INNOSAN TRAVELAIR S PLUS PORTABLE OXYGEN CONCENTRATOR, CODE NO. M407001853 (WITH AC POWER SUPPLY ACCESSORY)
K971947 · Sim Italia S.R.L. · CAW · Oct 3, 1997 · Anesthesiology
Device Facts
Record ID
K971947
Device Name
INNOSAN TRAVELAIR S PLUS PORTABLE OXYGEN CONCENTRATOR, CODE NO. M407001853 (WITH AC POWER SUPPLY ACCESSORY)
Applicant
Sim Italia S.R.L.
Product Code
CAW · Anesthesiology
Decision Date
Oct 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innosan Travelair® "S" Plus Portable Oxygen Concentrator is a mobile oxygen concentrator for use in an automobile, with power from the automobile's battery or for use with a power supply accessory (off of a mains current - 110 VAC, 60 Hz or 220 VAC, 50 Hz) when the patient is away from their primary home oxygen source, intended to provide a patient with supplemental oxygen. This device delivers oxygen to patients by physical means, using a molecular sieve bed oxygen concentrator and is designed to conserve the use of oxygen during such delivery.
Device Story
Innosan Travelair® "S" Plus is a mobile oxygen concentrator; extracts oxygen from ambient air using dual molecular sieve beds. Provides continuous (approx. 1.2 L/min) and demand (approx. 2.6 L/min) flow modes. Powered by automobile battery or AC mains via external switching power supply. Features include maneuverability (wheels/handle), hinged fluid-resistant cover, and alarms for low oxygen, high temperature, and blocked filter. Used by patients away from home; operated by patient. Output is supplemental oxygen; helps maintain oxygen therapy during travel. Healthcare provider prescribes device; patient monitors oxygen delivery via device operation.
Clinical Evidence
No clinical data. Bench testing only. Testing included electrical safety, mechanical shock/vibration, fluid spill resistance, and electromagnetic compatibility per FDA 1993 Draft Reviewer Guidance and ASTM F1464-93.
Technological Characteristics
Mobile oxygen concentrator; dual molecular sieve beds. Dimensions: 10.5" x 6.25" x 4.73" (AC power supply). Power: 110/240 VAC, 50/60 Hz or automobile battery. Connectivity: IEC-type receptacle. Standards: ASTM F1464-93, IEC-601-1. Non-software based.
Indications for Use
Indicated for patients requiring supplemental oxygen when away from their primary home oxygen source. Device is mobile, intended for use in automobiles or via AC power supply.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{1}------------------------------------------------
K971947
EXHIBIT #1 OCT - 3 1997 Page 1 of 4
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:_
#### 1. Submitter's Identification:
Ms. Susan Muratori Sim Italia s.r.I. Surgical International Marketing Via S. Donato, 156 40127 Bologna, Italy Tel: (0039) 51/502439-519161-519255 Fax: (0039) 51/504453
Date Summary Prepared: May 23, 1997
#### 2. Name of the Device:
ﺍﻟﻤﺘﺤﺪﺓ
ﻣﺴﺘﻘﺒﺔ
Innosan Travelair® "S" Plus Portable Oxygen Concentrator
#### Predicate Device Information: 3.
Innosan Travelair® "S" Portable Oxygen Concentrator, Code No. M407001853, K#963042.
#### 4. Device Description:
The Innosan Travelair® "S" Plus Portable Oxygen Concentrator is a mobile oxygen concentrator intended for use in an automobile, with power from the automobile's battery and, for use with an AC power supply accessory. The Travelair® uses two molecular sieves to extract oxygen from ambient air; this is similar to most domestic oxygen concentrators. The unit provides two switch-selectable operating modes, continuous and demand, with a nonadjustable flow of approximately 1.2 L/min continuous and approximately 2.6 L/min in the demand mode. The unit has two large wheels, one small caster, and an adjustable handle to provide maneuverability. It has a hinged cover across the top panel to protect against the entry of fluids. An alarm warns of low oxygen concentration, excessive internal temperature and blocked
{2}------------------------------------------------
air inlet filter.
# AC Power Supply Accessory Description:
The Innosan Travelair® "S" Plus Portable Oxygen Concentrator AC Power Supply Accessory, called the "Switching Power Supply for Travelair® "S" Plus Portable Oxygen Concentrator", is manufactured by an OEM vendor and packaged by Sim Italia s.r.l. specifically for the Travelair® Portable Oxygen Concentrator. The power supply has UL and TÜV certification. The code number for the power supply is 032228000. The AC power supply is housed in a plastic enclosure with approximate maximum dimensions of 10.5" x 6.25" x 4.73" high. The input power is supplied via a three-wire power cord connected to an input IEC-type receptacle. The output panel includes an ON-OFF switch and 6.5' input power cable that terminates in a connector intended for the Innosan Travelair® "S" Plus Portable Oxygen Concentrator. The input power requirements are 110/240 VAC, 50/60 Hz.
#### Intended Use: 5.
ﻟﻌﻘﺎﺑﺔ
ﻣﺘﻌﺪﻫﻪ
The Innosan Travelair® "S" Plus Portable Oxygen Concentrator is a mobile oxygen concentrator for use in an automobile, with power from the automobile's battery or for use with a power supply accessory (off of a mains current - 110 VAC, 60 Hz or 220 VAC, 50 Hz) when the patient is away from their primary home oxygen source, intended to provide a patient with supplemental oxygen. This device delivers oxygen to patients by physical means, using a molecular sieve bed oxygen concentrator and is designed to conserve the use of oxygen during such delivery.
#### 6. Comparison to Predicate Devices:
The Innosan Travelair® "S" Plus Portable Oxygen Concentrator, Code No. M407001853, is identical to our original 510(k) cleared device, K#963042, except for the addition of an AC power supply accessory.
The only difference between the Travelair® "S" device and the Travelair® "S" Plus device is that, in addition to the use of an automobile battery, it can now be line-powered with a standard power supply functioning off of a mains current (110 VAC 60 Hz/220 VAC/50 Hz).
{3}------------------------------------------------
# EXHIBIT #1 Page 3 of 4
### Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the Innosan Travelair® "S" Plus Portable Oxygen Concentrator with an AC Power Supply Accessory in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.
The following testing was conducted by CITECH:
- Input Voltage Variation a.
- Overcurrent Protection b.
- C. Controls Protection
- Connector Protective Incompatibility d.
- Mechanical Safety e.
ﺮ ﺍﻟﻤﺴﺎ
- Mechanical Vibration and Shock Resistance (IEC-601-1 (1988) f.
- Fluid Spill Resistance (IEC-601-1, Clause 44.6) g.
- Extreme Operating and Storage Temperature/Humidity h.
- Surface Temperatures ..
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was CITECH's conclusion that the Travelair® sample tested met all relevant requirements of the aforementioned test.
In addition, the following testing was conducted by CITECH:
- Radiated and Conducted Electromagnetic Energy and Magnetic Field a. Testing on the Travelair® device. Testing was conducted per the DCRND Reviewer's Guideline, November 1993.
- b. Testing was also conducted per ASTM F1464-93, "Standard Specification for Oxygen Concentrators for Domiciliary Use".
## Regarding the AC Power Supply:
The AC Power Supply Accessory for the Innosan Travelair® "S" Plus Portable Oxygen Concentrator was tested for compliance to electrical safety and environmental criteria when connected to a Travelair® Portable Oxygen
{4}------------------------------------------------
# EXHIBIT #1 Page 4 of 4
Concentrator to simulate active operating conditions in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND. The following testing was conducted by CITECH:
- Overcurrent protection a.
- Electrical leakage current b.
- Mechanical shock resistance C.
- Extreme storage and operating environments d.
- Sinusoidal vibration e.
- Surface temperatures f.
- Spill resistance . g.
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was CITECH's conclusions that the AC Power Supply sample tested met all relevant requirements of the aforementioned test.
In addition, the following testing was conducted by CITECH:
- Radiated and Conducted Electromagnetic Energy and Magnetic Field a. Testing on the AC Power Supply. Testing was conducted per the DCRND Reviewer's Guideline, November 1993.
#### 8. Discussion of Clinical Tests Performed:
Not Applicable
#### 9. Conclusions:
Status
We have demonstrated that the Innosan Travelair® "S" Plus Portable Oxygen Concentrator with an AC Power Supply Accessory is identical to the predicate device presently on the market, with the addition of an AC Power Supply, based on electrical, mechanical and environmental results as well as ASTM F1464-93, "Standard Specification For Domiciliary Use". We also adhered to FDA's Reviewer Guidance for Oxygen Concentrators (1991).
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of several curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 3 1997
Ms. Susan D. Goldstein-Falk SIM Italia s.r.l. c/o MDI Consultants, Inc. 55 Northern Boulevard, Suite 410 Great Neck, New York 11021
K971947 Re: Innosan Travelair "S" Plus Requlatory Class: II (two) 73 CAW Product Code: Dated: September 11, 1997 Received: September 15, 1997
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
EXHIBIT B Page 1 of of 1 1
510(k) Number (if known):
Device Name: Innosan Travelair(R) "S" Plus Portable Oxygen Concentrator, Code No. M407001853 (with AC Power Supply Accessory) Indications For Use:
> The Innosan Travelair® "S" Plus Portable Oxygen Concentrator is a mobile oxygen concentrator for use in an automobile, with power from the automobile's battery or for use with a power supply accessory (off of a mains current - 110 VAC, 60 Hz or 220 VAC, 50 Hz) when the patient is away from their primary home oxygen source, intended to provide a patient with supplemental oxygen. This device delivers oxygen to patients by physical means, using a molecular sieve bed oxygen concentrator and is designed to conserve the use of oxygen during such delivery.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div align="center"></div> <div align="center">(Division Sign-Off)</div> <div align="center">Division of Cardiovascular, Respiratory,</div> <div align="center">and Neurological Devices</div> |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | __________________________ |
| Prescription Use | <div align="left">✓</div> |
|----------------------|---------------------------|
| (Per 21 CFR 801.109) | |
OR
| | Over-The-Counter Use__________ |
|--|--------------------------------|
| | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.