ETEST FOR LEVOFLOXACIN

K971697 · Ab Biodisk · JWY · Jul 14, 1997 · Microbiology

Device Facts

Record IDK971697
Device NameETEST FOR LEVOFLOXACIN
ApplicantAb Biodisk
Product CodeJWY · Microbiology
Decision DateJul 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Intended Use

For in vitro diagnostic use: Etest is a quantitative technique for the determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, Pneumococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC) in ug/ml of individual antibiotics against bacteria as tested on agar media by overnight incubation. This Etest 510(k) application is for MIC determination of Levofloxacin in the range of 0.002 - 32 µg/ml with non-fastidious Gram negative and Gram positive aerobic bacteria, S. pneumoniae and H. influenzae.

Device Story

Etest is a quantitative in vitro diagnostic device for determining antimicrobial susceptibility; utilizes a predefined antibiotic gradient on a test strip; applied to agar media inoculated with bacterial samples; incubated overnight; produces a Minimum Inhibitory Concentration (MIC) value in µg/ml; used in clinical laboratories by trained personnel; provides clinicians with susceptibility data to guide antibiotic therapy decisions; benefits patients by enabling targeted antimicrobial treatment.

Clinical Evidence

No clinical data provided; device relies on bench testing for MIC determination.

Technological Characteristics

Quantitative antimicrobial susceptibility test; consists of a reagent-impregnated strip creating a stable antibiotic concentration gradient on agar; manual interpretation of MIC based on inhibition ellipse intersection with the strip; standalone diagnostic test.

Indications for Use

Indicated for in vitro diagnostic determination of antimicrobial susceptibility (MIC) of Levofloxacin (0.002 - 32 µg/ml) in non-fastidious Gram-negative and Gram-positive aerobic bacteria, S. pneumoniae, and H. influenzae.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Anne Bolmström President AB Biodisk Dalvägen 10 S-171 36 Solna Sweden JUL 1 4 1997 - 111 - - Re: K971697 Trade Name: Etest for Levofloxacin Regulatory Class: II Product Code: JWY Dated: May 6, 1997 Received: May 8, 1997 Dear Ms. Bolmström: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Paris 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {1}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Autman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page 1 of 1 510(k) Number (if known): Device Name: Etest® for Levofloxacin Indications For Use: For in vitro diagnostic use: Etest is a quantitative technique for the determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, Pneumococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC) in ug/ml of individual antibiotics against bacteria as tested on agar media by overnight incubation. This Etest 510(k) application is for MIC determination of Levofloxacin in the range of 0.002 - 32 µg/ml with non-fastidious Gram negative and Gram positive aerobic bacteria, S. pneumoniae and H. influenzae. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | <i>Air Perl</i> | |--|---------------------| | | (Division Sign-Off) | Division of Clinical Laboratory Devices | 510(k) Number | <i>K971697</i> | |---------------|----------------| |---------------|----------------| | Prescription Use | OR | Over-The Counter Use | |------------------------------------------------------------------------|----|-------------------------------------------------------| | (Per 21 CFR 801.109) <span style="text-decoration: overline;">✓</span> | | <span style="text-decoration: overline;">_____</span> | (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...