Intended Use
The Stryker Wedge Interference Screw System is intended for use in the surgical reconstruction of the anterior cruciate ligament (ACL) deficient knee to provide interference fixation of the femoral and tibial bony attachments of the patella bone-patellar tendon-tibial bone graft complex, the semi-membranosus tendon grafts, and the semi-tendinosus tendon grafts, or fixation of the allograft techniques such as ACL allografts and Achilles tendon allografts.
Device Story
Stryker Wedge Interference Screw System is a metallic bone fixation fastener used in orthopedic surgery. Device provides interference fixation for soft tissue and bone-tendon grafts during ACL reconstruction. Used in clinical settings by orthopedic surgeons. Device is provided sterile for single-use. Material is biocompatible titanium alloy. Surgeon inserts screw to secure graft within femoral or tibial bone tunnels. Fixation restores mechanical stability to the knee joint, facilitating ligamentous healing and patient recovery. Device is not for spinal pedicular fixation.
Clinical Evidence
Bench testing only. Biocompatibility verified per ASTM F136, ISO-10993, and G95-1. Sterilization validated for EtO (AAMI ST27), gamma (AAMI ST32), and autoclave (AAMI ST37).
Technological Characteristics
Metallic bone fixation fastener; material: ASTM F136 titanium alloy; sterilization: EtO (AAMI ST27), gamma (AAMI ST32), or autoclave (AAMI ST37); single-use; Class II device.
Indications for Use
Indicated for patients undergoing surgical reconstruction of an ACL-deficient knee requiring interference fixation of autograft (patellar tendon-bone, semi-membranosus, semi-tendinosus) or allograft (ACL, Achilles tendon) tissues. Contraindicated for pedicular screw fixation/attachment to the cervical, thoracic, or lumbar vertebral column.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Related Devices
- K230595 — OsteoCentric ACL Fixation System · OsteoCentric Technologies · Jun 29, 2023
- K193491 — restor3d Metallic Interference Screw · Restor3D · Mar 19, 2020
- K212547 — Quick-Start Screws · Innovate Orthopaedics Limited · Dec 30, 2021
- K993166 — STRYKER BIOABSORBABLE INTERFERENCE SCREW SYSTEM · Stryker Endoscopy · Aug 4, 2000
- K212197 — WasherCap Fixation System · Abanza Tecnomed S.L · Sep 22, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Stryker Endoscopy logo. The word "stryker" is in a bold, sans-serif font, with the "s" and "t" connected. There is a registered trademark symbol to the right of the "r". Below "stryker" is the word "ENDOSCOPY" in a smaller, sans-serif font.
K9 7 2233
AUG - 5 1997
DISPOSABLES GROUP
Doug Lorang, M.S.E.
2590 Walsh Avenue Santa Clara, CA 95054 Phone: (408) 567-9100 x2541 Fax: (408) 567-2507
## SUMMARY OF SAFETY AND EFFECTIVENESS
## Device Name:
| Classification Name: | Smooth or Threaded Metallic Bone Fixation Fasteners<br>CFR 888.3040, Class II |
|----------------------------|-------------------------------------------------------------------------------|
| Device Product Code: | Panel Code 87, Orthopedic Devices, HWC |
| Common and Usual Name: | ACL Interference Screw |
| Proprietary Name: | Stryker Wedge Interference Screw System |
| Regulatory Classification: | Class II |
## Safety and Effectiveness Summary:
This summary of 510(k) safety and effectiveness is being submitted in accordance with requirements of SMDA 1990.
The Stryker Wedge Interference Screw System is intended for use in the surgical reconstruction of the anterior cruciate ligament deficient knee to provide interference fixation of the femoral and tibial bony attachments of the patellar tendon-tibial bone graft complex, the semi-membranosus tendon grafts, and the semi-tendinosus tendon grafts, or fixation of the allograft techniques such as ACL allografts and Achilles tendon allografts. The Stryker Wedge Interference Screw System is equivalent in intended use, safety and effectiveness to other fixation devices in commercial distribution. The Stryker Wedge Interference Screw System will be provided sterile for single-use applications (ASTM 4169). This device is validated for EtO (AAMI ST27) and gamma (AAMI ST32) sterilization methods, as well as autoclave steam sterilization for re-sterilization (AAMI ST37 and SSSA), with a minimum SAL of 10-0. The material of construction (ASTM F136) and its overall design are equivalent to currently marketed products. The device material has been found to be biocompatible per ASTM F136, ISO-10993, and G95-1 standards.
The screw system does not raise any new safety and efficacy concerns when compared to other devices currently on the market. Therefore, the Strvker Wedge Interference Screw System is substantially equivalent to other ACL fixation devices.
Douglas M. Lorang, M.S.E. Design Engineer
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem is a stylized image of an eagle with three human profiles incorporated into its design. The eagle's wings and body are formed by the silhouettes of the human faces.
Ford and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Douglas M. Lorang, M.S.E. Design Enqineer Stryker Endoscopy 2590 Walsh Avenue Santa Clara, California 95054
AUG - 5 1997
Re : K972233 Stryker Wedge Interference Trade Name: Screw System Requlatory Class: II Product Code: HWC Dated: June 12, 1997 Received: June 16, 1997
Dear Mr. Lorang:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting a device for pedicular screw fixation/attachment would cause the device to be adulterated under 501(f)(1) of This device, if intended for use in pedicular screw the Act. fixation/attachment, would be found not substantially equivalent and would be a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
{2}------------------------------------------------
Page 2 - Douglas M. Lorang, M.S.E.
- 1. The package insert must prominently state that the device is intended for the specific use(s) described in the enclosure only; and
- 2. You may not label or in any way promote this device for pedicular screw attachment to, or fixation of the cervical, thoracic or lumbar vertebral column. If this device is a screw with outer diameters of 3 mm - 10 mm and overall lengths of 10 mm - 75 mm inclusively, the package insert must include the following statement, "WARNING: This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine." Any pedicular screw fixation/attachment to the cervical, thoracic or lumbar vertebral column of this device is considered investigational and may only be investigated as a siqnificant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conduct of the investigation.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{3}------------------------------------------------
Page 3 - Douglas M. Lorang, M.S.E.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the ´ market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
-Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1
510(k) Number (if known): K972233
Device Name: Stryker Wedge Interference Screw System . .
Indications For Use:
The Stryker Wedge Interference Screw System is intended for use in the surgical reconstruction of the anterior cruciate ligament (ACL) deficient knee to provide interference fixation of the femoral and tibial bony attachments of the patella bone-patellar tendon-tibial bone graft complex, the semi-membranosus tendon grafts, and the semi-tendinosus tendon grafts, or fixation of the allograft techniques such as ACL allografts and Achilles tendon allografts.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K972
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
OR
Over-The-Counter Use_