HUGHES BLADDER AND BOWEL TRAINING SYSTEM(AD-T-18, MED-18,MED-1,MED-T-1,MEDD-18,MED-38,MED-T-10,MED-T-25,MED-T-5

K971442 · Humed, Inc. · KPN · Jan 8, 1998 · Gastroenterology, Urology

Device Facts

Record IDK971442
Device NameHUGHES BLADDER AND BOWEL TRAINING SYSTEM(AD-T-18, MED-18,MED-1,MED-T-1,MEDD-18,MED-38,MED-T-10,MED-T-25,MED-T-5
ApplicantHumed, Inc.
Product CodeKPN · Gastroenterology, Urology
Decision DateJan 8, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.2040
Device ClassClass 2

Indications for Use

The proposed product, Hughes Bowel and Bladder System Model MeD-18, is an enuresis alarm, a device to detect urine and/or feces and to set off a vibratory alarm when detected, that may be used by an ambulatory adult to assist in the control of incontinence. Hughes Bowel and Bladder System model MeD-18 helps persons with incontinence manage or overcome bladder control problems by providing them with a private alarm system to detect when uncontrolled urination or excretion begins. Hughes Bowel and Bladder System, model MeD-18, assists in controlling incontinence so that the wearer can manage their conditon promptly and with complete discretion at the first sign of bladder or bowel release day or night.

Device Story

The Hughes Bowel and Bladder System Model MeD-18 is an enuresis alarm designed for ambulatory adults. The device detects the presence of urine or feces and provides a discreet, vibratory alert to the wearer. It functions via a pager-like motor that activates upon detection, allowing the user to manage incontinence promptly. The device is intended for use by the patient to assist in controlling incontinence during day or night activities.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Enuresis alarm; pager-like form factor; vibratory motor for alert; battery-operated; non-invasive; wearable.

Indications for Use

Indicated for ambulatory adults with incontinence to assist in the management of bladder and bowel control by providing a private alarm upon detection of urine or feces.

Regulatory Classification

Identification

An enuresis alarm is a device intended for use in treatment of bedwetting. Through an electrical trigger mechanism, the device sounds an alarm when a small quantity of urine is detected on a sensing pad. This generic type of device includes conditioned response enuresis alarms.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Submission Summary (Full Text)

{1}------------------------------------------------ 510(k) Sunimary for K971442 JAN - 8 1998 **Sponsor:** HuMed, Inc. Charles Hughes, President 709 N. Memorial Parkway Huntsville, Alabama 35801 K971442 P101 The proposed product, Hughes Bowel and Bladder System Model MeD-18, is an enuresis alarm, a device to detect urine and/or feces and to set off a vibratory alarm when detected, that may be used by an ambulatory adult to assist in the control of incontinence. Hughes Bowel and Bladder System model MeD-18 helps persons with incontinence manage or overcome bladder control problems by providing them with a private alarm system to detect when uncontrolled urination or excretion begins. Hughes Bowel and Bladder System, model MeD-18, assists in controlling incontinence so that the wearer can manage their conditon promptly and with complete discretion at the first sign of bladder or bowel release day or night. The device functions with a pager-like motor that vibrates to inform the wearer of urine / feces spillage. {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black against a white background. JAN - 8 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Charles Hughes President HuMed, Inc. 709 N. Memorial Parkway Huntsville, AL 35801 Re: K971442 Hughes Bowel and Bladder System Model MeD-18 Dated: December 15, 1997 Received: December 18, 1997 Regulatory Class: II 21 CFR 876.2040/Procode: 78 KPN Dear Mr. Hughes: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, h. William Yri Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page_of_ 510(k) Number (if known); K 2 7/1442 Device Name Husbal Reach and Blaker System nell-18 Indications For Use: The proposed product, Hughes Bowel and Bladder System Model MeD-18, is an enuresis alarm, a device to detect urine and/or feces and to set off a vibratory alarm when detected, that may be used by an ambulatory adult to assist in the control of incontinence. Hughes Bowel and Madder System model MeD-18 helps persons with incontinence manage of overcome bladder control problems by providing them with a private alarm system to detect when uncontrolled urination or excretion begins. Hughes Bowel and Bladder System, model MeD-18, assists in controlling incontinence so that the wearer can manage their comditon promptly and with complete discretion at the first sign of bladder or bowel release day or night. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) *Robert R Ratliff* (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number *K971442* Prescription Use***_*****_ OR Over-The-Counter Use_** (Per 21 CFR 801.109) (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...