OHMEDA INFANT RESUSCITATION SYSTEM

K971243 · Ohmeda Medical · BTM · Dec 2, 1997 · Anesthesiology

Device Facts

Record IDK971243
Device NameOHMEDA INFANT RESUSCITATION SYSTEM
ApplicantOhmeda Medical
Product CodeBTM · Anesthesiology
Decision DateDec 2, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5915
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Infant Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen and/or manual ventilation to the infant. These practices are well-described in collaborative guidelines established by the American Heart Association and the American Academy of Pediatrics. Resuscitation may be required whenever an infant fails to establish effective, adequate breathing patterns necessary to meet tissue oxygen demands and/or to rid the body of carbon dioxide.

Device Story

Infant Resuscitation System (IRS) provides equipment for pulmonary resuscitation of infants; used by clinicians to establish clear airways and deliver oxygen/manual ventilation. System components: suction device for clearing trachea/nasal passages; oxygen flowmeter for 100% oxygen delivery; airway pressure manometer with adjustable pressure relief valve (12-55 cm H2O) to monitor/limit pressure delivered via external resuscitation bag (not included). Optional gas blender allows air/oxygen mixing. Device connects to wall/cylinder gas sources via hoses. Clinician monitors airway pressure via manometer to facilitate lung pressurization and prevent over-inflation/alveolar rupture. Benefits include standardized, guideline-compliant resuscitation support during transition to extrauterine life.

Clinical Evidence

No clinical data. Bench testing conducted to ensure device meets specifications.

Technological Characteristics

Combination device including suction regulator, oxygen flowmeter, airway pressure manometer, and optional gas blender. Manometer features adjustable pressure relief valve (12-55 cm H2O). Connectivity via hoses to wall/cylinder gas outlets. Mechanical operation; no software or electronic components described.

Indications for Use

Indicated for infants requiring pulmonary resuscitation to establish a clear airway, provide oxygen, or manual ventilation due to failure to establish effective breathing patterns.

Regulatory Classification

Identification

A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} 1471243 # Ohmeda Infant Resuscitation System ## 510(k) Summary DEC - 2 1997 ## 1. Predicate Device Information The IRS is a combination of the following stand-alone medical devices: - Ohio Pressure Compensated Flowmeter - 510(k) No. K760102 - Ohio Continuous Suction Regulator (wall vacuum powered) - 510(k) No. K883229 - Ohio Continuous Suction Regulator (venturi powered) - Substantially equivalent to the wall vacuum powered configuration - Ohmeda Oxygen Blender - 510(k) No. K853905 - Airway Manometer - Substantially equivalent to the Airway Manometer included in the resuscitation section of the Air Shields Resuscitaire™ (a Class II Infant Radiant Warmer) In addition, the IRS is substantially equivalent to the resuscitation sections of the Hill Rom Stabilet 2000C Infant Warmer and to the Air Shields Resuscitaire™ Radiant Warmer. ## 2. Intended Use Statement The Infant Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen and/or manual ventilation to the infant. These practices are well-described in collaborative guidelines established by the American Heart Association and the American Academy of Pediatrics. Resuscitation may be required whenever an infant fails to establish effective, adequate breathing patterns necessary to meet tissue oxygen demands and/or to rid the body of carbon dioxide. ## 3. Functional Description The Infant Resuscitation System incorporates a suction device for routine clearance of mucous and debris from the trachea and nasal passages, as well as one oxygen flowmeter for delivery of oxygen to the infant requiring such therapy. The airway pressure manometer is attached to a resuscitation bag (not included) and measures the pressure delivered to the infant in order to facilitate adequate pressurization of the lungs. An optional gas blender mixes air and oxygen in percentages selected by the user. The Infant Resuscitation System includes hoses to connect the system to wall or cylinder outlets and/or assemblies. It does not include the resuscitation bag for manual ventilation. {1} 4. Assessment of Technological Characteristics The technological characteristics of the IRS are similar to those of the predicate devices to which the IRS is substantially equivalent. The IRS is a combination of devices which are either preamendment devices or have been cleared for commercial distribution via Premarket Notification. 5. Performance Data Pulmonary resuscitation of infants includes well established clinical practices. These practices are well-described in collaborative guidelines established by the American Heart Association and the American Academy of Pediatrics. The IRS provide clinicians with the means to implement these practices. Consequently, clinical testing to demonstrate safety and effectiveness is not required. Bench testing will be conducted to ensure that the device meets its specifications. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 2 1997 Mr. Alberto F. Profumo Ohmeda, Inc. 9065 Guilford Road Columbia, Maryland 21046-1801 Re: K971243 Ohmeda Infant Resuscitation System ("IRS") Regulatory Class: II (two) Product Code: 73 BTM Dated: September 4, 1997 Received: September 8, 1997 Dear Mr. Profumo: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Mr. Alberto F. Profumo This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} Page 1 of 1 510(k) Number (if known): K971243 Device Name: *Infant Resuscitation System* Indications For Use: ## Clinical Use Description *Infant Resuscitation System* The Infant Resuscitation System provides the basic equipment required for pulmonary resuscitation during infancy. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen and/or manual ventilation to the infant. These practices are well-described in collaborative guidelines established by the American Heart Association (AHA) and the American Academy of Pediatrics (AAP). Resuscitation may be required whenever an infant fails to establish effective, adequate breathing patterns necessary to meet tissue oxygen demands and/or to rid the body of carbon dioxide. The Infant Resuscitation System incorporates the following features for the practice of infant resuscitation: a suction device for routine clearance of the trachea and nasal passages; an oxygen flowmeter to deliver 100% oxygen to the infant requiring such therapy; and an airway pressure manometer with an adjustable pressure relief valve (12-55 cm H2O). The airway pressure manometer may be attached to a resuscitation bag (not included) and is used to measure the pressure being delivered to the infant in order to facilitate adequate pressurization of the lungs. An adjustable pressure relief valve is a safety feature which helps prevent the user from over-inflating the infant’s lungs, which could result in rupture of the alveoli. The valve may be adjusted to release excess gas at a user selected pressure point. The ability to adjust the pressure relief valve facilitates optimal oxygen delivery as lung compliance changes dynamically after birth during the transition to extrauterine life. The operative range (12-55 cm H2O) of this feature meets the recommendations by the AHA and AAP. The Infant Resuscitation System is outfitted with unregulated air and oxygen outlet connection which may be attached to an optional gas blender for the purposes of delivering air and oxygen mixes in the user selected percentages. The Resuscitation System may be equipped with optional hoses to connect the system to wall or cylinder outlets and/or yoke assemblies. The Infant Resuscitation System does not include the resuscitation bag for manual ventilation. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) Over- The Counter Use ☐ (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number _______________ (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...