← Product Code [HET](/productcode/HET) · K971255

# IMAGYN EXPANDABLE LAPAROSCOPIC INTRODUCER (K971255)

_Imagyn Medical, Inc. · HET · Apr 21, 1997 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K971255

## Device Facts

- **Applicant:** Imagyn Medical, Inc.
- **Product Code:** [HET](/productcode/HET.md)
- **Decision Date:** Apr 21, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Imagyn Expandable Laparoscopic Introducer is indicated for use in providing access to the peritoneal cavity for diagnostic and operative instruments during laparoscopic procedures.

## Device Story

Single-use, sterile laparoscopic introducer; metal and plastic construction. Used for second-site access during laparoscopic procedures. Incorporates valve system to maintain pneumoperitoneum (insufflation pressure) during instrument insertion/manipulation. Operated by surgeons in clinical/OR settings. Facilitates instrument passage while preventing gas loss; enables diagnostic and operative laparoscopic interventions.

## Clinical Evidence

Bench testing only. Performed visual, dimensional, leak, puncture, instrument passage, and mechanical testing to verify performance specifications and intended use.

## Technological Characteristics

Single-use, sterile device; metal and plastic components; integrated valve system for pneumoperitoneum maintenance.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- Imagyn MicroLap Introducer
- Innerdyne, Inc. Step Trocar Expandable Port

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

APR 21 1997

# 510(k) Summary

K971255

## Submitted By:

Imagyn Medical, Inc.
27651 La Paz Road
Laguna Niguel, CA 92677-3917
(714) 362-2500
Contact: Debra A. Rinderer

## Summary Preparation:

April 2, 1997

## Device:

Imagyn Expandable Laparoscopic Introducer

## Predicate Devices:

Imagyn MicroLap Introducer
Innerdyne, Inc. Step Trocar Expandable Port

## Device Description:

The Imagyn Expandable Laparoscopic Introducer is a single use, sterile device made from both metal and plastic components. For second-site use during laparoscopic procedures, the device incorporates a valve system that maintains adequate insufflation pressures while diagnostic or operative instruments are in use.

## Intended Use of Device:

The Imagyn Expandable Laparoscopic Introducer is indicated for use in providing access to the peritoneal cavity for diagnostic and operative instruments during laparoscopic procedures.

## Technological Comparisons:

The device has equivalent technological characteristics to the predicate devices to achieve laparoscope and/or laparoscopic instrument insertion without loss of pneumoperitoneum.

## Performance Summary:

Non-clinical tests were performed to demonstrate that the device performed according to its description and intended use. These included visual, dimensional, leak, puncture, instrument passage, and mechanical testing and demonstrated adequate performance for the intended use.

---

**Source:** [https://fda.innolitics.com/device/K971255](https://fda.innolitics.com/device/K971255)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
