K971715 · Sulzer Orthopedics, Inc. · KWS · Aug 7, 1997 · Orthopedic
Device Facts
Record ID
K971715
Device Name
SELECT SHOULDER OFFSET HUMERAL HEAD COMPONENTS
Applicant
Sulzer Orthopedics, Inc.
Product Code
KWS · Orthopedic
Decision Date
Aug 7, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Select Shoulder Offset Humeral Heads, when used with one of the Select Shoulder Humeral Stems and Glenoid Components (if applicable), are intended for use in treatment of the following: 1. Patient conditions, including but not limited to, noninflammatory degenerative joint disease (NIDJD), e.g., osteoarthritis or post-traumatic arthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. 2. Complex acute fractures, fracture-dislocations of the humeral head, malunion or non-union of a small osteoporotic head fragment, chronic, recurrent or acute dislocation with loss of humeral head cartilage, or large impression fractures. 3. Those patients with failed previous surgery where pain, deformity, or dysfunction persists. 4. Failed previous surgery, including joint reconstruction, internal fixation, nonunion of the humeral neck, arthrodesis or hemiarthroplasty. 5. Cuff tear arthropathy. 6. Avascular necrosis or osteonecrosis of the humeral head. 7. Tumor resection.
Device Story
Select Shoulder Offset Humeral Heads are metallic orthopedic implants manufactured from cobalt chrome alloy (CoCrMo). Designed for articulation with human glenoid or all-poly glenoid components; features female taper for attachment to Select Shoulder humeral stems. Underside includes clock-like numbering (1-12) for surgeon reference to select head position relative to glenoid. Offset taper location facilitates insertion, anatomic soft tissue balancing, and improved glenoid contact. Used by orthopedic surgeons in clinical settings for shoulder arthroplasty. Device provides structural replacement of humeral head to restore joint function, reduce pain, and improve stability in patients with degenerative or traumatic shoulder conditions.
Clinical Evidence
Bench testing only. Testing included pulloff strength measurements and contact area analysis at various levels of abduction to verify mechanical performance and joint contact characteristics compared to predicate devices.
Technological Characteristics
Metallic humeral head component; material: cobalt chrome alloy (CoCrMo). Features female taper for modular attachment to humeral stems. Articulating surface is polished; underside may be polished or grit blasted. Includes clock-like numbering (1-12) for orientation. Available in various heights and diameters.
Indications for Use
Indicated for patients requiring shoulder joint replacement due to noninflammatory degenerative joint disease (osteoarthritis, post-traumatic arthritis), inflammatory joint disease (rheumatoid arthritis), complex acute fractures, fracture-dislocations, malunion/non-union, chronic/recurrent dislocation, cuff tear arthropathy, avascular necrosis, or tumor resection.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Tornier Aequalis Shoulder System
Depuy Global Total Shoulder System
Biomet Bio-Modular Total Shoulder
Zimmer Fenlin Total Shoulder
Kirschner Modular Shoulder System
3M/Orthomet Modular Neer II Shoulder System
Encore Foundation Shoulder System
Submission Summary (Full Text)
{0}------------------------------------------------
I
## 510(k) SUMMARY
April 29, 1997
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21CFR 807, this is to serve as a 510(k) Summary for the Sulzer Orthopedics Inc. Select® Shoulder Offset Humeral Heads.
| Submitter: | Sulzer Orthopedics Inc.<br>9900 Spectrum Drive<br>Austin, TX 78717<br>(512) 432-9900 |
|----------------------|--------------------------------------------------------------------------------------|
| Contact Person: | Jacquelyn Hughes<br>Manager, Regulatory Affairs |
| Classification Name: | No formal classification has been established for humeral head<br>components. |
| Common/Usual Name: | Humeral Head Components |
| Trade/Proprietary: | Select® Shoulder Offset Humeral Heads |
## Product Description/Substantial Equivalence:
The Select Shoulder Offset Humeral Heads are metallic components manufactured from cobalt chrome alloy (CoCrMo). The heads, once impacted onto one of the Select Shoulder humeral stems, are designed to articulate with the normal human glenoid or with a replacement all-poly glenoid component. The humeral heads feature a female taper which allows for attachment to the male taper of the Sulzer Orthopedics Select Shoulder Humeral Stems. The articulating surface of the head is polished. The underside of the head, with the exception of the taper attachment feature, may be either polished or grit blasted. Similar to the face of a clock, the underside of the head is numbered from 1-12, giving the surgeon the ability to reference and select the position of the head relative to the glenoid. The location of the female taper is offset from center, thus permitting ease of insertion into the joint, allowing closer replication of the normal head anatomy permitting accurate anatomic soft tissue balancing, and providing greater contact with the glenoid when the humerus is in a normal resting position. The heads are available in a variety of heights and diameters. The humeral heads are designed for use with Sulzer Orthopedics Humeral Stems and/or glenoid components which come in a variety of sizes for increased stability of the glenohumeral joint.
Testing indicated that the pulloff strengths for the Select Shoulder Offset Humeral Heads were comparable to currently marketed devices. Contact area testing indicated that the offset heads provide adequate contact area at various levels of abduction.
The Select Shoulder CoCr Humeral Heads are similar to those of the Tornier Aequalis Shoulder System, the Depuy Global Total Shoulder System, the Biomet Bio-Modular Total Shoulder, the Zimmer Fenlin Total Shoulder, the Kirschner Modular Shoulder System, the 3M/Orthomet Modular Neer II Shoulder System, and the Encore Foundation Shoulder System.
## UUUZZA
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mitchell A. Dhority, RAC Senior Regulatory Affairs Specialist ---------SulzerMedica ¨ Sulzer Orthopedics, Inc. AUG - 7 1997 9900 Spectrum Drive Austin, Texas 78717
Re : K971715 Trade Name: Select Shoulder Offset Humeral Heads Regulatory Class: III Product Codes: KWS, KWT, and HSD Dated: May 7, 1997 Received: May 9, 1997
Dear Mr. Dhority:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that,
{2}------------------------------------------------
Page 2 - Mitchell A. Dhority, RAC
through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: " this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
ia M. Witten, Ph.D., M.D. Cel Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
|
|
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Select Shoulder Offset
Device Name: Humeral Heads
Indications for Use:
The Select Shoulder Offset Humeral Heads, when used with one of the Select Shoulder Humeral Stems and Glenoid Components (if applicable), are intended for use in treatment of the following: ... .. . ...........
- Patient conditions, including but not limited to, noninflammatory degenerative joint 1. disease (NIDJD), e.g., osteoarthritis or post-traumatic arthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
- Complex acute fractures, fracture-dislocations of the humeral head, malunion or non-2. union of a small osteoporotic head fragment, chronic, recurrent or acute dislocation with loss of humeral head cartilage, or large impression fractures.
- Those patients with failed previous surgery where pain, deformity, or dysfunction 3. persists.
- Failed previous surgery, including joint reconstruction, internal fixation, nonunion of 4. the humeral neck, arthrodesis or hemiarthroplasty.
- Cuff tear arthropathy. న.
- Avascular necrosis or osteonecrosis of the humeral head. 6.
- 7. Tumor resection.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
evice Evaluation (ODE)
(Division Sign-Off)
Division of Genarai Restorative Devices
510(%) Numbar. K971715
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Image /page/3/Picture/19 description: The image shows a sequence of numbers. The numbers are all zeros except for the last digit, which is an eight. The sequence of numbers is "000008".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.