KSEA ARTHROPUMP PLUS

K971649 · KARL STORZ Endoscopy-America, Inc. · HRX · Oct 29, 1997 · Orthopedic

Device Facts

Record IDK971649
Device NameKSEA ARTHROPUMP PLUS
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeHRX · Orthopedic
Decision DateOct 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Intended Use

These instruments are intended for use by qualified surgeons and provide irrigation/aspiration to the surgical site during arthroscopic surgical procedures.

Device Story

Arthropump Plus is a roller-type irrigation/aspiration pump; provides fluid management during arthroscopic surgery. Device features user-adjustable irrigation flow rate, irrigation pressure, and aspiration suction underpressure. Operated by surgeons in clinical/surgical settings to maintain surgical site visibility and environment. Output consists of controlled fluid delivery and suction; assists in clearing surgical field. Benefits include improved visualization and site management during arthroscopy.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Roller-type pump mechanism; user-adjustable controls for irrigation flow, pressure, and aspiration suction. Class II device.

Indications for Use

Indicated for use by qualified surgeons to provide irrigation and aspiration to the surgical site during arthroscopic surgical procedures.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K971649 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 2 9 1997 OCT - This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge. | Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe Drive<br>Culver City, CA 90230<br>(310) 558-1500 | |------------------------|---------------------------------------------------------------------------------------------------------------| | Contact: | Kevin Kennan<br>Regulatory Affairs Specialist | | Device Identification: | Common Name:<br>Suction/irrigation Pump | | | Trade Name: (optional) | Indication: The Karl Storz Arthropump Plus system is designed to provide irrigation and aspiration of the surgical site during arthroscopic surgical procedures. Karl Storz Arthropump Plus Device Description: The Karl Storz Arthropump Plus and accessories is a roller type irrigation/aspiration pump with user adjustable irrigation flow rate, irrigation pressure, and aspiration suction underpressure. Substantial Equivalence: The Karl Storz Arthropump Plus system is substantially equivalent to the predicate devices since the basic features and intended uses are similar. The minor differences between the Karl Storz Arthropump Plus and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or intended use of these devices. Signed: Blin Blanett Kevin Kennan Regulatory Affairs Specialist 000130 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines that suggest feathers or wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular fashion around the bird symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Kevin Kennan Regulatory Affairs Specialist Karl Storz Endoscopy 600 Corporate Pointe Culver City, California 90230-7600 OCT 2 9 1997 Re: K971649 Trade Name: Arthropump Plus Regulatory Class: II Product Code: HRX Dated: July 21, 1997 Received: August 14, 1997 Dear Mr. Kennan: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Mr. Kevin Kennan This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free would number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, fcolle Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Storz Endoscopy. The logo is in black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "Karl Storz Endoscopy" in a smaller font. The logo is simple and easy to read. 510(k) Number (if known): Not yet assigned K971649 Device Name: Arthropump Plus Indications for Use: These instruments are intended for use by qualified surgeons and provide irrigation/aspiration to the surgical site during arthroscopic surgical procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of General Restorative Devices | | | 510(k) Number | K971649 | | Prescription Use:<br>(Per 21 CFR 801.109) | X | OR Over-The-Counter Use: | |-------------------------------------------|---|--------------------------| |-------------------------------------------|---|--------------------------| (Optional Format 1-2-96)000003
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%