K971518 · KARL STORZ Endoscopy-America, Inc. · HIH · Jul 24, 1997 · Obstetrics/Gynecology
Device Facts
Record ID
K971518
Device Name
KSEA SEMI-RIGID HYSTEROSCOPE AND ACCESSORIES
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jul 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KSEA Semi-Rigid Hysteroscope and Accessories are designed to view the uterus, and, using additional accessories, to perform various diagnostic and therapeutic procedures. The Alken Motion Control Device attaches to an instrument channel to precisely guide the introduction of small instrument into the patient.
Device Story
KSEA Semi-Rigid Hysteroscope is a straight, graduated shaft endoscope with a remote eyepiece; used for uterine visualization. Alken Motion Control Device attaches to the instrument channel to guide accessory introduction and movement. Both are manually operated, reusable surgical instruments. Used by physicians in clinical settings for diagnostic and therapeutic gynecological procedures. Provides direct visualization of the uterine cavity to assist in biopsy, septa/adhesion transection, and foreign body removal. Benefits include precise instrument guidance and minimally invasive access to the uterus.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: stainless steel (chromium-plated Monel 400). Design: straight, graduated shaft endoscope with remote eyepiece. Operation: manual. Reusable. No software or electronic components.
Indications for Use
Indicated for diagnostic procedures including infertility, pregnancy wastage, abnormal uterine bleeding, intrauterine foreign body, pelvic pain, IUD monitoring, and evaluation of abnormal hysterosalpingogram or endometrial abnormalities. Also indicated for operative procedures including directed biopsy, transection of intrauterine septa, and transection of intrauterine adhesions. Absolute contraindication: acute PID. Relative contraindications: inability to distend uterus, cervical/vaginal infection, known pregnancy, uterine bleeding/menses, invasive cervical carcinoma, recent uterine perforation, medical contraindication, anesthesia intolerance, cervical stenosis.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Karl Storz Endoscopy. The logo is white text on a black background. The top line of the logo says "STORZ" in large, bold letters. The second line of the logo says "Karl Storz Endoscopy" in smaller letters.
# K97 1578
## JUL 2 4 1997
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document are accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe<br>Culver City, CA 90230<br>(310) 558-1500 |
|------------------------|---------------------------------------------------------------------------------------------------------|
| Contact: | Marlena Allen Piercy, Ph.D.<br>Senior Clinical Affairs Specialist |
| Device Identification: | Common Name<br>Semi-Rigid Hysteroscope and Accessories<br>Trade Name |
KSEA Semi-Rigid Hysteroscope KSEA Alken Motion Control Device
Indication: The KSEA Semi-Rigid Hysteroscope and Accessories are designed to view the uterus, and, using additional accessories, to perform various diagnostic and therapeutic procedures. The Alken Motion Control Device attaches to an instrument channel to precisely guide the introduction of small instrument into the patient.
Device Description: The KSEA Semi-Rigid Hysteroscope is a straight, graduated shaft endoscope with a remote eyepiece. The Alken Motion Control Device is an attachment to the instrument channel to control the introduction and movement of accessories. All devices are manually operated, reusable surgical devices. Body contact materials are stainless steel (chromium-plated Monel 400®).
Substantial Equivalence: The KSEA Semi-Rigid Hysteroscope and Accessories are substantially equivalent to the predicate devices since the basic features, design, and intended uses are the same or similar. The minor differences in dimensions between the KSEA Semi-Rigid Hysteroscope and Accessories and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function, or intended use of these devices,
Signed: Mhelina Allen Piercey
Mariana Allen Piercey, Ph.D.
Marlena Allen Piercy, Ph.D. Senior Clinical Affairs Specialist
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are written around the eagle in a circular pattern. The eagle is a symbol of the United States, and the seal represents the department's mission to protect the health of all Americans.
Semi-Rigid Hysteroscope and Alken Motion
21 CFR §884.1690/Product code: 85 HIH
Control Device . ..........................
Dated: April 23, 1997
Regulatory class: II
Received: April 25, 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Marlena Allen Piercy, Ph.D. Senior Clinical Affairs Specialist Karl Storz Endoscopy-America, Inc. . 600 Corporate Pointe Culver City, California 90230-7600
JUL 2 4 1997
Dear Dr. Piercy:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the intenstated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Re: K971518
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 80) and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
h.7liau Yu
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Kan Storz Endoscopy-America, Inc.
600 Corporate Pointe Culver City, California 90230-7600 Phone 310 558 1500
Toll Free 800 421 0837 Fax 310 410 5527
### 510(k) Number (if known): Not yet assigned
### Device Name: Semi-Rigid Hysteroscope with Accessories
Indications for Use: The Karl Storz Semi-Rigid Hysteroscope and Accessories are indicated in the following conditions:
- Diagnostic
infertility and pregnancy wastage abnormal uterine bleeding intrauterine foreign body pelvic pain monitoring of IUD status evaluation of abnormal hysterosalpingogram evaluation of abnormalities of the endometrium
### Operative
directed biopsy transection of intrauterine septa transection of intrauterine adhesions
Absolute Contraindications for Use: acute PID
#### Relative Contraindications for Use:
inability to distend uterus cervical/vaginal infection known pregnancy uterine bleeding or menses invasive carcinoma of the cervix recent uterine perforation medical contraindication intolerance to anesthesia cervical stenosis
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--------------------------------------------------------------------------------|
| Prescription Use: | <span style="text-decoration: overline;">✓</span> |
| (Per 21 CFR 801.109) | OR Over-The-Counter Use: |
| | |
| | (Optional Format 1-2-96) |
| | <span style="text-decoration: overline;">Rolera satting</span> |
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal, ENT, and Radiological Devices |
| 510(k) Number | <span style="text-decoration: overline;">K971518</span> |
| | 000003 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.