Control Level Theta Plasma is intended for use in a laboratory program of quality assurance for activated partial thromboplastin time (APTT) and prothrombin time (PT) results with short clotting times. PT and APTT clotting times at the low end, or below the locally determined normal reference range are often related to the presence of elevated levels of fibrinogen and factor VIII. Both of these proteins are elevated in response to an inflammatory stimulus, malignant disease process or trauma. Control Level Theta Plasma provides a means for verifying the lower extent of the reportable range of the PT and APTT, particularly the imprecision of short clotting times often encountered with these screening tests.
Device Story
Control Level Theta is a lyophilized, human citrated plasma preparation containing stabilizers, buffers, and elevated levels of fibrinogen and factor VIII. It is used in clinical laboratories to verify the lower limit of the reportable range for PT and APTT assays. The device is reconstituted with deionized or distilled water and analyzed on standard coagulation analyzers (e.g., MLA-1000C, MLA-1600, MLA-700, ACL-3000, Fibrintimer) alongside patient samples. By producing abnormally short clotting times, it allows clinicians to assess the precision of screening tests in the low-time range. The product is handled as potentially infectious biological material. It provides a quality control mechanism to ensure accurate reporting of PT/APTT results, aiding in the monitoring of patients with inflammatory or malignant conditions.
Clinical Evidence
Bench testing only. Precision studies (within-run and between-run) compared Control Level Theta to Dade Ci-Trol Level I using various reagents and instruments (MLA-1000C, MLA-1600, MLA-700, ACL-3000). Within-run CVs for PT and APTT were consistently ≤ 1.35%. Reconstituted stability was validated for 6 hours at room temperature with < 5% change in clot time. Testing across nine PT and seven APTT reagents confirmed consistently shorter clotting times compared to pooled normal plasma.
Technological Characteristics
Lyophilized human citrated plasma with added stabilizers and buffers. Reconstituted with 1.0 mL deionized/distilled water. Functions as a biological control for coagulation analyzers. Compatible with various end-point detection methods (mechanical, optical). No electronic components or software algorithms; performance is dependent on the chemical composition of the plasma and the user's coagulation instrumentation.
Indications for Use
Indicated for use in clinical laboratories as a quality assurance control for PT and APTT testing, specifically for verifying the lower end of the reportable range where clotting times are abnormally short due to elevated fibrinogen and factor VIII levels associated with inflammation, malignancy, or trauma.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Dade® Ci-Trol® Coagulation Control, Level I
Submission Summary (Full Text)
{0}------------------------------------------------
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
# PREMARKET NOTIFICATION 510(K) SUMMARY
| Submitter: | Laura A. Worfolk, Ph.D.<br>Pacific Hemostasis<br>11515 Vanstory Drive<br>Huntersville, NC 28078<br>(704) 875-0494<br>Fax # (704) 875-6671 |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-------------------------------------------------------------------------------------------------------------------------------------------|
The same as above. Contact Person:
4/16/97 Date:
Coagulation Control Plasma (Abnormal) Level 6 Trade Name:
Not applicable Common Name:
Classification Name: Plasma, Coagulation Control (per 21 CFR section 864.5425)
Dade® Ci-Trol® Coagulation Control, Level I Equivalent Device:
#### Description of Coagulation Control Plasma (Abnormal) Level θ:
Control Level 6 Plasma is a lyophilized preparation of fresh human citrated plasma with added stabilizers and buffers. The reconstitution volume is 1.0 mL (with deionized or distilled water). Control Level Theta contains markedly elevated levels of human fibrinogen and factor VIII, and produces an abnormally short Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) with a variety of end-point detection methods. Each unit of source material used in the preparation of this product has been tested by an FDA approved method and found non-reactive for HBAg (Hepatitis B Surface antigen) and negative for antibodies to HIV and HCV. However, since no known test method can offer complete assurance that product derived from human blood will not transmit
{1}------------------------------------------------
Hepatitis, AIDS, or other infectious diseases, this product should be handled as potentially infectious biological material.
### Intended Use of Coagulation Control Plasma (Abnormal) Level 9:
Control Level 0 Plasma is intended for use in a laboratory program of quality assurance for activated partial thromboplastin time (APTT) and prothrombin time (PT) results with short clotting times. PT and APTT clotting times at the low end, or below the locally determined normal reference range are often related to the presence of elevated levels of fibrinogen and factor VIII. Both of these proteins are elevated in response to an inflammatory stimulus, malignant disease process or trauma. Control Level 6 Plasma provides a means for verifying the lower extent of the reportable range of the PT and APTT, particularly the imprecision of short clotting times often encountered with these screening tests.
## Summary of Data Comparing Control Level 8 Plasma to Dade® Ci-Trol® Level I:
Pacific Hemostasis Control Level Theta was compared to Dade® Ci-Trol® Level I. a pre-amendment device. Both Control Level Theta and Dade® Ci-Trol® Level I plasmas are lyophilized preparations of human citrated plasmas with added stabilizers and buffers. The intended use for both products is indistinguishable: as a control plasma for Prothrombin and Activated Partial Thromboplastin Time testing in a clinical laboratory program of quality assurance. Control Level Theta contains elevated levels of human fibrinogen and Factor VIII. which result in clotting times below the normal reference range. In contrast, Dade® Ci-Trol® Level I yields clotting times within the normal reference range.
Within-run precision studies performed on two different instruments (MLA®-1000C™ & MLA®-1600) using both Dade and Pacific Hemostasis brand PT and APTT reagents yielded comparable data. The coefficient of variation (CV) for Control Level Theta and Dade® Ci-Trol® I was well below the
{2}------------------------------------------------
performance specifications. (For Pacific Hemostasis Control Level Theta the CV for both PT and APTT must be ≤ 5%. The stated CV for Dade® Ci-Trol® I is ≤ 5% and < 9% for PT and APTT, respectively.) Testing performed using Pacific Hemostasis brand PT and APTT reagents yielded a 1.23% (PT) and 0.74% (APTT) CV with Control Level Theta. The CV's for Dade® Ci-Trol® Level I using Pacific Hemostasis brand reagents was 0.82% (PT) and 0.79% (APTT). Testing was also performed using Dade brand PT and APTT reagents. For Control Level Theta, a 1.35% and 0.55% CV was obtained for PT and APTT testing respectively. Dade Ci-Trol Level I yielded a 0.74% and 0.48% CV for PT and APTT testing respectively. The following table summarizes this data:
| Table 14. Summary of Within-Run Precision Studies | | | | | |
|---------------------------------------------------|---------------|---------------|---------------|---------------|--|
| | PT | | APTT | | |
| | Theta Control | Dade Ci-Trol® | Theta Control | Dade Ci-Trol® | |
| Dade Reagents | 1.35 | 0.74 | 0.55 | 0.48 | |
| PH Reagents | 1.23 | 0.82 | 0.74 | 0.79 | |
Results are %CV.
Since the results for the between-run precision studies were indistinguishable between the two manufacturer's reagents, day-to-day precision studies were performed using Pacific Hemostasis brand reagents only. Testing was performed on the MLA®-1000C™. Control Level Theta had a 1.46% and 1.29% CV for PT and APTT testing respectively; Dade® Ci-Trol® Level I yielded a 1.87% and 1.41% CV for PT and APTT testing.
The reconstituted stability claim for Control Level Theta is for 6 hours at room temperature. (The manufacturing specification for Control Level Theta is for less than a 5% change in clot time over a 6 hour storage period.) Stability studies support this claim; there was a 1.39% change for PT and 2.43% change for APTT over a 6-hour time period. Dade Ci-Trol® Level I was also tested at 6
{3}------------------------------------------------
hours and yielded comparable data; 1.37% and -0.19% for PT and APTT respectively. (These studies were also run on the MLA®-1000C™.)
Control Level Theta yields an abnormally short PT and APTT (when compared to normal plasma) with a wide variety of end-point detection methods. Control Level Theta and Pooled Normal Plasma were tested for the PT and APTT using nine PT reagents and seven APTT reagents. Testing for both the PT and APTT was performed on four different coagulation analyzers. The instruments used for this analysis were the manual LAbor Fibrintimer®, the semiautomated MLA®-700, and the fully automated ACL-3000™6 and the MLA -1000C™. This combination of analyzers represents approximately 80% of the instruments currently in use (within the USA). All reagent-instrument testing combinations yielded shorter PT and APTT clot times for Control Level Theta when compared to Pooled Normal Plasma. This testing was not performed for Dade® Ci-Trol® Level I since it is intended primarily for use with Dade brand PT and APTT reagents.
In summary, within-run, between-run and reconstituted stability studies support the substantial equivalence claim for Pacific Hemostasis Coagulation Control Level Theta to Dade® Ci-Trol® Level I. The one new performance characteristic, shorter clotting times, is intended to extend the verifiable limits of PT and APTT testing with more precision than controls that have longer clot times. Therefore based on the data provided, it is our conclusion that Control Level Theta is substantially equivalent to Dade® Ci-Trol® Level I.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an emblem featuring a stylized eagle with three wavy lines extending from its wing.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUN 17 1997
Lisa A. Worfolk, Ph.D. · Research Scientist Pacific Hemostasis 11515 Vanstory Drive, Suite 125 Hunterville, North Carolina 28078
K971428 Re : Coagulation Control Plasma (Abnormal) Level θ Requlatory Class: II Product Code: GGN April 16, 1997 Dated: Received: April 17, 1997
Dear Dr. Worfolk:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰੇ substantially equivalent determination assumes compliance with the Good Manufacturing -Practice -for Medical Devices: - General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
ન્દ્ર
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as ----described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Sitman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
5 10(k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: Coagulation Control Plasma_Abnormal Level Theta
ndications For Use:
Pacific Hemostasis Control Plasma (Abnormal) Level Theeided for use in a laboratory program of quality assurance for ARCT and Pro Moter Mith short clotting
laboratory program of quality assurance for Annus on blockho locally dotermined laboratory program or quality assurance the low the locally de semined
times. PT and APTT clotting times at the researce of olyves of times. Pr and APTI Cloceing cimes at the for char of elevated levels of norwal reference runge are of these proteins are elevated in response to an inflammatory stimulus, malignant disease process or trauma. Control Level Theta P inflamatory Scimaras) Marighant Girouse proven when of the reportable range of Flasma provent, particularly the imprecision of short clotting times often encountered with these screening tests.
IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF EDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <i>Veronica Celuia for Dr. Montgomery</i> (Division Sign-Off) |
|-----------------------------------------|---------------------------------------------------------------|
| Division of Clinical Laboratory Devices | |
| 510(k) Number | 0971428 |
| scription Use | | OR | Over-The-Counter Use |
|-------------------|--|----|--------------------------|
| r 21 CFR 801.109) | | | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.