The intended use of the device is to inject medicaments subcutaneously, intramuscularly, or via I.V. access ports. This device is intended for all adult patient populations based upon the appropriate needle size selection by the physician or clinician, and is governed by the specific drug being administered. Any pediatric applications is at the discretion of the physician and governed by the specific drug being administered.
Device Story
Gettig Styrl-Ject® Delivery System is a sterile, single-use hypodermic syringe cartridge; consists of pre-filled glass carpule and plastic syringe. Carpule features plastic retainer with frangible button to maintain sterility of needle path until use; eliminates need for alcohol swab. User snaps off button, inserts carpule into plastic syringe; syringe needle rear extension punctures rubber diaphragm to create fluid path. Plunger rod engages plunger for aspiration and injection. Configurable for various needle sizes; supports subcutaneous, intramuscular, and I.V. applications via butyl and needleless access ports. Used by physicians or clinicians in clinical settings to administer prescribed medicaments.
Clinical Evidence
Bench testing only. Performance evaluations included carpule leak testing (vacuum dye), button snap-off force, glass neck integrity, plunger extrusion force, needle sharpness/penetration drag, needle retention, and cover pull-off force. No clinical data provided.
Technological Characteristics
Sterile, single-use hypodermic syringe cartridge. Components: glass carpule, plastic syringe. Materials utilize USP class VI designations with established history in medical manufacturing. Connectivity: none. Energy source: manual. Sterilization: not specified (implied sterile).
Indications for Use
Indicated for subcutaneous, intramuscular, or I.V. administration of medicaments in adult patients; pediatric use at physician discretion. Contraindications governed by specific drug administered.
Regulatory Classification
Identification
A cartridge syringe is a device intended to inject anesthetic agents subcutaneously or intramuscularly. The device consists of a metal syringe body into which a disposable, previously filled, glass carpule (a cylindrical cartridge) containing anesthetic is placed. After attaching a needle to the syringe body and activating the carpule by partially inserting the plunger on the syringe, the device is used to administer an injection to the patient.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
SEP 11 1997
K972285
# Section A
ﺎ ﻣﺴ
# 510 (k) Summary
## Submitted By:
Gettig Pharmaceutical Instrument Company A Division of Gettig Technologies, Inc. One Streamside Place West P.O. Box 85 Spring Mills, PA 16875
# Establishment Registration Number - #2511670
Telephone Number - 814-422-8892 Fax Number - 814-422-8011 email - kfee@gettig.com
Contact Person: Kevin Fee
Date of Application - June 17, 1997
#### Device Name:
| • Proprietary Name -- | Gettig Styrl-Ject® Delivery System |
|-----------------------------------|--------------------------------------------|
| • Common Name -- | Hypodermic Syringe Cartridge |
| • Proposed Classification Name -- | Syringe, Cartridge<br>21 CFR Sec. 872.6770 |
| • Class II | |
| • Product Code -- | 76 E J I |
#### Predicate Device - Substantial Equivalence
The Gettig Stryl-Ject® Delivery System is similar to various predicate and preamendment devices currently and previously marketed having the same intended use for medicament administration. The device has attributes similar to the Astra Aspirating Syringe (K851903), the Terumo Hypodermic Syringe (K771205), the Wyeth Tubex® (Pre-amendment), and the Monoject® Prefilled Syringe (K812934).
{1}------------------------------------------------
# Device Description
The Gettig Styrl-Ject® Delivery System is a sterile, single use hypodermic syringe cartridge device designed to be similar to current single use products. The device consists of two basic constituents - a previously filled glass carpule, and plastic syringe. The carpule features a plastic retainer with a frangible button. When ready to use, the carpule retainer button is snapped off, exposing the rubber diaphragm insert. Until, the button is snapped the carpule needle path remains sterile, thus eliminating the need for an alcohol swab. The plastic syringe is also sterile until ready for use via the plunger rod encapsulating the needle rear extension. After removal of the threaded plunger rod, the carpule is then inserted into the clear plastic syringe allowing the syringe needle rear extension to puncture the rubber diaphragm creating a path for medicament administration. Then the syringe plunger rod is engaged with the plunger and the cover (needle or luer) is removed, to allow aspiration and injection. The syringe can be configured for a variety of needle sizes for subcutaneous injection as well as for I.V. applications via butyl and needleless access ports.
## Biocompatibility
Reasonable assurance of biocompatibility of the materials comprising the device is provided by their established history in medical product manufacturing, USP class VI designations, and DMF file information for the materials (see Section H - Appendix).
#### Materials
See Section F for materials description and Section H Appendix for material specifications.
## Physical Properties
See Section H Appendix for physical properties
## Statement of Intended Use
The intended use of the device is to inject medicaments subcutaneously, intramuscularly, or via I.V. access ports. This device is intended for all adult patient populations based upon the appropriate needle size selection by the physician or clinician, and is governed by the specific drug being administered. Any pediatric applications is at the discretion of the physician and governed by the specific drug being administered.
## Indication Statement Differences
There are no differences for indications with the predicate device(s).
## Comparison of Technologies, Materials, and Design with Predicate Device(s).
See Section E, Tables 11 & 12 for comparison data.
{2}------------------------------------------------
## Performance Bench Tests and Data to Show Safety and Effectiveness
The non-clinical bench tests performed on the Gettig Styrl-Ject® Delivery System are the same protocols and standards used for the evaluation of our standard syringe products. See Section C for performance data and Section H Appendix for performance test summary.
#### Listing of Performance Bench Tests
Carpule Leak Test Vacuum Dye Air Gauge - Federal DimensionAire Button Snap Tests Before Sterilization After Sterilization Glass Neck Snap-Off Schott revision 1 & 2 Plunger Extrusion Force
Using O-Ring Universal Plunger Chattilon Gauge
Needle Sharpness - 22 ga & 25 ga Penetration Drag
Needle Rear Extension Penetration
Needle Retention - Instron Pull
22 ga & 25 ga
Cover Pull-off Force - Instron Pull 22 ga & 25 ga
22 ga & 23 ga
N. H. P. Fastener Corp.
Needle Rear Extension Coring 22 ga & 25 ga Rear Beveled
22 ga Blunt w/Rear Blunt
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP | | 1997
Mr. Kevin Fee Gettig Pharmaceutical Instrument Company 1 Stream Side Place West Spring Mills, Pennsylvania 16875-0085
K972285 Re: Trade Name: Gettig Styrl-Ject Delivery System Requlatory Class: II Product Code: EJI Dated: June 17, 1997 Received: June 19, 1997
Dear Mr. Fee:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug
{4}------------------------------------------------
Page 2 - Mr. Fee
Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.f.a.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timot Ulatowski Direct or Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
{5}------------------------------------------------
Page of
510(k) Number (if known):
Device Name: Gettig Styrl-Ject® Delivery System
Indications For Use:
The indications for use of the device is to inject medicaments subcutaneously, intramuscularly, or via I.V. access ports. This device is subedules for all adult patient populations based upon the appropriate needle size selection by the physician or clinician, and is governed by the specific drug being administered. Any pediatric applications is at the discretion of the physician and governed by the specific drug being administered.
There are no differences for indications with the predicate device(s).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Dental, Infection Contr and General Hospital D 5 i O(k) Number
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use.
(Optional Format 1-2-96)
.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.