ULTIMA AND ULTIMA LX CEMENTED FEMORAL STEM
K972435 · Johnson & Johnson Professionals, Inc. · JDI · Sep 11, 1997 · Orthopedic
Device Facts
| Record ID | K972435 |
| Device Name | ULTIMA AND ULTIMA LX CEMENTED FEMORAL STEM |
| Applicant | Johnson & Johnson Professionals, Inc. |
| Product Code | JDI · Orthopedic |
| Decision Date | Sep 11, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The ULTIMA and ULTIMA LX Femoral Hip Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, traumatic and non-union of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques. The ULTIMA and ULTIMA LX Cemented Femoral stem is indicated for use only with PMMA bone cement.
Device Story
Femoral component for total or partial hip arthroplasty; implanted with PMMA bone cement. Used in conjunction with acetabular component and modular hip head. Features include collared or collarless (ULTIMA LX) designs, polished or grit blast finishes, and proximal slot for instrumentation. Available in 135-degree (centralized) or 130-degree (medialized) neck angles to facilitate anatomical head positioning. Used by orthopedic surgeons in clinical settings. Provides structural support for hip joint; restores mobility; reduces pain in patients with severe joint damage.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and functional comparison to predicate device.
Technological Characteristics
Cemented femoral hip prosthesis. Materials: metal/polymer. Design: collared/collarless, polished/grit blast finish, proximal slot. Neck angles: 130/135 degrees. Modular compatibility with P.F.C. Hip Heads. Implanted with PMMA bone cement.
Indications for Use
Indicated for patients requiring total or partial hip replacement due to severe pain/disability from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, femoral fractures, congenital hip dysplasia, protrusio acetabuli, slipped capital epiphysis, or previous fusion with inadequate bone stock. Must be used with PMMA bone cement.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
- ULTIMA Total Hip System Cemented Femoral Stem (K924379)
Reference Devices
- P.F.C. Modular Hip Heads (K893872)
Related Devices
- K963109 — HERITAGE HIP SYSTEM · Zimmer, Inc. · Jan 9, 1997
- K191056 — Conformity stem, cemented · United Orthopedic Corporation · Jan 3, 2020
- K964357 — NATURAL-HIP SYSTEM COCR STEM · Intermedics Orthopedics · Jan 27, 1997
- K961799 — NATURAL HIP SYSTEM - DRG HIP STEM · Intermedics Orthopedics · Jul 15, 1996
- K954598 — OSTEONICS OFFSET HIP STEMS · Osteonics Corp. · Apr 4, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
# SEP | | 1997
## Summary of Safety and Effectiveness Data for the ULTIMA® and ULTIMA® LX Cemented Femoral Stem
## Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350
K972435
### Contact Person________________________________________________________________________________________________________________________________________________________________
Mary E. Gray Associate Regulatory Affairs Specialist Phone: (508) 828-3545 (508) 828-3212 Fax:
#### Name of Device
| Proprietary Name: | Ultima ® and Ultima® LX Cemented Femoral Stem |
|----------------------|-----------------------------------------------------------------|
| Common Name: | Hip Prosthesis |
| Classification Name: | Hip/joint metal/polymer semi-constrained cemented<br>prosthesis |
| Regulatory Class: | Class II by 21 CFR 888.3350 |
| Product Code: | 87 JDI |
| Owner/Operator No.: | 9001269 |
### Device Classification__________________
This device has been placed in Class II for Hip joint metal/polymer semiconstrained cemented prosthesis per 21 CFR § 888.3350.
### Statement of Substantial Equivalence__________________________________________________________________________________________________________________________________________
The modifications to the ULTIMA Cemented Femoral Stem which include surface finishes of polished and grit blast, collared and collarless (ULTIMA LX) and an increase in offset, and a larger proximal slot are substantially equivalent and identical in material and in function to the ULTIMA Total Hip System Cemented Femoral Stem cleared for marketing under premarket notification #K924379 (February 12, 1993).
The subject device is composed of similar materials to the predicate device mentioned above. Further, the intended use, design, and manufacture of the modified ULTIMA and ULTIMA LX Cemented Femoral Stem are substantially equivalent to those currently distributed. Additionally, the packaging and method of sterilization utilized for the ULTIMA and ULTIMA LX Cemented Femoral Stem are the same as those used for the previously mentioned predicate device.
ULTIMA and ULTIMA LX Cemented Femoral Stem
{1}------------------------------------------------
The ULTIMA and ULTIMA LX Cemented Femoral Hip Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, traumatic and non-union of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques.
The ULTIMA and ULTIMA LX Cemented Femoral Stem is indicated for use only with PMMA bone cement.
### Physical Description
The ULTIMA and ULTIMA LX Cemented stem is a femoral component for use in total or partial hip arthroplasty procedures. It is designed to be implanted with Polymethylmethacrylate bone cement (PMMA) and used in conjunction with a total hip acetabular component and modular hip head. The ULTIMA and ULTIMA LX Femoral Hip Stems can be used in either the right or left hip.
The stem body design is derived from that of the primary ULTIMA stems K#924379. These stems will be offered in both a collared and collarless (ULTIMA LX) design, with either a polished or grit blast finish and a larger proximal slot. The stems have also been designed with a proximal slot to allow the surgeon an additional instrumentation choice during the seating of the implant.
Both the ULTIMA and ULTIMA LX femoral components are available with either a neck angle of a 135 (centralized) or a neck angle of 130 (medialized) to allow for more accurate anatomical head positioning for the patient. The ULTIMA and ULTIMA LX femoral components can be utilized with the P.F.C. Modular Hip Heads (previously cleared K# 893872) of various neck heads diameters (22.225, 26, 28, and 32 mm) and three neck lengths of standard, +5 mm and +10 mm. The 28 mm head and the 32 mm head are also available with a neck length of -3 mm and -5 mm, respectively.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with a flowing, wave-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 11 1997
Ms. Mary E. Gray ·Associate Regulatory Affairs Specialist Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350
K972435 Re: Ultima® and Ultima® LX Cemented Femoral Stem Regulatory Class: II Product Code: JDI June 27, 1997 Dated: Received: June 30, 1997
Dear Ms. Gray:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - Ms. Mary E. Gary
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
coleps
f Cella M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use for the ULTIMA® and ULTIMA® LX Cemented Femoral Stems
Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350
## Indications of Use
The ULTIMA and ULTIMA LX Femoral Hip Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, traumatic and non-union of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques.
The ULTIMA and ULTIMA LX Cemented Femoral stem is indicated for use only with PMMA bone cement.
**Prescription Use**
(Per 21 CFR 801.109)
Division Sign-Off
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K972435