SYSMEX IMMATURE RETICULOCYTE FRACTION (IRF) PARAMETER

K971736 · Sysmex Corp. · GKZ · Aug 26, 1997 · Hematology

Device Facts

Record IDK971736
Device NameSYSMEX IMMATURE RETICULOCYTE FRACTION (IRF) PARAMETER
ApplicantSysmex Corp.
Product CodeGKZ · Hematology
Decision DateAug 26, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Intended Use

The IRF is a clinical parameter obtained from the Sysmex™ R-Series automated reticulocyte analyzers. It is intended for in-vitro use in clinical laboratories. Its clinical use is to monitor bone marrow suppression and recovery in terms of erythropoiesis in situations of previous cancer chemotherapeutic bone marrow suppression and bone marrow transplantation. Qualified laboratory personnel are responsible for review of all abnormal results.

Device Story

Sysmex R-Series analyzers (R-1000, R-3000, RAM-1) function as automated flow cytometers. Input: whole blood samples. Process: dilution and staining with fluorescent dye (Auromine-O); measurement of fluorescence intensity and light scatter. Output: scattergram displaying reticulocyte fluorescence intensity, categorized into Low (LFR), Middle (MFR), and High (HFR) fluorescence regions. The IRF parameter is calculated as the sum of MFR and HFR ratios. Used in clinical laboratories by qualified personnel to monitor erythropoiesis. Output assists clinicians in assessing bone marrow suppression and recovery status following chemotherapy or transplantation.

Clinical Evidence

Clinical studies performed on patients receiving myeloablative chemotherapy and patients undergoing bone marrow transplantation. Evidence supports the IRF parameter as a predictive indicator of bone marrow suppression and recovery, showing agreement with WBC and ANC counts and comparable or superior performance to standard reticulocyte counts.

Technological Characteristics

Automated flow cytometry; uses fluorescent dye (Auromine-O) for cell staining. Measures fluorescence intensity and light scatter. Table-top analyzer form factor. Parameters reported as ratios/percentages of total reticulocyte count. Software-based categorization of fluorescence intensity into LFR, MFR, and HFR regions.

Indications for Use

Indicated for monitoring bone marrow suppression and recovery in patients undergoing cancer chemotherapy or bone marrow transplantation. For use in clinical laboratories by qualified personnel.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness Sysmex Corporation of America Gilmer Road 6699 RFD Long Grove, IL 60047-9596 (847)726-3662 Fax: (847)726-3505 Attention: Cathy Trester Date Prepared: May 5, 1997 AUG 26 1997 ## Sysmex™ IRF parameter (MFR+HFR) on R-1000, R-3000, RAM-1 analyzers, Trade name and Common name The Sysmex™ IRF parameter is substantially equivalent to the reticulocyte count, and shows agreement to the WBC count as predictive indicators of bone marrow suppression and recovery. The IRF parameter is obtained from the Sysmex R-1000(Premarket Notification K894539) R-3000 (Premarket Notification K912494) or SE/RAM-1(Premarket Notification K964375). The Sysmex™ R-3000, R-1000 and SE/RAM-1 are classified as Class II devices. The IRF parameter is classified as a Class III device. #### Intended Use The IRF is a clinical parameter obtained from the Sysmex™ R-Series automated reticulocyte analyzers. It is intended for in-vitro use in clinical laboratories. Its clinical use is to monitor bone marrow suppression and recovery in terms of erythropoiesis in situations of previous cancer chemotherapeutic bone marrow suppression and bone marrow transplantation. Qualified laboratory personnel are responsible for review of all abnormal results. ### Device Description and Principles The R-1000, R-3000 and RAM-1 are table-top analyzer systems for automated reticulocyte counting. These instruments are dedicated flow cytometers which dilute and stain whole blood with a fluorescent dye (Auromine-O), then count and measure fluorescence and scatter of stained blood cells. The fluorescence intensity is measured, and the analyzer identifies the reticulocytes based on fluorescence and scatter. The fluorescent intensity of the reticulocytes is displayed on the analyzer as a scattergram, and this display is separated into three regions: Low Fluorescence Intensity (LFR), Middle Fluorescence Intensity (MFR), and High Fluorescence Intensity (HFR). These regions are reported as a ratio or percentage(sum = 100%), in addition to the analyzer's reporting total reticulocyte count and RBC count. The IRF parameter is determined by the sum of the MFR (Middle Fluorescence Ratio) and HFR (High Fluorescence Ratio) {MFR+HFR=IRF}. ### Comparison to Predicate Methodology and Clinical Data In the area of assessment of bone marrow production, the predicate methodology is use of the white blood cell count (WBC), the absolute neutrophil count (ANC), or the reticulocyte count. The Immature Reticulocyte Fraction is similar to WBC and ANC, and similar or better than reticulocyte count as a monitor for bone marrow production. Clinical studies have been performed which support this claim for patients receiving myeloablative {1}------------------------------------------------ chemotherapy. In addition, several clinical studies have been published which support this claim for patients having undergone bone marrow transplantation. ## Conclusion The Sysmex™ Immature Reticulocyte Fraction (IRF) parameter provided by the R-Series reticulocyte analyzers has been shown to be a parameter of predictive value for bone marrow suppression and recovery. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is centered horizontally. The words are capitalized, and the overall appearance is clean and professional. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Catherine M. Trester, M.T. (ASCP) Clinical and Regulatory Affairs SYSMEX™ Corporation Gilmer Road, 6699 RFD AUG 26 1997 Long Grove, Illinois 60047-9596 ... ... Re : K971736/S1 Trade Name: Sysmex™ Immature Reticulocyte Fraction (IRF) Parameter Requlatory Class: III Product Code: GKZ July 16, 1997 Dated: Received: July 18, 1997 Dear Ms. Trester: We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requiation (QS) for Medical Devices: General requiation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page_ 510(k) Number (if known): Device Name: Immature Reticulocyte Fraction on R-Series Analyzers Indications For Use: The IRF parameter of the R-Series analyzers is intended for in-vit-vouse in --clinical laboratories. Its clinical use is to monitor bone marrow suppression and recovery in terms of erythropoiesis in situations of previous cancer chemotherapeutic bone marrow suppression and bone marrow transplantation. Qualified laboratory personnel are responsible for review of all abnormal results. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number **__** Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use **__** (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...