K971582 · Nipro Medical Corp. · LJT · Dec 22, 1997 · General Hospital
Device Facts
Record ID
K971582
Device Name
NIPRO CORELESS NEEDLE/CORELESS NEEDLE SET
Applicant
Nipro Medical Corp.
Product Code
LJT · General Hospital
Decision Date
Dec 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
Device Story
Nipro Coreless Needle and Needle Set; designed to prevent coring during access to implanted reservoirs. Device consists of luer hub, stainless steel (Type 304) cannula, and protector. Needle sets include soft PVC tubing, roller clamp, and luer connectors; some models feature Y-shaped tubing for mixture injection. Used by clinicians to deliver drug solutions into implanted reservoirs. Prevents coring of the reservoir septum, ensuring integrity of the implanted system.
Clinical Evidence
No clinical data provided; substantial equivalence is based on identical design and technical characteristics to the predicate device.
Indicated for the injection of drug solutions into an implanted reservoir in the body.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
Related Devices
K113469 — NIPRO HUBER INFUSION, EXCEL HUBER NEEDLE · Nipro Medical Corporation · Oct 18, 2012
K113471 — NIPRO HUBER INFUSION SET, EXCEL HUBER INFUSION SET · Nipro Medical Corporation · Oct 18, 2012
K211121 — Intera Non-coring (Huber) Refill Needles, Intera Non-coring (Huber) Special bolus Needles and OR Prep Kit · Intera Oncology, Inc. · Oct 14, 2021
K071846 — MODIFICATION TO EZ HUBER SAFETY INFUSION SET · Pfm Medical, Inc. · Aug 30, 2007
K060580 — K-SHIELD PORT ACCESS INFUSION SET · Kawasumi Laboratories, Inc. · Jun 6, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
K971582
Image /page/0/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized, abstract symbol to the left of the word "NIPRO" in bold, sans-serif font. The symbol appears to be two interlocking shapes, possibly representing a stylized "N".
DEC 22 1997
CORPORATION 10877 N.W. 33rd Street Miami. Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
## SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO® CORELESS NEEDLE AND CORELESS NEEDLE SET
\$807.92 (a)(1)
Contact Person:
Luis Candelario General Manager
Address:
Nipro Medical Corporation 10877 NW 33rd Street Miami, Florida 33172 (305) 599-7174 (telephone) (305) 599-8454 (fax)
Date of Summary Preparation:
April 30, 1997
\$807.92 (a)(2)
Nipro® Coreless Needle and Nipro Coreless Needle Set Trade Name:
Hypodermic single lumen needle or I.V. Administration Set Common Name:
Classification Name: Hypodermic single lumen needle (21 CFR §880.5570) or Intravascular administration set (21 CFR §880.5440)
\$807.92 (a)(3)
Legally Marketed Substantially Equivalent Device: Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
8807.92 (a)(4)
Description of Device: Nipro Coreless Needles and Nipro Coreless Needle Sets are described in Hypodermic single lumen needle (21 CFR 8880.5570) or Intravascular administration set (21 CFR \$880.5440). The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
{1}------------------------------------------------
The Nipro Coreless Needles and Nipro Coreless Needle Sets that we intend to market include 2 different types: the straight type (7 sizes) and the 90° angle type (7 sizes). The Nipro Coreless Needle consists of a luer hub, a cannula, and a protector over the cannula. The needle base and the needle protector are composed of polypropylene (PP) and the cannula is made of stainless steel (Type 304).
The Nipro Coreless Needle Sets consist of a needle connected to a length (100 mm or 450 mm) of soft polyvinylchloride (PVC) tubing with a clamp roller along it and a luer connector and luer cap at the other end. Models in which there is a Yshaped tube for mixture injection include a second length of PVC tubing connected at the first luer connector with a clamp roller along it which is then connected to another luer connector and cap.
The materials used for the components include stainless steel (Type 304), polyethylene (PE), polyvinylchloride (PVC), polypropylene (PP), polycarbonate (PC), natural rubber (NR), and polyoxymethylene (POM).
\$807.92 (a)(5)
Intended Use:
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
§807.92 (a)(6)
Comparison of Technical Characteristics:
The Nipro Medical Corporation and Exel International, Inc. devices are exactly the same devices, therefore, all technical characteristics are identical.
s
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 1997
Nipro Medical Corporation C/O Kaelyn B. Hadley Consultant 1384 Copperfield Court Lexington, Kentucky 40514-1268
Re: K971582 Trade Name: NIPRO® Huber Needles and Huber Needle Sets Requlatory Class: II Product Code: LJT Dated: October 30, 1997 Received: November 3, 1997
Dear Ms. Hadley:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does
{3}------------------------------------------------
Page 2 - Ms. Hadley
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours
Timothy A. Ulatowski
Directo Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
K971582
Image /page/4/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized, abstract symbol to the left of the word "NIPRO" in bold, sans-serif font. The symbol appears to be two interlocking shapes, possibly representing a stylized "N".
DEC 22 1997
CORPORATION 10877 N.W. 33rd Street Miami. Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
## SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO® CORELESS NEEDLE AND CORELESS NEEDLE SET
\$807.92 (a)(1)
Contact Person:
Luis Candelario General Manager
Address:
Nipro Medical Corporation 10877 NW 33rd Street Miami, Florida 33172 (305) 599-7174 (telephone) (305) 599-8454 (fax)
Date of Summary Preparation:
April 30, 1997
\$807.92 (a)(2)
Nipro® Coreless Needle and Nipro Coreless Needle Set Trade Name:
Hypodermic single lumen needle or I.V. Administration Set Common Name:
Classification Name: Hypodermic single lumen needle (21 CFR §880.5570) or Intravascular administration set (21 CFR §880.5440)
\$807.92 (a)(3)
Legally Marketed Substantially Equivalent Device: Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
8807.92 (a)(4)
Description of Device: Nipro Coreless Needles and Nipro Coreless Needle Sets are described in Hypodermic single lumen needle (21 CFR 8880.5570) or Intravascular administration set (21 CFR \$880.5440). The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
{5}------------------------------------------------
The Nipro Coreless Needles and Nipro Coreless Needle Sets that we intend to market include 2 different types: the straight type (7 sizes) and the 90° angle type (7 sizes). The Nipro Coreless Needle consists of a luer hub, a cannula, and a protector over the cannula. The needle base and the needle protector are composed of polypropylene (PP) and the cannula is made of stainless steel (Type 304).
The Nipro Coreless Needle Sets consist of a needle connected to a length (100 mm or 450 mm) of soft polyvinylchloride (PVC) tubing with a clamp roller along it and a luer connector and luer cap at the other end. Models in which there is a Yshaped tube for mixture injection include a second length of PVC tubing connected at the first luer connector with a clamp roller along it which is then connected to another luer connector and cap.
The materials used for the components include stainless steel (Type 304), polyethylene (PE), polyvinylchloride (PVC), polypropylene (PP), polycarbonate (PC), natural rubber (NR), and polyoxymethylene (POM).
\$807.92 (a)(5)
Intended Use:
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
§807.92 (a)(6)
Comparison of Technical Characteristics:
The Nipro Medical Corporation and Exel International, Inc. devices are exactly the same devices, therefore, all technical characteristics are identical.
s
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in black and have a flowing, wave-like quality. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 1997
Nipro Medical Corporation C/O Kaelyn B. Hadley Consultant 1384 Copperfield Court Lexington, Kentucky 40514-1268
Re: K971582 Trade Name: NIPRO® Huber Needles and Huber Needle Sets Requlatory Class: II Product Code: LJT Dated: October 30, 1997 Received: November 3, 1997
Dear Ms. Hadley:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does
{7}------------------------------------------------
Page 2 - Ms. Hadley
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours
Timothy A. Ulatowski
Directo Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known) Device Name: Nipro Coreless Needle and Nipro Coreless Needle Set
Indications for Use: The Nipro Coreless Needle and the Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ... 1
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|-----------------------------------------------------------------------------------------|----|----------------------|
| (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices | | |
| 516(k) Number Use | OR | Over-the-Counter Use |
| (Per 21 CFR 801.109) | | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.