SVL-9130 DEFIBRILLATION ELECTRODE

K971149 · Survivalink Corp. · MKJ · Jun 23, 1997 · Cardiovascular

Device Facts

Record IDK971149
Device NameSVL-9130 DEFIBRILLATION ELECTRODE
ApplicantSurvivalink Corp.
Product CodeMKJ · Cardiovascular
Decision DateJun 23, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The SVL-9130 electrodes are single use and intended to be used in conjunction with semi-automatic external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The electrodes are intended to be used with AEDs that specify the use of disposable electrodes that meet AAMI DF-39 standards. These AEDs must have compatible electrode connectors. The VivaLink AED and the V2 are examples of such AEDs. The electrodes are intended for short term use and must be used before the expiration date listed on the packaging. The AEDs are used for emergency treatment of cardiac arrest patients who weigh more than 90 pounds. The user assesses the patient’s condition and confirms that the patient is unconscious, pulseless and is not breathing prior to applying the electrodes to the skin.

Device Story

SVL-9130 is a pair of hydrogel polymeric self-adhesive electrode pads; used as a single-use accessory for semi-automatic external defibrillators (AEDs). Electrodes interface with AEDs to monitor cardiac signals and deliver defibrillation energy to patients in cardiac arrest. Operated by emergency responders or trained users; user assesses patient condition (unconscious, pulseless, non-breathing) before application. Electrodes must meet AAMI DF-39 standards and be compatible with specific AED connectors. Device facilitates emergency life-saving intervention by enabling the AED to analyze heart rhythms and deliver necessary electrical shocks.

Clinical Evidence

Bench testing only. The device was tested against AAMI DF-39 specifications and internal company specifications to ensure performance and safety. No clinical data was required or provided.

Technological Characteristics

Hydrogel polymeric self-adhesive electrode pads. Designed for single-use. Compliant with AAMI DF-39 standards for disposable electrodes. Passive electrical interface for monitoring and defibrillation energy delivery.

Indications for Use

Indicated for emergency treatment of cardiac arrest patients weighing >90 lbs. Used with semi-automatic external defibrillators (AEDs) to monitor and deliver defibrillation energy. Requires patient to be unconscious, pulseless, and not breathing.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Submission Summary (Full Text)

{0} K971149 JUN 23 1997 # ATTACHMENT H: 510(K) SUMMARY H-1 {1} # 510(k) Summary ## H-1. ADMINISTRATIVE INFORMATION ### H-1.1 Name and address Submitted by: SurVivaLink Corporation 5430 Feltl Road Minneapolis, MN 55343 Contact Person: Sew-Wah Tay, Ph.D. Telephone No.: 612-939-4181 ext. 142 Facsimile No.: 612-939-4191 Date Prepared: March 26, 1997 ### H-1.2 Device Name Common or Usual Name: Disposable Polymer (Hydrogel) External Monitoring and Defibrillation Electrode Device Name: SVL-9130 Trade Name: SVL-9130 ### H-1.3 Classification Name Disposable Single Use Accessory (Electrode) to: a) Semi-automatic low energy DC defibrillator 21CFR§870.5300; Class II b) Cardiac Monitor (Cardiotachometer and Rate Alarm) 21CFR§870.2300; Class II Note: FDA has determined that Automatic External Defibrillators are currently classified as class III devices ### H-1.4 Applicant Applicant’s Name: SurVivaLink, Corporation 5430 Feltl Road Minneapolis, MN 55343 ## H-2. PREDICATE DEVICE 1. SurVivaLink Model 9010 (9010) electrodes manufactured for SurVivaLink Corporation by Katecho Inc.models (K940445) 2. Katecho K-Defib/Pace electrodes (K914955) ## H-3. INDICATION FOR USE The SVL-9130 electrodes are single use and intended to be used in conjunction with semi-automatic external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The electrodes are intended to be used with AEDs that specify the use of H-2 {2} disposable electrodes that meet AAMI DF-39 standards. These AEDs must have compatible electrode connectors. The VivaLink AED and the V2 are examples of such AEDs. The electrodes are intended for short term use and must be used before the expiration date listed on the packaging. The AEDs are used for emergency treatment of cardiac arrest patients who weigh more than 90 pounds. The user assesses the patient’s condition and confirms that the patient is unconscious, pulseless and is not breathing prior to applying the electrodes to the skin. ## H-4 DEVICE DESCRIPTION The SVL-9130 electrodes consists of a pair of hydrogel polymeric self-adhesive electrode pads of equal dimension. The electrode are packaged in such a way that the two conductive areas are in electrical contact. ## H-5. SUBSTANTIAL EQUIVALENCE The Company’s SVL-9130 electrodes covered by this submission are substantially equivalent to other legally marketed electrodes for semi-automatic low power DC defibrillators. Specifically, the SVL-9130 electrode is substantially equivalent to SVL-9010 electrodes (K940455) and Katecho’s D-Defib/Pace electrodes (K 914955). ## H-6. PERFORMANCE DATA The SVL-9130 electrodes meet all the specifications for single use hydrogel electrodes of the AAMI DF-39 specifications and SurVivaLink’s internal specifications. In all instances, the SVL-9130 electrodes functioned as intended. {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 23 1997 Sew-Wah Tay, Ph.D. SurVivaLink Corporation 5420 Feltl Road Minneapolis, Minnesota 55343 Re: K971149 SurVivaLink 9130 Defibrillation Electrode Regulatory Class: III (three) Product Code: 74 MKJ Dated: March 27, 1997 Received: March 28, 1997 Dear Dr. Tay: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Sew-Wah Tay, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} # 8. INDICATION FOR USE The SVL-9130 electrodes are single use and intended to be used in conjunction with semi-automatic external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The electrodes are intended to be used with AEDs that specify the use of disposable electrodes that meet AAMI DF-39 standards. These AEDs must have compatible electrode connectors. The VivaLink AED and the V2 are examples of such AEDs. The electrodes are intended for short term use and must be used before the expiration date listed on the packaging. The AEDs are used for emergency treatment of cardiac arrest patients who weigh more than 90 pounds. The user assesses the patient’s condition and confirms that the patient is unconscious, pulseless and is not breathing prior to applying the electrodes to the skin. Prescription Use (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K971149
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...