← Product Code [GEI](/productcode/GEI) · K972558

# BIPOLAR SCISSORS (K972558)

_Enable Medical Corp. · GEI · Oct 7, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K972558

## Device Facts

- **Applicant:** Enable Medical Corp.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 7, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ENABLE Bipolar Scissors with ENABLE Bipolar Disposable Cartridges are intended to cut tissue and control bleeding through coagulation during general surgery.

## Device Story

Reusable electrosurgical instrument consisting of handle and stainless steel scissors blades; utilizes disposable bipolar cartridges. Surgeon operates device by placing blades across target tissue and closing handle; electrosurgical generator (20-30 watts) activated via footswitch to deliver current through electrodes to blades. Simultaneous cutting and coagulation occurs; surgeon modulates current flow based on tissue type and bleeding. Used in general surgery; intended for physician use. Benefits include efficient tissue management and hemostasis.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Reusable handle with stainless steel blades; disposable bipolar cartridges. Bipolar electrosurgical energy source; power range 20-30 watts. Device length 4-11 inches; blade length 1-2 inches. Manual operation via handle and footswitch.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Symbiosis Bipolar Scissors ([K951387](/device/K951387.md))
- Everest Medical Bipolar Scissors ([K945975](/device/K945975.md) and [K955001](/device/K955001.md))
- CardioThoracic Systems MIDCAB/SVH Bipolar Scissors ([K963930](/device/K963930.md))
- Ethicon PowerStar Bipolar Scissors ([K960476](/device/K960476.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

0CT - 7 1997

# 510(k) SUMMARY BIPOLAR SCISSORS WITH ENABLE BIPOLAR DISPOSABLE CARTRIDGES 510(k) NOTIFICATION K_9725558

### GENERAL INFORMATION

| Manufacturer:        | ENABLE Medical Corporation<br>6345 Centre Park Drive<br>West Chester, OH 45069-3863<br>(513) 755-7600<br>(513) 755-7676<br>Est. Reg. No. 1530251 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Michael D. Hooven<br>President/CEO<br>ENABLE Medical Corporation                                                                                 |
| Date Prepared:       | --------                                                                                                                                         |
| DEVICE DESCRIPTION   |                                                                                                                                                  |
| Classification:      | Class II                                                                                                                                         |
| Trade Name:          | ENABLE Bipolar Scissors with ENABLE Bipolar<br>Disposable Cartridges                                                                             |
| Generic/Common Name: | Electrosurgical cutting and coagulation device and<br>accessories<br>21CFR878.4400                                                               |

## PREDICATED DEVICES

- 1. Symbiosis Bipolar Scissors (K951387)
- 2. Everest Medical Bipolar Scissors (K945975 and K955001)
- 3. CardioThoracic Systems MIDCAB/SVH Bipolar Scissors (K963930)
- 4. Ethicon PowerStar Bipolar Scissors (K960476)

{1}------------------------------------------------

### INTENDED USE

The ENABLE Bipolar Scissors with ENABLE Bipolar Disposable Cartridges are intended to cut tissue and control bleeding through coagulation during general surgery.

## PRODUCT DESCRIPTION

The ENABLE Bipolar Scissors is a reusable electrosurgical instrument consisting of an instrument handle with a pair of scissors blades which attaches to the ENABLE Bipolar Disposable Cartridges and an electrosurgical generator. Electric current flows from the electrodes in the cartridges to the stainless steel cutting surfaces of the blades. The electrosurgical generator controls the flow of the electric current down the electrode. The overall length of the device is between 4 and 11 inches. The power cord connecting the handle to the power control unit is approximately 120 inches. The blades are between 1 and 2 inches in length.

The surgeon places the opened scissors across the tissue to be cut and closes the scissors handle while moving the scissors across the tissue. Energizing the electrosurgical generator via the generator's footswitch, the surgeon can simultaneously cut and coagulate the target tissue. The surgeon can vary the current flow rate according to the tissue and the amount of bleeding encountered. The power setting for the ENABLE Bipolar Scissors is 20-30 watts.

The ENABLE Bipolar Scissors with ENABLE Bipolar Disposable Cartridges are substantially equivalent to the above-identified predicate devices with regard to intended use, function, physical characteristics, materials and sterilization method. ENABLE, CardioThoracic, Symbiosis, Everest and Ethicon devices are all bipolar scissors that cut tissue and coagulate soft tissue through the use of bipolar technology. All the bipolar scissors are connected to the same or similar electrosurgical generators and use similar power ranges for operation.

### SUMMARY

As contained in this 510(k) summary, the ENABLE Bipolar Scissors with ENABLE Bipolar Disposable Cartridges are substantially equivalent to the predicate devices identified.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

0CT - 7 1997

Mr. Michael D. Hooven President ENABLE Medical Corporation 6345 Centre Park Drive West Chester, Ohio 45069-3863

Re: K972558 ENABLE Bipolar Scissors with ENABLE Disposable Cartridges Requlatory Class: II Product Code: GEI Dated: July 7, 1997 Received: July 9, 1997

Dear Mr. Hooven:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{3}------------------------------------------------

Page 2 - Mr. Michael D. Hooven

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Colia M. Witton, Ph.D.

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{4}------------------------------------------------

(

(

K979558

# Bipolar Scissors 510(k) Premarket Notification

# STATEMENT OF INDICATIONS OF USE

The ENABLE Bipolar Scissors with ENABLE Bipolar Disposable Cartridges is intended to cut tissue and control bleeding through coagulation during general surgery.

Prescription Use *X*
(Per 21 CFR 801.109)

Acolla

(Divis

---

**Source:** [https://fda.innolitics.com/device/K972558](https://fda.innolitics.com/device/K972558)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
