IMPACT (INFORMING MOBILE PERSONNEL AND CARE TRACKING) PAGER SYSTEM
K971868 · Marquette Electronics, Inc. · MSX · Nov 13, 1997 · Cardiovascular
Device Facts
Record ID
K971868
Device Name
IMPACT (INFORMING MOBILE PERSONNEL AND CARE TRACKING) PAGER SYSTEM
Applicant
Marquette Electronics, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Nov 13, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Intended Use
The IMRACT Pager System is intended to be used for the annunciation of events within a zone coverage area that is secondary to the primary care that is provided. Primary care via the central station, patient bedside monitor, or telemetry system remains unchanged. This device is intended to be used within the hospital/facility environment.
Device Story
IMPACT Pager System acts as secondary alarm enunciator in hospital environments; receives patient data (arrhythmia type, heart rate, waveforms) from central stations or bedside monitors via Ethernet; transmits alerts to mobile personnel pagers. System components include CardioPager Server, Micro Serial Server, and Pager Transmitter; these operate outside immediate patient environment. Device provides zone-specific information to draw attention to patient conditions; does not replace primary monitoring systems. Healthcare providers use pager output to identify patient status; primary clinical decision-making remains dependent on primary monitoring systems. Benefits include timely notification of mobile staff regarding patient events.
Clinical Evidence
Bench testing only. No clinical data provided. Functional performance testing verified the accuracy of data replication (arrhythmia type, heart rate, waveforms) from the primary monitoring system to the pager. Testing exercised the administrator application and Unity alarm monitoring services to ensure correct information transmission.
Technological Characteristics
System components: CardioPager Server, Micro Serial Server, Pager Transmitter. Connectivity: Ethernet-based data acquisition from primary monitoring systems. Power: External power modules with U.L. listing/classification. Form factor: Server-based system with mobile pager units. Software: Administrator application and Unity alarm monitoring services.
Indications for Use
Indicated for hospital/facility personnel requiring secondary annunciation of patient events and zone information. Not intended for primary patient monitoring or alarm notification.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
K994430 — IMPACT.WF MOBILE WAVEFORM RECEIVING SYSTEM · General Electric Medical Systems Information Techn · Mar 28, 2000
K992848 — ALARMVIEW SYSTEM · Data Critical Corp. · Nov 19, 1999
K973527 — CARDIO-PAGER SYSTEM · Data Critical Corp. · Mar 31, 1998
K992749 — ULTRAVIEW WAVEFORM PAGER SYSTEM, MODEL 91841 · Spacelabs Medical, Inc. · Jan 14, 2000
K013156 — MODIFICATION TO ALARMVIEW WIRELESS DATA NETWORK SYSTEM · Data Critical Corp. · Oct 19, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
14718608
NOV 1 3 1997
# 510(k) Summary of Safety and Effectiveness
#### 1. Submitter
Marquette Medical Systems 8200 West Tower Avenue Milwaukee, WI 53223 USA
Establishment Registration Number 2124823
Contact Name / Telephone Number Dianne Schmitz Corporate Regulatory Affairs Marquette Medical Systems
Phone: (414) 362-3230
Date: 8 May 97
### General Information 2.
### Trade/Proprietary Name
Marquette's name for this device is IMPACT (Informing Mobile Personnel and Care Tracking) Pager System.
### Common/Usual Name
This device is commonly known as a pager.
### Device Classification
This device is unclassified according to a review performed of the CDRH Manual -FDA 91-4246 Classification Names for Medical Devices and In Vitro Diagnostic Products, as well as the Diogenes - The 510(k) Register and database.
## Performance Standards
Performance standards (Section 514 of the Act) have not yet been established for the device that is the subject of this premarket notification submission.
### Legally Marketed Predicate Device(s) 3.
The legally marketed predicate product is the ADU/ Pager-LAN System, K962827 .
#### র্বা Device Description and Intended Use
The IMPACT pager is intended for the annunciation of events to provide zone information that is secondary to the primary care provided. It provides specific information within a zone and alerts or draws an identified individual's aftention to a defined patient condition in a timely manner. Primary care via the central station, telemetry system, or patient bedside monitor remains unchanged by the addition of the IMPACT pager.
{1}------------------------------------------------
### Test Summary & Conclusion 6.
The CardioPager Server, Micro Serial Server, and the Pager Transmitter are not used in the patient environment. These devices are purchased from the manufacturer with appropriate U.L. listing or classification on the device and/or the external power modules. The devices' U.L. Listing or Classification was reviewed by Marquette's Technical Coordinator and found to be appropriate. Marquette's reliability group also performed additional electrical and mechanical performance and safety tests which were specified and included in the submission.
While the CardioPager system functions as a secondary enunciator of alams, the bedside monitor and central station make up the patient monitoring system. Patient waveform acquisition, determination of alarm conditions, and alam notifications are primary functions of the patient monitoring system, which has been previously verified for accuracy. Because the CardioPager replicates arthythmia type, heart rate, and waveform information obtained from the monitoring system via Ethemet, this Information was compared for correctness during the functional performance testing.
Functional performance testing of the CardioPager system was accomplished and provided in the submission. This test plan fully exercised the administrator application running on the server as well as the operation of the Unity™ alarm monitoring services of the CardioPager server. Additionally, the information at the pager was checked for correctness, as mentioned above.
Validation test results indicated that the IMPACT Pager System met the requirements of its intended use. Again, this information is secondary to the primary care provided. Primary care remains unchanged by the addition of the IMPACT pager.
Marquette Medical Systems has demonstrated that use of the IMPACT Pager System is as safe and effective, and performs substantially equivalent to use of the ADU/ Pager-LAN System.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a partial view of a document with a logo and some text. On the left side, there is a logo that appears to be a stylized representation of human faces or profiles. To the right of the logo, there is some text, but it is not fully visible in the image. The text seems to be part of a heading or title, possibly related to the organization or department associated with the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1 3 1997
Ms. Dianne Schmitz Marquette Medical Systems 8200 West Tower Avenue Milwaukee, Wisconsin 53223
Re : K971868 IMPACT (Informing Mobile Personnel and Care Tracking) Pager System Regulatory Class: III (three) Product Code: 74 MSX Dated: September 5, 1997 Received: September 8, 1997
Dear Ms. Schmitz:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Dianne Schmitz
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial ----------equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Unknown; 510(k) filed on May 16, 1997
Stop(s) Number (if known)
IMPACT (Informing Mobile Personnel and Care Tracking) Device Name: Pager System
Indications For Use:
The IMRACT Pager System is intended to be used for the annunciation of events within a zone coverage area that is secondary to the primary care that is provided. Primary care via the central station, patient bedside monitor, or telemetry system remains unchanged.
This device is intended to be used within the hospital/facility environment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chaison Formen for Arc
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Prescription Use
(Per 21 CFR 801.109)
Image /page/4/Picture/16 description: The image shows a sequence of numbers. The numbers are 000005. There are five zeros followed by a five.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.