K971542 · Rehabilicare, Inc. · IPF · Aug 14, 1997 · Physical Medicine
Device Facts
Record ID
K971542
Device Name
ORTHO DX NEUROMUSCULAR STIMULATOR
Applicant
Rehabilicare, Inc.
Product Code
IPF · Physical Medicine
Decision Date
Aug 14, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ortho Dx, a non-invasive electrical stimulator that combines symmetrical biphasic neuromuscular stimulation with high volt pulsed direct current neuromuscular stimulation to be used under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions. It is used to, relax muscle spasm, prevent or retard disuse atrophy, increase local blood circulation, reeducate muscles, prevent venous thrombosis with immediate post surgical stimulation in calf muscles, and increase range of motion.
Device Story
Ortho Dx is a non-invasive, prescription-only neuromuscular stimulator; operates under medical supervision. Device combines symmetrical biphasic neuromuscular stimulation and high volt pulsed direct current; modes function simultaneously or independently. System delivers electrical pulses to patient musculature; intended to relax spasms, retard atrophy, improve circulation, reeducate muscles, prevent venous thrombosis, and increase range of motion. Clinician adjusts parameters including amplitude, pulse width, pulse rate, treatment time, and ramp times. Output affects muscle contraction and physiological response to aid rehabilitation. Device utilizes UL-approved power supply.
Clinical Evidence
Bench testing only. Verification of output waveforms, amplitude, pulse width, pulse rate, treatment time, and ramp times performed at room temperature (72°C) and 50% relative humidity using calibrated measurement equipment. Results confirmed parameters are within or identical to predicate device specifications.
Technological Characteristics
Non-invasive electrical stimulator; symmetrical biphasic and high volt pulsed direct current output modes. Parameters: amplitude, pulse width, pulse rate, treatment time, ramp up/down times. Powered by UL-approved power supply. Operates at room temperature (72°C).
Indications for Use
Indicated for patients undergoing rehabilitation for medical diseases and conditions requiring adjunctive electrical stimulation therapy to relax muscle spasms, prevent/retard disuse atrophy, increase local blood circulation, reeducate muscles, prevent venous thrombosis (post-surgical calf stimulation), and increase range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
Predicate Devices
GV II Neuromuscular Stimulator, Model 9700/7000S (K881082)
TRI-STIM Stimulator, Muscle, Electrical Powered (Registration No. not known)
Submission Summary (Full Text)
{1}------------------------------------------------
# 510(k) Summary
# Ortho D. Neuromuscular Stimulator
K971542
#### Model 6700S
#### Product:
Ortho Dx Neuromuscular Stimulator, Model 6700S
# AUG / 4 097
## Classification:
Class II
#### Manufacture:
Rehabilicare Inc.
· 1811 Old Highway 8 New Brighton, MN 55112 Phone No. (612) 631-0590
## Contact:
Gary L. Moore Phone No. (612) 638-0437
## Predicate Products:
- GV II ™ Neuromuscular Stimulator, Model 9700/7000S, K881082 -
- Neuromuscular II (NMIII TM), Model 6800S, K840346 -
- TRI-STIM Stimulator, Muscle, Electrical Powered, Registration No. not known -
## Indications For Use:
The Ortho Dx, a non-invasive electrical stimulator that combines symmetrical biphasic neuromuscular stimulation with high volt pulsed direct current neuromuscular stimulation to be used under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions. It is used to, relax muscle spasm, prevent or retard disuse atrophy, increase local blood circulation, reeducate muscles, prevent venous thrombosis with immediate post surgical stimulation in calf muscles, and increase range of motion.
{2}------------------------------------------------
#### Device Description:
The Ortho De is a non-invasive electrical stimulator that combines symmetrical biphasic neuromuscular stimulation with high volt pulsed direct current neuromuscular stimulation that will operate in both modes simultaneously or independently. This system is a prescription device intended for use on patients who are undergoing rehabilitation for medical diseases and conditions on the order of a licensed practitioner.
#### Summary of Testing:
The output wave form, amplitude, pulse width, pulse rate, treatment time, NMS off time, NMS ramp up time and NMS ramp down time have been verified to the technical requirements listed in the operators manual. All measurements were taken at room temperature (72°C) with a relative humidity of 50% using a calibrated 5V power supply with a maximum capacity of 3 ampere continuous current and calibrated measurement equipment. These values are within and/or identical to the operating parameters of the predicate devices. The Ortho Dx neuromuscular stimulator uses the same power supply and proposes no additional safety considerations than the legally marketed predicate devices (NM II/III and GV II). The Line Powered Repac10 employees a UL approved power supply and is legally marketed with the predicate devices (NM II/II and GV II).
#### Substantial Equivalence Summary:
The Ortho Dx ™ is a neuromuscular stimulator with a symmetric biphasic output wave form identical to the Mentor Neuromuscular II (NM III ™ ), Model 6800S, Neuromuscular Stimulator, Registration Number, K840346, and a high volt pulsed current output wave form identical to the GV II M Model 9700, High Voltage Pulsed Galvanic Stimulator, Registration Number, K881082. This type of unit, combining both types of output wave forms, is similar to a predicate device, the TRI-STIM ™ , Stimulator, Muscle, Electrical Powered, manufactured by Henley, a Division of Maxxim Medical, 104 Industrial Blvd., Sugarland, TX 77478 (registration number unknown).
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Gary L. Moore Director of Engineering Rehabilicare Inc. 1811 Old Hiqhway 8 New Brighton, Minnesota 55112
AUG / 4 1997
Re : K971542 Ortho Dx™ Neuromuscular Stimulator Regulatory Class: II Product Code: IPF Dated: July 29, 1997 Received: August 4, 1997
Dear Mr. Moore:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
# Page 2 - Mr. Gary L. Moore
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA described in your 510(K) prematice notification: - tho legally Finding of Bubboandar organisation a classification for your markets production permits your device to proceed to the market.
If you desire specific advice for your device on our labeling i regulation (21 CFR Part 801 and additionally 809.10 for in regaradiagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact che office of Compliance at (301) 594-4639. Also, please note the Office or compriation as (compranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
ia M. Witten, Ph.D., M.D. Cel Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{5}------------------------------------------------
### Statement for Indication of Use:
The Ortho Dx, a non-invasive electrical stimulator that combines symmetrical biphasic neuromuscular stimulation with high volt pulsed direct current neuromuscular stimulation to be used under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions. It is used - to, relax muscle spasm, prevent or retard disuse … atrophy, increase local blood circulation, reeducate muscles, prevent venous thrombosis with immediate post surgical stimulation in calf muscles, and increase range of motion.
Sign-Off
51.JK, Number
Prescription Use
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.