← Product Code [MRH](/productcode/MRH) · K972664

# STORZ MILLENNIUM VISCOUS FLUID SYSTEM (K972664)

_Storz Instrument Co. · MRH · Oct 10, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K972664

## Device Facts

- **Applicant:** Storz Instrument Co.
- **Product Code:** [MRH](/productcode/MRH.md)
- **Decision Date:** Oct 10, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Storz Millennium™ Viscous Fluid System is a new device intended for the injection and aspiration of viscous fluids into and out of the eye during ophthalmic posterior vitreoretinal surgery. The Storz viscous fluid system consists of the Storz CX5700 module to be used with the Storz Millennium™ Microsurgical System and the Storz CX5710 sterile single use accessory pack.

## Device Story

System comprises Storz CX5700 module and CX5710 disposable injection pack; used during posterior vitreoretinal surgery. Module integrates with Storz Millennium™ Microsurgical System to inject/aspirate viscous fluids (e.g., silicone oil) for retinal tamponade. Operated by surgeon via foot pedal and touch screen interface. System provides dual linear foot control for pressure/vacuum regulation (0-70 PSI injection; 0-600 mmHg aspiration). Disposable pack includes tubing, syringe, and needle assembly. Output allows surgeon to manipulate intraocular fluid dynamics to fixate retinal tears/detachments.

## Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and design comparison to legally marketed predicate devices.

## Technological Characteristics

Modular pump system; AISI 303 stainless steel patient contact materials; dual linear foot controller; touch screen interface; sterile, single-use disposable accessory packs; packaged in Tyvek® pouches.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Storz Millennium™ Microsurgical System ([K961310](/device/K961310.md))
- Escalon® Viscous Fluid System ([K963434](/device/K963434.md))
- Richard James Viscous Fluid Transfer System ([K902835](/device/K902835.md))
- Escalon® Fluid Delivery Pack ([K963434](/device/K963434.md))

## Reference Devices

- Alcon Accurus™ Vitreoretinal Surgical System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K972664

Image /page/0/Picture/1 description: The image shows the word "storz" in white letters on a black background. The letters are stylized and bold. There is a registered trademark symbol in the upper right corner of the image. The background is a rounded rectangle.

Storz Millennium™ Viscous Fluid System Premarket Notification

0CT 1 0 1997

#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

The following information is submitted in accordance with the requirements of 21 CFR 807.92:

Contact Person: Patrick G. Balsmann, Regulatory Affairs Associate Storz, 3365 Tree Court Industrial Bldv., St. Louis, MO 63122-6694 Phone: (314) 225- 5051, ext. 5538.

Date Prepared: July 15, 1997

Proprietary Name: Storz Millennium™ Viscous Fluid System

Common/Usual Name: Viscous Fluid Injection/Aspiration System(s).

Classification Name: Infusion Pump, 86 MRH; 21CFR §880.5725.

Device Description/Intended Use: The Storz Millennium™ Viscous Fluid System is intended for the injection and aspiration of viscous fluids into and out of the eye during posterior ophthalmic vitreoretinal surgery. The system is designed to inject and aspirate viscous fluids, such as silicone oil which act as a tamponade in fixating large retinal tears and detachments. The Storz viscous fluid system consists of the Storz CX5700 Millennium™ Viscous Fluid System Module and the Storz CX5710 Viscous Fluid Injection Pack. The Storz CX5700 module is a self-contained pump system used as an individual module with the Storz Millennium™ Microsurgical System for the injection and aspiration of viscous fluids. The Storz CX5710 is the disposable accessory pack intended for single use to be used with the Storz viscous fluid module.

Predicate Device: The Storz CX5700 Millennium™ Viscous Fluid System Module is similar in design and function to the Storz Millennium™ Microsurgical System (K961310), Escalon® Viscous Fluid System (K963434), the Richard James Viscous Fluid Transfer System (K902835), and the Alcon Accurus™ Vitreoretinal Surgical System. The Storz CX5710 Viscous Fluid Injection Pack is similar in design, composition, and function to the Escalon® Fluid Delivery Pack (K963434.)

Predicate Comparison: A chart comparing characteristics of the Storz CX5700 Millennium™ Viscous Fluid System Module and the Storz CX5710 Viscous Fluid Injection Pack to those of the predicate devices is attached.

**Submitted by:**

Robert H. Johnson

Patrick G. Balsmann Regulatory Affairs

1 × 4

{1}------------------------------------------------

#### Storz CX5700 Millennium™ Viscous Fluid System Module Device Comparison Chart

(

(

| Device<br>Description                              | Storz CX5700<br>Viscous Fluid<br>System                                                      | Storz<br>Millennium™<br>Microsurgical<br>System                                                                                       | Alcon Accurus™<br>Vitreoretinal<br>Surgical System                                                                                                      | Escalon® Viscous<br>Fluid System                                                  | Richard James<br>Viscous Fluid<br>Transfer System                                 |
|----------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 510(k)                                             | current                                                                                      | K961310                                                                                                                               | unknown                                                                                                                                                 | K963434                                                                           | K902835                                                                           |
| Intended Use                                       | Posterior segment<br>ophthalmic<br>surgery - injection<br>& aspiration of<br>viscous fluids. | Anterior & posterior<br>segment ophthalmic<br>surgery.                                                                                | Posterior segment<br>ophthalmic surgery.                                                                                                                | Posterior ophthalmic<br>surgery - injection<br>& aspiration of<br>viscous fluids. | Posterior ophthalmic<br>surgery - injection &<br>aspiration of viscous<br>fluids. |
| Modular<br>Design                                  | Yes                                                                                          | Yes.                                                                                                                                  | No.                                                                                                                                                     | No.                                                                               | No.                                                                               |
| Ophthalmic<br>Features                             | Viscous fluid<br>injection/<br>aspiration.                                                   | Irrigation/aspiration,<br>phacoemulsification/<br>fragmentation,<br>vitrectomy, scissors,<br>bipolar, illumination,<br>& IOP control. | Viscous fluid<br>injection/aspiration,<br>Irrigation/aspiration,<br>fragmentation,<br>vitrectomy, scissors,<br>bipolar, illumination,<br>& IOP control. | Viscous fluid<br>injection/aspiration.                                            | Viscous fluid<br>injection/aspiration.                                            |
| Programmable<br>For Multiple<br>Surgeons           | Yes.                                                                                         | Yes.                                                                                                                                  | Yes.                                                                                                                                                    | No.                                                                               | No.                                                                               |
| Device<br>Description                              | Storz CX5700<br>Viscous Fluid<br>System                                                      | Storz<br>MillenniumTM<br>Microsurgical<br>System                                                                                      | Alcon AccurusTM<br>Vitreoretinal<br>Surgical System                                                                                                     | Escalon® Viscous<br>Fluid System                                                  | Richard James<br>Viscous Fluid<br>Transfer System                                 |
| 510(k)                                             | current                                                                                      | K961310                                                                                                                               | unknown                                                                                                                                                 | K963434                                                                           | K902835                                                                           |
| Disposable<br>Accessories                          | Yes.                                                                                         | Yes.                                                                                                                                  | Yes.                                                                                                                                                    | Yes.                                                                              | Yes.                                                                              |
| Viscous Fluid<br>User Interface                    | Foot pedal &<br>touch screen.                                                                | Foot pedal & touch<br>screen.                                                                                                         | Foot pedal & touch<br>screen.                                                                                                                           | Foot pedal &<br>control knob.                                                     | Foot pedal & control<br>knob.                                                     |
| Foot<br>Controller -<br>Dual Linear                | Yes                                                                                          | Yes.                                                                                                                                  | No.                                                                                                                                                     | No.                                                                               | No.                                                                               |
| Viscous Fluid<br>Injection,<br>Pressure<br>Range   | 0 - 70 PSI                                                                                   | Not Applicable.                                                                                                                       | 0 - 80 PSI.                                                                                                                                             | 0 - 70 PSI.                                                                       | 0 - 70 PSI.                                                                       |
| Viscous Fluid<br>Injection,<br>Aspiration<br>Range | 0 - 600 mmHg.                                                                                | Not Applicable.                                                                                                                       | 0 - 600 mmHg.                                                                                                                                           | 0 - 600 mmHg.                                                                     | 0 - 650 mmHg.                                                                     |

ﻟ

{2}------------------------------------------------

# Storz CX5700 Millennium™ Viscous Fluid System Module
Comparison

(

ﺴﺮ ﻣ

{3}------------------------------------------------

#### Storz CX5710 Viscous Fluid Injection Pack Device Comparison Chart

ﻟﻤﺴﺘﻌﻤﺎ

-

| Device Description                  | Storz CX5710 Viscous Fluid Injection Pack<br>Escalon® Fluid Delivery Pack                                                                                                                                                                                                                                                                                     |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                              | current<br>K963434                                                                                                                                                                                                                                                                                                                                            |
| Pack Components                     | Modified hose barb, tubing, male luer with locking nut, filter, 10cc syringe adaptor, 10cc modified piston, needle/cover assembly, 10cc standard piston, & 10cc syringe assembly.<br>Hose barb, tubing, male luer with locking nut, filter, 10cc syringe adaptor, 10cc modified piston, needle/cover assembly, 10cc standard piston, & 10cc syringe assembly. |
| Applicable Ophthalmic Surgical Unit | Storz Millennium™ Microsurgical System<br>Escalon® Viscous Fluid System                                                                                                                                                                                                                                                                                       |
| Provided Sterile                    | Yes.<br>Yes.                                                                                                                                                                                                                                                                                                                                                  |
| Labeled For Single Patient Use      | Yes.<br>Yes.                                                                                                                                                                                                                                                                                                                                                  |
| Patient Contact Material            | AISI 303 stainless steel.<br>AISI 303 stainless steel.                                                                                                                                                                                                                                                                                                        |
| Packaging                           | Sealed Tyvek® pouches.<br>10 packs/box.<br>Sealed Tyvek® pouches.<br>10 packs/box.                                                                                                                                                                                                                                                                            |

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular pattern.

OCT 10 1997

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Patrick G. Balsmann Domestic Regulatory Affairs Associate Storz Instrument Company 3365 Tree Ct. Industrial Blvd. St. Louis, MO 63122-6694

Re: K972664

> Trade Name: Storz Millennium™ Viscous Fluid System Regulatory Class: II Product Code: 86 MRH Dated: July 15, 1997 Received: July 16, 1997

Dear Mr. Balsmann:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. Patrick G. Balsmann

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A. Roepke Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo for Storz. The logo is in black and white and features the word "storz" in a stylized font. The word is set against a black background with rounded corners. There is a registered trademark symbol to the right of the word.

## Storz Millennium™ Viscous Fluid System Premarket Notification

### INDICATIONS FOR USE

Device Name:

Storz Millennium™ Viscous Fluid System

Indications for Use:

The Storz Millennium™ Viscous Fluid System is a new device intended for the injection and aspiration of viscous fluids into and out of the eye during ophthalmic posterior vitreoretinal surgery. The Storz viscous fluid system consists of the Storz CX5700 module to be used with the Storz Millennium™ Microsurgical System and the Storz CX5710 sterile single use accessory pack.

Enette TK Beers

ivision of Ophthalmic 510(k) Number

**Prescription Use**
**(Per 21 CFR 801.109)**

(114

---

**Source:** [https://fda.innolitics.com/device/K972664](https://fda.innolitics.com/device/K972664)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
