HP ENDOVAGINAL/ENDORECTAL PROBE, HP IMAGE POINT ULTRASOUND SYSTEM
Applicant
Hewlett-Packard Co.
Product Code
ITX · Radiology
Decision Date
Jul 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
Diagnostic ultrasound imaging or Doppler analysis of the human body as follows: Mode of Operation (Clinical Application: Fetal, Small Organ (prostate), Trans-rectal, Trans-vaginal; Modes: B, M, PWD, Color Doppler, Power (Amplitude) Doppler Imaging, Combined (B+M, B+PW)). Prescription Use (Per 21 CFR 801.109).
Device Story
The 21336A is an endovaginal/endorectal ultrasound probe designed for use with the HP M2410A ultrasound system. It functions as a transducer, converting electrical signals into acoustic energy to image internal structures and back into electrical signals for processing by the host system. It is operated by clinicians in a clinical setting to perform diagnostic imaging and Doppler analysis. The device provides real-time visual feedback to the physician, aiding in clinical assessment and decision-making for gynecological, obstetrical, and urological (prostate) conditions. The primary modification from predicate devices is the use of new patient contact materials, which have been verified for biocompatibility. The device is intended for prescription use only.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, specifically biocompatibility testing of new patient contact materials and acoustic output measurements confirming compliance with FDA limits.
Technological Characteristics
Endovaginal/endorectal ultrasound transducer. Operates with HP M2410A system. Features new patient contact materials verified for biocompatibility. Acoustic output is within FDA limits. Connectivity is via integration with the M2410A ultrasound system.
Indications for Use
Indicated for diagnostic ultrasound imaging and Doppler analysis in fetal, small organ (prostate), trans-rectal, and trans-vaginal applications. Contraindicated for in vitro fertilization and percutaneous umbilical blood sampling.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{1}------------------------------------------------
JUL 2 4 1997
K972348
, I
# 510(k) Summary of Safety and Effectiveness
#### per 21 CFR 807.92
- Submitter's Name / Contact Person: Rob Butler 1) Address: 3000 Minuteman Road, Andover, MA 01810 Telephone Number: (508) 659-2785 Date Summary was prepared: June 23, 1997
- Trade Name: 21336A probe for M2410A ultrasound imaging system 2) Common Name: endovaginal/endorectal probe for "ImagePoint" ultrasound system Classification Pro Codes: 90-ITX, 90-IYO, and 90-IYN
- Identification of Predicate Devices: 21370A endovaginal probe and 21371A endorectal 3) probes used on HP-Philips SD800 ultrasound system (K935923) and HP M2410A ultrasound system (K954028)
- Description of the device or modification being submitted for premarket approval. 4)
Functionality: 21336A EV/ER probe has equivalent functionality to the 21370A EV and 21371A ER probes now being used with M2410A; acoustic output for 21336A is within FDA limits
Scientific Concepts: same as existing probes on M2410A
Significant Characteristics of the Modification: new patient contact materials for 21336A EV/ER probe
- Statement of Intended Use: same as intended use for current devices (M2410A, 21370A, 5) 21371A): endovaginal, gynecological, obstetrical, endorectal, urological (small parts prostate)
- Predicate Device Comparison: 21336A on M2410A is very similar to predicate devices in 6) all respects except patient contact materials; this 510(k) submission includes biocompatibility test results which demonstrate biocompatibility for the new materials
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol, with three wavy lines representing the flow of health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Rob Butler Regulatory Approval Hewlett-Packard Medical Products Group Hewlett-Packard Co. 3000 Minuteman Road Andover, MA 01810-1087
JUJI 2 4 1997
Re: K972348
HP Model 21336A Endovaginal/Endorectal Probe Dated: June 23, 1997 Received: June 24, 1997 Regulatory class: II 21 CFR 892.1570/Procode: 90 ITX
Dear Mr. Butler:
We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Image Point (HP Model 2410A) Diagnostic Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
#### 21336A
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for medical Devices: General regulation (CFR) Part 820) and that, through periodic QS inspections, the Food and
{3}------------------------------------------------
### Page 2 - Rob Butler
Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's February 17, 1993 "Revised 510(k) Diagnostic Ultrasound Guidance for 1993." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and . prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
{4}------------------------------------------------
Page 3 - Rob Butler
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "dsmo@fdadr.cdrh.fda.gov".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David A. Asbjorn
Lillian Yin, Ph.D.
for
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
## FDA Ultrasound Device Indications Statement
Page 1
510(k) Number (if known) :
Device Name: 21336A Endovaginal/Endorectal Probe for M2410A Ultrasound System
Fill out one form for each ultrasound system or transducer = = . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Intended Use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows: Mode of Operation
:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler<br>Imaging | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|---------------------------------|----|----|----|-----|-----|------------------|--------------------------------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Fetal | NA | ✓ | ✓ | ✓ | NA | ✓ | ✓ | NA | ✓ | NA |
| Abdominal | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Intra-operative (specify) | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Intra-Operative<br>Neurological | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Pediatric | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Small Organ (prostate) | NA | ✓ | ✓ | ✓ | NA | ✓ | ✓ | NA | ✓ | NA |
| Neonatal Cephalic | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Adult Cephalic | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Cardiac Adult | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Cardiac Pediatric | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Trans-esophageal | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Trans-rectal | NA | ✓ | ✓ | ✓ | NA | ✓ | ✓ | NA | ✓ | NA |
| Trans-vaginal | NA | ✓ | ✓ | ✓ | NA | ✓ | ✓ | NA | NA | NA |
| Intra-Luminal | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Trans-urethral | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Peripheral Vessel | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Laparoscopic | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Musculo-Skeletal Conv. | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Musculo-Skeletal Superf. | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Other (specify) | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
Other Indications or Modes: combined modes are: B+M, B+PW
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use . (Per 21 CFR 801.109) 14
Vamil a. Segonm
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K97234
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.