N-GENEOUS LDL CHOLESTEROL REAGENT, N-GENEOUS LDL CHOLESTEROL CALIBRATOR, GENZYME LDL CHOLESTEROL CONTROL SET
Applicant
Genzyme Corp.
Product Code
LBR · Clinical Chemistry
Decision Date
Jun 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1475
Device Class
Class 1
Indications for Use
Reagents: For the direct, quantitative determination of low-density lipoprotein cholesterol (LDL-C) in human serum or plasma. Calibrators: For the calibration of the N-geneous™ LDL Cholesterol assay in serum or plasma. Controls: To monitor the performance of Genzyme LDL Cholesterol Reagents.
Device Story
Device is a two-reagent homogeneous assay for direct quantitative determination of LDL-C in human serum or plasma. Principle: first detergent selectively solubilizes non-LDL lipoproteins, removed by cholesterol enzymes; LDL particles remain intact. Hydrogen peroxide from enzyme reaction with non-LDL cholesterol consumed by peroxidase/4-aminoantipyrine to yield colorless product. Second detergent releases LDL cholesterol; enzyme reaction with coupler produces color proportional to LDL-C concentration. Used in clinical laboratory settings by trained personnel. Output is colorimetric measurement used by clinicians to assess lipid profiles and manage patient cardiovascular health.
Clinical Evidence
Bench testing only. Comparative study of 92 serum samples (LDL 68.1-214.5 mg/dL) against Direct LDL and β-Quantification methods. Correlation coefficients (r) were 0.97 and 0.96 respectively. Mean percent differences were -2.0% and -1.9%. Precision studies at three NCEP target levels (<130, 130-159, ≥160 mg/dL) showed within-run CVs of 0.62%-0.73% and between-run CVs of 1.73%-2.27%.
Technological Characteristics
Two-reagent homogeneous enzymatic colorimetric assay. Utilizes selective detergents for lipoprotein solubilization and peroxidase-based colorimetric detection. Designed for use on clinical chemistry analyzers. No specific materials of construction or connectivity standards provided.
Indications for Use
Indicated for the quantitative determination of LDL-C in human serum or plasma for clinical laboratory use.
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
Reference Devices
Direct LDL Cholesterol Immunoseparation Reagent
β-Quantification
Submission Summary (Full Text)
{1}------------------------------------------------
K911573
510(k) PREMARKET NOTIFICATION
Genzyme Corporation One Kendall Square Cambridge, MA 02139
June 18, 1997
Reagent and C
### 510(k) Summary Of Safety and Effectiveness Information Upon Which An Equivalence Determination Could be Made
| Trade or Proprietary Name: | Genzyme N-geneous™ LDL Cholesterol Reagent<br>Genzyme N-geneous™ LDL Cholesterol Calibrator<br>Genzyme LDL Control Set |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Homogeneous assay for low density lipoprotein cholesterol |
| Classification Name: | Low density lipoprotein cholesterol test<br>Calibrator, Primary<br>Low density lipoprotein control |
| Manufacturer: | Genzyme Diagnostics<br>One Kendall Square<br>Cambridge, MA 02139-1562 |
| Contact Person: | Nancy E. Isaac, Associate Director, Regulatory Affairs (617) 374-7431 or<br>Beth A. Crowley, Regulatory Affairs Associate (617) 252-7669 |
The use of the Genzyme N-geneous™ LDL Cholesterol Reagents in the clinical laboratory setting is substantially equivalent to the Direct LDL Cholesterol Immunoseparation Reagent and the B-Quantification methods.
The Genzyme N-geneous™ LDL Cholesterol Reagent is a two-reagent homogeneous method for the direct quantitative determination of low density lipoprotein cholesterol (LDL-C) in human serum and plasma.
The principle of the test is based upon a unique detergent which selectively solubilizes only the non-L.DL lipoproteins, allowing them to be removed by cholesterol enzymes prior to the LDL cholesterol reaction. The LDL particles remain intact. The hydrogen peroxide produced by the reaction of the enzymes with the released cholesterol is consumed by a peroxidase reaction with 4-aminoantipyrine. yielding a colorless product. A second detergent, capable of releasing LDL cholesterol molecules, is added. The enzyme reaction with LDL cholesterol in the presence of the coupler produces color which is proportional to the amount of LDL cholesterol in the sample.
Comparative performance studies were conducted using the N-geneous™ LDL Cholesterol Reagents and two reference methods: Direct LDL Cholesterol Immunoseparation Reagent and 3-Quantification.
ﺎ ﻓﻲ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ
{2}------------------------------------------------
K971573
### 510(k) PREMARKET NOTIFICATION
Ninety-two serum samples, with LDL values between 68.1 and 214.5 mg/dL, were tested at Genzyme Corporation using the N-geneous™ LDL and the Direct LDL methods. Of these 92 samples, 54 were tested with the ß-Quantification reference method.
| N-geneous™ LDL | vs.<br>β-Quantification<br>(n = 54) | vs.<br>Direct LDL<br>(n = 92) |
|-----------------------------|-------------------------------------|-------------------------------|
| Slope | 0.96 | 0.94 |
| Intercept | 3.02 | 4.46 |
| Correlation Coefficient (r) | 0.96 | 0.97 |
| Mean (mg/dL) | Ng: 122.5<br>βQ: 125.1 | Ng: 120.0<br>Dir: 122.8 |
| Standard Deviation (mg/dL) | Ng: 30.7<br>βQ: 30.9 | Ng: 30.5<br>Dir: 31.6 |
| Mean Difference | -2.62 | -2.81 |
| Mean Percent Difference | -1.9% | -2.0% |
Precision studies were conducted using the N-geneous™ LDL Cholesterol Reagents. Both within-run and between-run studies were performed using frozen serum pools at three target levels of LDL cholesterol as determined by the National Cholesterol Education Program (NCEP): <130 mg/dL (low); 130-159 mg/dL (mid); and ≥160 mg/dL (high).
| Within-Run | Low<br>(<130 mg/dL) | Mid<br>(130-159 mg/dL) | High<br>(≥160 mg/dL) |
|----------------------|---------------------|------------------------|----------------------|
| n | 20 | 20 | 20 |
| Sample Range (mg/dL) | 97.1 - 99.6 | 144.9 - 148.7 | 207.0 - 211.5 |
| Mean (mg/dL) | 98.1 | 146.5 | 209.8 |
| SD (mg/dL) | 0.72 | 0.96 | 1.31 |
| %CV | 0.73% | 0.66% | 0.62% |
| Between-Run | Low<br>(<130 mg/dL) | Mid<br>(130-159 mg/dL) | High<br>(≥160 mg/dL) |
|----------------------|---------------------|------------------------|----------------------|
| n | 40 | 40 | 40 |
| Sample Range (mg/dL) | 93.0 - 101.9 | 137.5 - 148.6 | 199.3 - 213.9 |
| Mean (mg/dL) | 98.1 | 142.7 | 207.3 |
| SD (mg/dL) | 2.2 | 2.8 | 3.6 |
| %CV | 2.27% | 1.95% | 1.73% |
These data demonstrate that the performance of the N-geneous™ LDL Cholesterol Reagents in the clinical laboratory is substantially equivalent to the performance of the Direct LDL Cholesterol Immunoseparation Reagent and the ß-Quantification methods.
In lieu of a 510(k) statement under 513(i) of the Act, this information is provided as a 510(k) summary for disclosure to any other persons/companies without the specific written authorization from Genzyme Corporation.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUN 18 1997
Nancy E. Isaac Associate Director, Requlatory Affairs Genzyme Corporation One Kendall Square Cambridge, Massachusetts 02139
駅e: K971573 N-geneous™ LDL Cholesterol Reagent/Cholesterol Calibrator Genzyme LDL Cholesterol Control Set Regulatory Class: I & II Product Code: LBR, JIS, JIS, JIX Dated: April 29, 1997 Received: April 30, 1997
Dear Ms. Isaac:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle or bird design, with three curved lines forming the body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## JUN 18 1997
Nancy E. Isaac Associate Director, Regulatory Affairs Genzyme Corporation -------One Kendall Square Cambridge, Massachusetts 02139
K971573 駅e: N-geneous™ LDL Cholesterol Reagent/Cholesterol Calibrator Genzyme LDL Cholesterol Control Set Regulatory Class: I & II Product Code: LBR, JIS, JIS, JIX Dated: April 29, 1997 April 30, 1997 Received:
Dear Ms. Isaac:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in Please note: this response to your the Federal Register. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
તેમ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page 1 of 1
Genzyme Corporation One Kendall Square Cambridge, MA 02139-1562 252-7500 Telephone (617)
510(k) Number (if known):
N-geneous™ LDL Cholesterol Reagent Device Name: N-geneous™ LDL Cholesterol Calibrator Genzyme LDL Cholesterol Control Set 参
Indications For Use:
Reagents:
For the direct, quantitative determination of low-density lipoprotein cholesterol (LDL-C) in human serum or plasma.
### Calibrators:
For the calibration of the N-geneous™ LDL Cholesterol assay in serum or plasma.
Controls:
To monitor the performance of Genzyme LDL Cholesterol Reagents.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cinical Laboratory Devices
510(k) Nurnber k 971537
న Prescription Use (Per 21 CFR 801.109)
ರ್ಯ
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Sk-19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.