NPB-195 PULSE OXIMETER

K972468 · Nellcor Puritan Bennett, Inc. · DQA · Sep 25, 1997 · Cardiovascular

Device Facts

Record IDK972468
Device NameNPB-195 PULSE OXIMETER
ApplicantNellcor Puritan Bennett, Inc.
Product CodeDQA · Cardiovascular
Decision DateSep 25, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Intended Use

The intended use of the NPB-195 Pulse Oximeter is the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. For use with neonatal, pediatric and adult patients, in hospital, hospital-type, intra-hospital transport and home environments. For prescription use only.

Device Story

NPB-195 Pulse Oximeter provides continuous, non-invasive monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate. Device utilizes compatible Nellcor Puritan Bennett sensors containing red and infrared LEDs; photodiode detects light absorption variations during pulsatile cycle. Internal engine processes analog sensor signals to calculate SpO2, pulse rate, and pulse amplitude. Output displayed on LCD as digital values, 'blip bar' pulse amplitude, and pleth waveform. Device includes audible/visual alarms for saturation, pulse rate, and sensor search. Used in hospitals, transport, and home environments by clinicians or patients. Serial port (EIA-232) supports real-time ASCII data output for printers or nurse call systems. Powered by AC or internal rechargeable lead-acid battery. Benefits include real-time physiological monitoring to assist clinical decision-making regarding patient oxygenation status.

Clinical Evidence

Noninvasive controlled hypoxia studies conducted to establish SpO2 accuracy and verify sensor performance specifications. No other clinical data provided.

Technological Characteristics

Measures light absorption via red/infrared LEDs and photodiode. Powered by AC or internal 6V rechargeable lead-acid battery. Features LCD display, serial (EIA-232) interface, and Flash ROM upgradability. Software algorithm and engine circuitry identical to predicate NPB-40.

Indications for Use

Indicated for continuous, non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate in neonatal, pediatric, and adult patients in hospital, transport, or home settings. Prescription use only.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 11-1 171 11 #### Section 16. 510(k) Summary Section 16.a Date Summary Prepared 30 June 1997 SEP 2 5 199 Section 16.b Company Information Establishment: Official Correspondent: Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588 Roger D. Brown Regulatory Affairs Specialist Nellcor Puritan Bennett Inc. 11150 Thompson Avenue Lenexa, KS 66219 (913) 495-7146 (direct phone) (913) 495-7285 (fax) ## Section 16.c Name of Device Proprietary: Common/Usual: NPB-195 Pulse Oximeter Pulse Oximeter Classification: Altrades plant Oximeter (§870.2700/74DQA) # Section 16.d Equivalent Devices Substantial equivalence to the following legally marketed predicate devices with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for predicate devices and the NPB-195, as well as testing to accepted industry standards. In addition, noninvasive controlled hypoxia studies were conducted to establish the NPB-195's accuracy and to ensure that the sensors meet their currently published accuracy specifications with the NPB-195. The predicate devices are as follows: 1. NPB-40 Handheld Pulse Oximeter, Nellcor Puritan Bennett Inc., K963707 2. Ohmeda 3800 Pulse Oximeter, Ohmeda Inc., K962127 847 {1}------------------------------------------------ ## Section 16.e Device Description The NPB-195 Pulse Oximeter is designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate by use of one of a range of compatible Nellcor Puritan Bennett oxygen transducers (sensors) The NPB-195 displays digital values of SpO2 and pulse rate, and pulse amplitude by means of a "blip bar" presentation. The NPB-195 is also capable of displaying a Pleth Waveform and trends on its LCD display. The NPB-195 can be powered by an internal power supply operating on AC from a standard electrical utility receptacle (manually switchable from 115V to 230V) or alternatively by an integral sealed 6V rechargeable lead-acid battery. The NPB-195 is intended for prescription use with adult, pediatric and neonatal patients in hospital, hospital-type, intra-hospital transport and home environments. Audible and visual alarms for high/low saturation, pulse rate and pulse search are provided. The NPB-195 also includes adjustable alarm silence duration and other configurable power on settings. The NPB-195 provides an audible low battery warning to alert the user of impending loss of power and consequent loss of monitoring capability. The NPB-195 Pulse Oximeter has visual indicators for pulse search, power mode (i.e. battery or AC) and alarm silence in addition to alarm features. In addition to the above mentioned device features, the instrument has been designed to satisfy the needs of both the user and the patient. A convenient carrying handle is incorporated into the case. There is also a serial port (EIA-232 interface) that provides ASCII output of real-time data every two seconds. This data can be printed on serial printers and can interface with nurse call systems through the rear connector. The device is also Flash ROM upgradable. #### Section 16.f Intended Use The intended use of the NPB-195 Pulse Oximeter is for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and neonate patients. The environments of use are hospital, hospitaltype facilities, intra-hospital transport and the home. The intended use, patient population and environment of use are the same or similar to the predicate devices, the NPB-40 and Ohmeda 3800, #### Section 16.g Technological Characteristics The NPB-195 Pulse Oximeter measures functional oxygen saturation by calculating the light absorption of tissue, bone, and blood in the sampling light beam path during the pulsatile cycle. Red and infrared LED's are utilized as light sources. A photodiode acting as a photodetector senses the signal strength of the two wavelengths of light, which vary inversely with the amount of light transmitted through the tissue. The NPB-195 receives this electrical information from the sensor and processes the information by use of an SpO2 software algorithm to provide a real time value of SpO2. The software algorithm is contained in the NPB-195 "engine". The "engine" (part of the instrument that processes the analog signals from the sensors) consists of electrical circuitry and processors that calculate the SpO2, pulse rate, and pulse amplitude. The electrical circuitry of the NPB-195 engine is the same as the engine used by the predicate device, NPB-40, cleared under K963707. The NPB-195 also uses the same SpO2 software algorithm to process the information from the sensor as the predicate device, NPB-40, cleared under K963707. 0 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 5 1997 Mr. Roger D. Brown Nellcor Puritan Bennett Inc. 11150 Thompson Avenue Lenexa, Kansas 66219-2301 K972468 Re: NPB-195 Pulse Oximeter Regulatory Class: II (two) 74 DQA Product Code: Dated: June 30, 1997 Received: July 1, 1997 Dear Mr. Brown: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Mr. Roger D. Brown This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | Page | 1 of 1 | |------|--------| |------|--------| | 510(k) Number (if known): | K972468 | |---------------------------|------------------------| | Device Name: | NPB-195 Pulse Oximeter | Indications For Use: The intended use of the NPB-195 Pulse Oximeter is the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate. For use with neonatal, pediatric and adult patients, in hospital, hospital-type, intra-hospital transport and home environments. For prescription use only. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) | |------------------------------------------|----|--------------------------------------------------| |------------------------------------------|----|--------------------------------------------------| | (Division Sign-Off)<br>Division of Cardiovascular, Respiratory,<br>and Neurological Devices<br>510(k) Number | K972468 | |--------------------------------------------------------------------------------------------------------------|---------| |--------------------------------------------------------------------------------------------------------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...