K971706 · Nobel Biocare Uas, Inc. · NHA · Jul 21, 1997 · Dental
Device Facts
Record ID
K971706
Device Name
TIADAPT ABUTMENT SYSTEM
Applicant
Nobel Biocare Uas, Inc.
Product Code
NHA · Dental
Decision Date
Jul 21, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nobel Biocare's TiAdapt™ Abutment System is a set of screw retained preparable abutments that are secured to an endosseous implant and are intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.
Device Story
TiAdapt Abutment System consists of preparable titanium abutments, abutment screws, counter torque devices, adaptors, screwdrivers, temporary cylinders, and impression copings. System secures to Brånemark System dental implants via gold abutment screw. Functions as anchor for cemented single crowns or multi-unit bridges. Used by dental professionals in clinical settings to restore patient chewing function. Components facilitate prosthetic attachment to endosseous implants.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System components constructed of titanium. Includes preparable abutments, gold abutment screws, counter torque devices, and manual screwdrivers. Designed for use with Brånemark System endosseous dental implants. Mechanical fixation via screw-retained connection. No software or electronic components.
Indications for Use
Indicated for patients requiring dental restoration via endosseous implants to restore chewing function; used as an anchor for cemented prosthetic devices such as artificial teeth.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
JUL 21 1997
Pre-market Notification Page -3-�������������������������������
### 510(k) Summary of Safety and Effectiveness VII.
### A. Name and Address
This Summary of Safety and Effectiveness is being submitted by Nobel Biocare USA, Inc., 777 Oakmont Lane, Suite 100, Westmont, IL 60559. The telephone number is: (630) 654-9100, extension 2002, and the contact person will be Betsy Brown, the Vice President, Regulatory Affairs.
# B. Name of the Device
This device system consists of various preparable titanium abutments, abutment screws, a counter torque device with handle, adaptors, machine Uni-grip screwdrivers with autoclavable box, temporary cylinders, temporary tubes, laboratory screws and impression copings.
## C. The Predicate Product
The predicate products used in this Pre-market Notification are similar components marketed by Nobel Biocare including the CeraOne Abutments (K910611), Ceramic Abutment (K913255), CeraOne™ Abutment Screw (K961737), Guide Pin (K911969), Abutment Lab-Screw (K925769), CerAdapt™, Machine Counter Torque Device (K953775), Estheticone™ Temporary Cylinder (K925766), Miruscone Temporary Cylinder (K962403), Temporary Tube (K926766), Titanium Impression Coping (K952448), Plastic Impression Coping (K910611), CeraOne™ Screwdrivers (K922805) and Fixture Guide Set Tray (K944963).
# D. Description of Device
The Nobel Biocare TiAdapt™ Abutment System is comprised of preparable titanium abutments which are secured to the Brånemark System® dental implants with a gold abutment screw. This system functions as an anchor to which a single crown can be cemented. When more than one preparable abutment is used, a multi-unit bridge can be cemented to the abutments. The TiAdapt Abutment System also includes a counter torque device and adaptors, temporary cylinders, impression copings and machine Uni-grip screwdrivers.
{1}------------------------------------------------
Pre-market Notification Page -4-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
E. Intended Use of the Device
Nobel Biocare's TiAdapt™ Abutment System is a set of screw retained preparable abutments that are secured to an endosseous implant and are intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.
F. Comparison of Technological Characteristics
The technological characteristics between the components of the TiAdapt Abutment System and the corresponding predicate products found in the Brånemark System® are identical.
### Summary Of Safety And Effectiveness Concerns
Safety and effectiveness problems that have been encountered with similar Abutment Systems used with endosseous implants that are currently being marketed include:
- 1. occasional fractures of the screw attaching the abutment to the fixtures usually due to functional overload from masticastory forces:
- 2. screws working loose (usually because torque force below recommended values was applied when the abutment was attached to the fixture) may lead to the formation of granulation tissue at the level of the fixture and abutment connection which may, in turn, result in infection:
- 3. improper initial seating of abutments resulting in gingival inflammation and fistulae formation, both conditions resolve when proper seating of the abutment is accomplished.
- 4. occasional fracture of the abutment screw which is usually caused by poorly designed and/or fabricated restorations that creates overloads or cause's metal fatigue.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem featuring a stylized depiction of an eagle or bird-like figure with outstretched wings. The emblem is rendered in a blue color, contrasting with the white background of the seal.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 1 7 2010
Ms. Phuong Nguyen Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Lind, California 92887
Re: K971706
Trade/Device Name: TiAdapt Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 28, 2010 Received: July 29, 2010
Dear Mr. Nguyen:
This letter corrects our substantially equivalent letter of July 29, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Nguyen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Supa Viner
Anthony D. Watson, B.S., M.S., M.B.A.
Director
Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: TiAdapt™ Abutment System_
Indications For Use:
Nobel Biocare's TiAdapt Abutment System is a set of screw retained preparable abutments which are secured to an endosseous implant and is intended to function as an anchor to which prosthetic devices, such as artificial teeth, may be attached using dental cement to restore a patient's chewing function.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | Field W. Shippen |
|----------------------------------------|------------------|
| Division of Dental, Infection Control, | |
| and General Hospital Devices | |
| 510(k) Number | K971706 |
Prescription Use. (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.