SYNCHRON SYSTEMS THYROXINE REAGENT

K971203 · Beckman Instruments, Inc. · KLI · May 16, 1997 · Clinical Chemistry

Device Facts

Record IDK971203
Device NameSYNCHRON SYSTEMS THYROXINE REAGENT
ApplicantBeckman Instruments, Inc.
Product CodeKLI · Clinical Chemistry
Decision DateMay 16, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1700
Device ClassClass 2

Indications for Use

The SYNCHRON Systems Thyroxine (T4) Reagent is intended for the quantitative determination of total thyroxine concentrations in serum on SYNCHRON Clinical Systems.

Device Story

The SYNCHRON Systems Thyroxine (T4) Reagent is an in vitro diagnostic reagent kit used on Beckman SYNCHRON Clinical Systems. It quantitatively measures total thyroxine (T4) concentrations in serum samples. The device utilizes a chemical reaction principle identical to the predicate CEDIA T4 MAb Reagent. The system processes serum samples to provide T4 concentration values, which clinicians use to assess thyroid function. The device is intended for use in clinical laboratory settings by trained laboratory personnel. It supports the use of Sodium Heparin and Lithium Heparin as anticoagulants.

Clinical Evidence

Bench testing only. Method comparison study (n=97) against the predicate CEDIA MAb Thyroxine Reagent yielded a slope of 1.018, intercept of 0.12 μg/dL, and correlation coefficient (r) of 0.992. Imprecision studies (n=80 per level) showed within-run %C.V. ranging from 1.7% to 4.3% and total imprecision %C.V. ranging from 2.4% to 8.8% across three concentration levels (4.2–17.1 μg/dL). On-board stability is validated for 30 days.

Technological Characteristics

In vitro diagnostic reagent kit for quantitative T4 determination. Utilizes chemical reaction principle identical to predicate. Compatible with SYNCHRON Clinical Systems. Reagent on-board stability: 30 days. Specimen types: serum, Sodium Heparin, Lithium Heparin.

Indications for Use

Indicated for the quantitative determination of total thyroxine (T4) concentrations in human serum to aid in the assessment of thyroid function.

Regulatory Classification

Identification

A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{1} Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON Systems Thyroxine (T4) Reagent Summary of Safety & Effectiveness K971203 MAY 16 1997 Summary of Safety & Effectiveness SYNCHRON Systems Thyroxine (T4) Reagent ## 1.0 Submitted By: Lucinda Stockert Senior Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 961-3777 FAX: (714) 961-4457 ## 2.0 Date Submitted: 7 May 1997 ## 3.0 Device Name(s): ### 3.1 Proprietary Names SYNCHRON® Systems Thyroxine (T4) Reagent ### 3.2 Classification Name Total thyroxine test system (21 CFR § 862.1700) ## 4.0 Predicate Device(s): | SYNCHRON Reagent | Predicate | Predicate Company | Docket Number | | --- | --- | --- | --- | | SYNCHRON Systems Thyroxine (T4) Reagent | CEDIA T4 MAb Reagent | Boehringer Mannheim Corp., | K935880 | | SYNCHRON Systems Thyroxine (T4) Reagent | CEDIA T4 MAb Reagent- SYNCHRON Application | Boehringer Mannheim Corp., | K942575 | t4_510k.sse {2} Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON Systems Thyroxine (T4) Reagent Summary of Safety & Effectiveness # 5.0 Description: The SYNCHRON Systems Thyroxine (T4) Reagent kit, which includes the T4 Reagent and the T4 Calibrators, in intended for use on Beckman's SYNCHRON Clinical Systems. # 6.0 Intended Use: The SYNCHRON Systems Thyroxine (T4) Reagent is intended for the quantitative determination of total thyroxine concentrations in serum on SYNCHRON Clinical Systems. # 7.0 Comparison to Predicate(s): The following tables show similarities and differences between the predicates identified in Section 4.0 of this summary. ## SIMILARITIES to the PREDICATE | Reagent | Aspect/Characteristic | Comments | | --- | --- | --- | | SYNCHRON Systems Thyroxine (T4) Reagent | Intended use | Same as the predicate | | | Chemical Reaction | Same principle as the predicate | | | Formulation | Same source, processing as predicate | | | Calibration | Same as the predicate | | | Reagent Kit Configuration | Same as the predicate | ## DIFFERENCES from the PREDICATE | Reagent | Aspect/Characteristic | Comments | | --- | --- | --- | | SYNCHRON Systems Thyroxine (T4) Reagent | Specimen Type | SYNCHRON T4 Reagent claims Sodium Heparin and Lithium Heparin as acceptable anticoagulants while the Boehringer Mannheim CEDIA T4 MAb Reagent claims Sodium Heparin, Lithium Heparin, and Sodium EDTA as acceptable anticoagulants. | file: t4_510k.sse {3} Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON Systems Thyroxine (T4) Reagent Summary of Safety & Effectiveness ## 8.0 Summary of Performance Data: The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, imprecision studies, and stability studies that relate results obtained from the SYNCHRON Systems Thyroxine (T4) Reagent and the Boehringer Mannheim CEDIA MAb Thyroxine Reagent with a SYNCHRON Application. ### Method Comparison Study Results #### SYNCHRON Systems T4 Reagent | Analyte (Reagent) | Slope | Intercept (μg/dL) | r | n | Predicate Method | | --- | --- | --- | --- | --- | --- | | Thyroxine (T4) | 1.018 | 0.12 | 0.992 | 97 | Boehringer Mannheim CEDIA MAb Thyroxine Reagent | ### Estimated Imprecision #### SYNCHRON Systems T4 Reagent | Sample | Mean (μg/dL) | S.D. (μg/dL) | %C.V. | N | | --- | --- | --- | --- | --- | | Within-Run Imprecision | | | | | | Level 1 | 4.2 | 0.2 | 4.3 | 80 | | Level 2 | 8.6 | 0.2 | 2.4 | 80 | | Level 3 | 17.1 | 0.3 | 1.7 | 80 | | Total Imprecision | | | | | | Level 1 | 4.2 | 0.4 | 8.8 | 80 | | Level 2 | 8.6 | 0.4 | 4.3 | 80 | | Level 3 | 17.1 | 0.4 | 2.4 | 80 | ### On-Board Stability Study Results | Reagent | Product Claim | | --- | --- | | SYNCHRON Thyroxine (T4) | 30 days | This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92. file: t4_510k.sse
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...