instrument contains a shaft retractor button on both sides of the<br>instrument for unloading and reloading knot/grasper elements. |
| | The instrument may be reloaded during a single procedure. Do not reload the<br>instrument for more than a total of eight knot deployments per instrument.<br><br>Continued on next page |
| | ETHIBOND EXCEL Polyester Suture is a nonabsorbable, braided, sterile,<br>surgical suture composed of Poly (ethylene terephthalate). It is prepared from<br>fibers of high molecular weight, long-chain, linear polyesters having recurrent<br>aromatic rings as an integral component. ETHIBOND EXCEL suture is<br>uniformly coated with polybutilate or poly [oxy-1, 4 butanediyloxy 91, 6<br>hexanediyl)]. |
| | The highly adherent coating is a relatively nonreactive nonabsorbable<br>compound which acts as a lubricant to mechanically improve the physical<br>properties of the uncoated suture by improving ease of passage through<br>tissues and by providing overall handling qualities as contrasted to the<br>braided, uncoated fiber. |
| | ETHIBOND EXCEL sutures are braided for optimal handling properties and,<br>for good visibility in the surgical field, are dyed green. ETHIBOND EXCEL<br>sutures meet all requirements established by the United States Pharmacopeia<br>(U.S.P.) for nonabsorbable sutures. |
| Intended use | For approximation of soft tissues. |
| Indications<br>statement | The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) is intended<br>for use in minimally invasive surgical applications where soft tissue is being<br>approximated with interrupted stitches. |
| | ETHIBOND EXCEL Polyester Suture is intended for use in general soft<br>tissue approximation and/or ligation, including use in cardiovascular,<br>ophthalmic and neurological procedures. |
| Technological<br>characteristics | The technological characteristics of the New Device is the same as the<br>Predicate Devices. |
| | Continued on next page |
| Performance<br>data | Bench testing and pre-clinical laboratory evaluations were performed to<br>ensure that the device could be used as designed. |
| | The studies demonstrated that the ETFS system facilitated laparoscopic<br>suturing, the ASD, together with the KEC, allowed for one-handed knot<br>deployment, knot security appeared adequate, and the Self-Right<br>Needleholder (Modified ENDOPATH® 5mm Reusable Needleholder) allowed<br>for rapid needle orientation and facilitated needle management. |
| Conclusion | Based on the 510(k) summaries and 510(k) statements (21 CFR §807) and<br>the information provided herein, we conclude that the New Device is<br>substantially equivalent to the Predicate Devices under the Federal Food,<br>Drug and Cosmetic Act. |
| Contact | Lorri Chavez, Project Manager<br>Ethicon Endo-Surgery, Inc.<br>4545 Creek Road<br>Cincinnati, Ohio 45242 |
| Date | July 15, 1997 |
{2}------------------------------------------------
### Appendix A: 510(k) Summary of Safety and Effectiveness, Continued
#### Device description (continued)
: 小 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
{3}------------------------------------------------
# Appendix A: 510(k) Summary of Safety and Effectiveness,
and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the commen
Continued
ﻂ ، ﻣﺴﺎ
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lorri Chavez Project Manager Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242
SEP 1 9 1997
K972679 Re:
> Trade Name: Endopath® Endoscopic Tissue Fastening System Regulatory Class: Class II Product Code: GAT Dated: July 15, 1997 Received: July 16, 1997
Dear Ms Chavez:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System
{5}------------------------------------------------
Page 2 - Ms. Lorri Chavez
Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used in the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:
- 1. The labeling, packaging and method of sterilization of the suture cannot be changed without prior notification, review and clearance by FDA.
- 2. The supplier of the sutures used in your device cannot be changed without prior notification, review and clearance by FDA.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers
{6}------------------------------------------------
Page 3 - Ms. Lorri Chavez
Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Appendix B: Indications for Use Statement
Indications for Use Statement: Statement
510(k) Number: K972679
Device Name: ENDOPATH® Endoscopic Tissue Fastening System (ETFS)
Indications for Use: The ENDOPATH® Endoscopic Tissue Fastening System (ETFS) is intended for use in minimally invasive surgical applications where soft tissue is being approximated with interrupted stitches.
ETHIBOND EXCEL Polyester Suture is intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
fcoQylo
(Dividion Sign-Off) Division of General Restorative 510(k) Number
**Prescription Use**
**(Per 21 CFR 801.109)**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.