K971265 · C.R. Bard, Inc. · DRF · Oct 23, 1997 · Cardiovascular
Device Facts
Record ID
K971265
Device Name
VIKING DIAGNOSTIC ELECTRODE CATHETER
Applicant
C.R. Bard, Inc.
Product Code
DRF · Cardiovascular
Decision Date
Oct 23, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
Bard Electrophysiology’s fixed curve diagnostic electrode catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
Device Story
Viking Diagnostic Electrode Catheter is a closed-lumen, non-steerable device used for electrophysiological mapping of cardiac structures. It features 6F diameter with 2-10 electrodes, varying inter-electrode spacings, and curve styles. Operated by clinicians in a clinical setting, the device is inserted into the heart to sense, record, stimulate, and provide temporary pacing. Output consists of electrical signals captured from cardiac tissue, which are used by physicians to evaluate cardiac arrhythmias and guide clinical decision-making. The device provides diagnostic information to assist in identifying arrhythmia mechanisms.
Clinical Evidence
Bench testing only. Performance evaluated against Cordis Webster and Bard Woven electrode catheters per March 1995 FDA Draft Guidance for Electrode Recording Catheters. Biocompatibility testing performed. Results confirmed device met all acceptance criteria and was clinically acceptable.
Technological Characteristics
Closed-lumen, non-steerable electrode recording catheter. 6F diameter, 2-10 electrodes. Materials include standard medical-grade components with minor variations in colorants. Biocompatibility verified. No active energy source or software.
Indications for Use
Indicated for temporary intracardiac sensing, recording, stimulation, and pacing during cardiac arrhythmia evaluation. Contraindicated for transseptal approach in patients with left atrial thrombus, myxoma, or interatrial baffle patch; contraindicated for retrograde transaortic approach in patients with aortic valve replacement.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}
C.R. Bard, Inc.
129 Concord Road
P.O. Box 566
Billerica, MA 01821
(508) 667-2511
K971265
BARD
OCT 23 1997
# 510(k) SUMMARY FOR THE VIKING DIAGNOSTIC ELECTRODE CATHETER
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990.
## A. Submitter’s Information
- Name: C.R. Bard, Inc.
- Address: 129 Concord Road, Billerica, MA 01821
- Phone: (508) 667-2511 extension 1065
- Fax: (508) 667-8594
- Contact Person: Douglas E. Ferguson, Senior Regulatory Affairs Coordinator
- Date of Preparation: March 31, 1997
## B. Device Name:
- Trade Name: Viking Diagnostic Electrode Catheter
- Common/Usual Name: Electrode Recording Catheter
- Classification Name: Electrode Recording Catheter
## C. Predicate Device Name(s):
- Cordis Webster Electrophysiology Catheter
- Bard Woven Electrode Catheter
## D. Device Description/Indications for Use:
### Description
The Viking Diagnostic Electrode Catheter is a closed lumen, non-steerable device. Typical of electrode recording catheters currently sold, the Viking catheter will be offered in 6F diameter with 2-10 electrodes with a variety of inter-electrode spacings and curve styles.
### Indications:
Bard Electrophysiology’s fixed curve diagnostic electrode catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
Viking Diagnostic Electrode Catheter 510(k) Submission
65
{1}
Viking Diagnostic Electrode Catheter 510(k) Submission
66
E. Technological Characteristics/Performance Data Summary
The "510(k) Substantial Equivalence Decision-Making Process (Detailed)" decision tree (CDRH 510(k) Manual 92-4158) was utilized in conjunction with the technological characteristics and performance testing results to make a determination of substantial equivalence as follows:
1. Does New Device Have Same Indication Statements?
Yes. Although the indications for the Viking catheter are more detailed than those of the Cordis Webster catheter, the uses of these devices are the same; they are to be used during the evaluation of cardiac arrhythmias for electrophysiological mapping of cardiac structures. The longer description in the indications for the Viking catheter identifies standard uses of diagnostic electrode catheters which are also contained in the indications for the currently sold preamendment Bard Woven electrode catheters and are referred to in the indications and instructions for use in the Bard® Steerable catheters (K921872; later referred to as the EP•XT™ catheters) and the Bard® Dynamic Tip and Tip Deflecting catheters (K891908, K904080, and K912213).
2. Does New Device Have Same Technological Characteristics, e.g., Design, Materials, etc.?
No. Although the Viking catheter has similar technological characteristics and component materials, they are not identical.
3. Could the New Characteristics Affect Safety or Effectiveness?
No. The only technological difference is in electrode width. However, since the distal and proximal electrode widths of the Viking catheter are within the range of those of the predicate devices, this is not an issue. Differences in materials are minor, such as the use of different colorants. Although these minor material differences are unlikely to affect safety, biocompatibility was verified through testing.
4. Are the Descriptive Characteristics Precise Enough to Ensure Equivalence?
No. Although the materials and technological characteristics are very similar, differences between vendors and manufacturing techniques necessitate bench testing to confirm equivalence.
{2}
5. Are Performance Data Available to Assess Equivalence?
Yes. The Viking catheter was tested for safety and performance based on the required characteristics of electrode recording catheters. Tests were chosen and developed based on the March, 1995 Draft Version of the “Electrode Recording Catheter Preliminary Guidance, Data to be Submitted to the Food and Drug Administration in Support of Premarket Notifications.” Where appropriate, comparative testing was done using the Cordis Webster catheter and/or a Bard Woven electrode catheter. In addition, as previously mentioned, the Viking catheter was subjected to biocompatibility testing.
6. Do Performance Data Demonstrate Equivalence?
Yes. The performance of the Viking catheter was found to meet all testing acceptance criteria and was therefore clinically acceptable. Where comparative testing was conducted, the Viking catheter performance was equivalent to, better than, or between the performances of the Cordis Webster and Bard Woven electrode catheters or the performance was determined to be clinically acceptable in every case. The Viking catheter passed all tests of biocompatibility.
SUBSTANTIALLY EQUIVALENT DETERMINATION:
The Viking Catheter is substantially equivalent to currently sold electrode recording catheters.
Viking Diagnostic Electrode Catheter 510(k) Submission
67
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
OCT 23 1997
Mr. Douglas E. Ferguson
C. R. Bard, Inc.
129 Concord Road
P.O. Box 566
Billerica, Massachusetts 01821
Re: K971265
Trade Name: Viking Diagnostic Electrode Catheters (38 models)
and SureLink Extension Cables (8 models)
Regulatory Class: II
Product Code: 74DRF
Dated: July 24, 1997
Received: July 28, 1997
Dear Mr. Ferguson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Douglas E. Ferguson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
# D. INDICATIONS FOR USE
Device Name: Viking Diagnostic Electrode Catheter
## Indications for Use:
Bard Electrophysiology’s fixed curve diagnostic electrode catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.
## Contraindications:
The transseptal approach is contraindicated in patients with left atrial thrombus or myxoma, or interatrial baffle patch. The retrograde transaortic approach is contraindicated in patients with aortic valve replacement.
---
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division:
Division of:
and Neurological:
510(k) Number | (Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number ~~K97265~~ |
| --- | --- |
Prescription Use ___________ (OR Over-the-Counter Use ___________)
(Per 21 CFR 801.109)
Viking Diagnostic Electrode Catheter 510(k) Submission
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.