KINEMATIC ALL-POLYETHYLENE PATELLA WITH THREE PEGS
K971550 · Howmedica Corp. · JWH · Jul 8, 1997 · Orthopedic
Device Facts
| Record ID | K971550 |
| Device Name | KINEMATIC ALL-POLYETHYLENE PATELLA WITH THREE PEGS |
| Applicant | Howmedica Corp. |
| Product Code | JWH · Orthopedic |
| Decision Date | Jul 8, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3560 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Kinematio® All-Polyethylene Patella with Three Pegs is intended to be used with the femoral and tibial components of the Kinematic® and Kinematic® II total knee systems, and the distal femoral component of the Howmedica Modular Replacement System in primary and/or revision total knee arthroplasty. This component is intended to resurface the articular surface of the patella. It is intended to be implanted with bone cement.
Device Story
Kinematic® All-Polyethylene Patella with Three Pegs is a prosthetic component for total knee arthroplasty; designed to resurface patellar articular surface. Device features rounded, dome-shaped articular surface and undersurface with triangular cement recess and three straight pegs for fixation. Implanted by orthopedic surgeons in clinical/OR settings using bone cement. Articular surface mates with intercondylar recess of Kinematic®, Kinematic® II, and Howmedica Modular Replacement System femoral components. Device replaces existing Kinematic® Patella (central keel design) to standardize instrumentation across Howmedica knee systems. Benefits include improved surgical compatibility and restoration of patellofemoral joint function.
Clinical Evidence
No clinical data. Evidence consists of bench testing, specifically patello-femoral contact area and lateral subluxation testing.
Technological Characteristics
Material: Ultra-High Molecular Weight Polyethylene (UHMWPE) conforming to ASTM F-648. Design: Three-peg undersurface with triangular cement recess; rounded dome articular surface. Sterilization: Gamma radiation (room air or inert environment with heat treatment).
Indications for Use
Indicated for patients undergoing primary or revision total knee arthroplasty requiring patellar resurfacing. Must be used with bone cement. Contraindicated for non-cemented fixation.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
- Kinematic® All Poly Patella
- Duracon® All Polyethylene Patella II
Related Devices
- K972967 — OSTEONICS SCORPIO TOTAL KNEE UNIVERSAL DOME PATELLAR COMPONENT;OSTEONICS SCORPIO TOTAL KNEE RECESSED PATELLAR COMPONENT · Osteonics Corp. · Jan 16, 1998
- K972643 — DEPUY AMK 3 PEG PATELLA · Depuy, Inc. · Jan 21, 1998
- K061340 — 23 MM SINGLE-PEG PATELLA COMPONENT · Biomet Manufacturing, Inc. · Sep 25, 2006
- K240299 — Persona the Personalized Knee System (Persona OsseoTi 3-Peg Patella) · Zimmer, Inc. · Apr 26, 2024
- K092803 — CARDO MEDICAL ALIGN 360 MEDIALIZED PATELLA COMPONENT · Cardo Medical Corp. · Dec 23, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
JUL - 8 1997
## Kinematic® All-Polyethylene Patella with Three Pegs Device:
The Kinematio® All-Polyethylene Patella with Three Pegs is intended to be used with the femoral and tibial components of the Kinematic® and Kinematic® II total knee systems, and the distal femoral component of the Howmedica Modular Replacement System in primary and/or revision total knee arthroplasty. This component is intended to resurface the articular surface of the patella. It is intended to be implanted with bone cement.
The Kinematic® All-Polyethylene Patella withThree Pegs is available in three sizes: small, medium, and large. This component has a rounded, dome articular surface that is intended to mate with the intercondylar recess of the above-referenced femoral components. The articular surface of the subject device is identical to the articular surface of the previously cleared Kinematio® Patella.
The undersurface design of the Kinematic® All Polyethylene Patella with Three Pegs has a triangular cement recess with three straight pegs. This undersurface design is identical to the Duracon® All Polyethylene Patella II undersurface. The existing Kinematio® Patella has a central keel with fixation holes. This undersurface is being modified to three pegs to allow common instrumentation to be used throughout Howmedica's total knee systems.
The Kinematic® All-Polyethylene Patella with Three Pegs is fabricated from Ultra-High Molecular Weight Polyethylene, which conforms to ASTM specification F-648. The Kinematio® All Polyethylene Patella with Three Pegs will be made available in two versions: one will be sterilized via a room air gamma sterilization method, and the second version will be sterilized via a gamma sterilization method in an inert environment followed by heat treatment.
The Kinematic® All Polyethylene Patella with Three Pegs is substantially equivalent to other legally marketed Howmedica devices: 1) Kinematic® All Poly Patella and 2) Duracon® All Polyethylene Patella II. This substantial equivalence determination is based on similarities in design, materials, intended use and operational principles.
Patello-femoral contact area and lateral subluxation testing of the patella were presented.
Margaret F. Crowe For information contact:
Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 (201) 507-7431
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The bird is oriented towards the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Margaret F. Crowe Manager, Requlatory Affairs Howmedica Inc. ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… Pfizer Hospital Products Group 359 Veterans Boulevard Rutherford, New Jersey 07070-2584
JUL - 8 1997
- K971550 Re : Kinematic All-Polyethylene Patella with 3 Pegs Requlatory Class: II Product Code: JWH Dated: April 25, 1997 Received: April 28, 1997
Dear Ms. Crowe:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone cement." You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations :
- This device may not be labeled or promoted for 1. non-cemented use.
- All labeling for this device, including package label and 2. labeling included within the package, must prominently state that the device is intended for cemented use only.
{2}------------------------------------------------
Page 2 - Ms. Margaret F. Crowe
- Any non-cemented fixation of this device is considered 3. investigational and may only be investiqated as a siqnificant risk device in accordance with the investigational device exemption (IDE) requlation under 21 CFR, Part 812. All users of the device for non-cemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investiqation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations .
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
{3}------------------------------------------------
Page 3 - Ms. Margaret F. Crowe
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K971550
Kinematic® All-Polyethylene Patella with Three Pegs Device Name:
Indications for Use:
The Kinematio® All-Polyethylene Patella with Three Pegs is intended to be used with the femoral and tibial components of the Kinematic® and Kinematic® II total knee systems,
and the distal femoral component of the Howmedica Modular Replacement System in primary and/or revision total knee arthroplasty. This component is intended to resurface the articular surface of the patella. It is intended to be implanted with bone cement.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109 Over-The-Counter Use
(Optional Format 1-2-96)
Format 1-2-96)
471556