K971582 · Nipro Medical Corp. · LJT · Dec 22, 1997 · General Hospital
Device Facts
Record ID
K971582
Device Name
NIPRO CORELESS NEEDLE/CORELESS NEEDLE SET
Applicant
Nipro Medical Corp.
Product Code
LJT · General Hospital
Decision Date
Dec 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
Device Story
Nipro Coreless Needle and Needle Set; designed to prevent coring during access to implanted reservoirs. Device consists of luer hub, stainless steel (Type 304) cannula, and protector. Needle sets include soft PVC tubing, roller clamp, and luer connectors; some models feature Y-shaped tubing for mixture injection. Used by clinicians to deliver drug solutions into implanted reservoirs. Prevents coring of the reservoir septum, ensuring integrity of the implanted system.
Clinical Evidence
No clinical data provided; substantial equivalence is based on identical design and technical characteristics to the predicate device.
Indicated for the injection of drug solutions into an implanted reservoir in the body.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
Submission Summary (Full Text)
{0}------------------------------------------------
K971582
Image /page/0/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized, abstract symbol to the left of the word "NIPRO" in bold, sans-serif font. The symbol appears to be two interlocking shapes, possibly representing a stylized "N".
DEC 22 1997
CORPORATION 10877 N.W. 33rd Street Miami. Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
## SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO® CORELESS NEEDLE AND CORELESS NEEDLE SET
\$807.92 (a)(1)
Contact Person:
Luis Candelario General Manager
Address:
Nipro Medical Corporation 10877 NW 33rd Street Miami, Florida 33172 (305) 599-7174 (telephone) (305) 599-8454 (fax)
Date of Summary Preparation:
April 30, 1997
\$807.92 (a)(2)
Nipro® Coreless Needle and Nipro Coreless Needle Set Trade Name:
Hypodermic single lumen needle or I.V. Administration Set Common Name:
Classification Name: Hypodermic single lumen needle (21 CFR §880.5570) or Intravascular administration set (21 CFR §880.5440)
\$807.92 (a)(3)
Legally Marketed Substantially Equivalent Device: Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
8807.92 (a)(4)
Description of Device: Nipro Coreless Needles and Nipro Coreless Needle Sets are described in Hypodermic single lumen needle (21 CFR 8880.5570) or Intravascular administration set (21 CFR \$880.5440). The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
{1}------------------------------------------------
The Nipro Coreless Needles and Nipro Coreless Needle Sets that we intend to market include 2 different types: the straight type (7 sizes) and the 90° angle type (7 sizes). The Nipro Coreless Needle consists of a luer hub, a cannula, and a protector over the cannula. The needle base and the needle protector are composed of polypropylene (PP) and the cannula is made of stainless steel (Type 304).
The Nipro Coreless Needle Sets consist of a needle connected to a length (100 mm or 450 mm) of soft polyvinylchloride (PVC) tubing with a clamp roller along it and a luer connector and luer cap at the other end. Models in which there is a Yshaped tube for mixture injection include a second length of PVC tubing connected at the first luer connector with a clamp roller along it which is then connected to another luer connector and cap.
The materials used for the components include stainless steel (Type 304), polyethylene (PE), polyvinylchloride (PVC), polypropylene (PP), polycarbonate (PC), natural rubber (NR), and polyoxymethylene (POM).
\$807.92 (a)(5)
Intended Use:
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
§807.92 (a)(6)
Comparison of Technical Characteristics:
The Nipro Medical Corporation and Exel International, Inc. devices are exactly the same devices, therefore, all technical characteristics are identical.
s
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 1997
Nipro Medical Corporation C/O Kaelyn B. Hadley Consultant 1384 Copperfield Court Lexington, Kentucky 40514-1268
Re: K971582 Trade Name: NIPRO® Huber Needles and Huber Needle Sets Requlatory Class: II Product Code: LJT Dated: October 30, 1997 Received: November 3, 1997
Dear Ms. Hadley:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does
{3}------------------------------------------------
Page 2 - Ms. Hadley
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours
Timothy A. Ulatowski
Directo Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
K971582
Image /page/4/Picture/1 description: The image shows the logo for Nipro. The logo consists of a stylized, abstract symbol to the left of the word "NIPRO" in bold, sans-serif font. The symbol appears to be two interlocking shapes, possibly representing a stylized "N".
DEC 22 1997
CORPORATION 10877 N.W. 33rd Street Miami. Florida 33172 Tel .: (305) 599-7174 Fax: (305) 599-8454
## SUMMARY OF SAFETY AND EFFECTIVENESS FOR NIPRO® CORELESS NEEDLE AND CORELESS NEEDLE SET
\$807.92 (a)(1)
Contact Person:
Luis Candelario General Manager
Address:
Nipro Medical Corporation 10877 NW 33rd Street Miami, Florida 33172 (305) 599-7174 (telephone) (305) 599-8454 (fax)
Date of Summary Preparation:
April 30, 1997
\$807.92 (a)(2)
Nipro® Coreless Needle and Nipro Coreless Needle Set Trade Name:
Hypodermic single lumen needle or I.V. Administration Set Common Name:
Classification Name: Hypodermic single lumen needle (21 CFR §880.5570) or Intravascular administration set (21 CFR §880.5440)
\$807.92 (a)(3)
Legally Marketed Substantially Equivalent Device: Huber Needle and Huber Needle Set, Exel International, Inc. (K895769 and K895770)
8807.92 (a)(4)
Description of Device: Nipro Coreless Needles and Nipro Coreless Needle Sets are described in Hypodermic single lumen needle (21 CFR 8880.5570) or Intravascular administration set (21 CFR \$880.5440). The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
{5}------------------------------------------------
The Nipro Coreless Needles and Nipro Coreless Needle Sets that we intend to market include 2 different types: the straight type (7 sizes) and the 90° angle type (7 sizes). The Nipro Coreless Needle consists of a luer hub, a cannula, and a protector over the cannula. The needle base and the needle protector are composed of polypropylene (PP) and the cannula is made of stainless steel (Type 304).
The Nipro Coreless Needle Sets consist of a needle connected to a length (100 mm or 450 mm) of soft polyvinylchloride (PVC) tubing with a clamp roller along it and a luer connector and luer cap at the other end. Models in which there is a Yshaped tube for mixture injection include a second length of PVC tubing connected at the first luer connector with a clamp roller along it which is then connected to another luer connector and cap.
The materials used for the components include stainless steel (Type 304), polyethylene (PE), polyvinylchloride (PVC), polypropylene (PP), polycarbonate (PC), natural rubber (NR), and polyoxymethylene (POM).
\$807.92 (a)(5)
Intended Use:
The Nipro Coreless Needle and Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
§807.92 (a)(6)
Comparison of Technical Characteristics:
The Nipro Medical Corporation and Exel International, Inc. devices are exactly the same devices, therefore, all technical characteristics are identical.
s
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in black and have a flowing, wave-like quality. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 1997
Nipro Medical Corporation C/O Kaelyn B. Hadley Consultant 1384 Copperfield Court Lexington, Kentucky 40514-1268
Re: K971582 Trade Name: NIPRO® Huber Needles and Huber Needle Sets Requlatory Class: II Product Code: LJT Dated: October 30, 1997 Received: November 3, 1997
Dear Ms. Hadley:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does
{7}------------------------------------------------
Page 2 - Ms. Hadley
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours
Timothy A. Ulatowski
Directo Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known) Device Name: Nipro Coreless Needle and Nipro Coreless Needle Set
Indications for Use: The Nipro Coreless Needle and the Nipro Coreless Needle Set, for which coring is prevented, are intended to be used to inject a drug solution into a reservoir implanted in the body.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ... 1
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|-----------------------------------------------------------------------------------------|----|----------------------|
| (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices | | |
| 516(k) Number Use | OR | Over-the-Counter Use |
| (Per 21 CFR 801.109) | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.