For in vitro diagnostic use to identify and enumerate percentages and absolute counts of T and NK lymphocytes in blood.
Device Story
System uses conjugated monoclonal antibodies (CD3 FITC, CD16+CD56 PE, CD45 PerCP) to stain leukocyte surface antigens in whole blood samples. Samples processed via erythrocyte lysis; analyzed using flow cytometry. For absolute counts, TRUCOUNT tubes containing known bead quantities are used. Device identifies lymphocyte subsets via fluorescence gating on CD45-positive populations, reducing non-lymphocyte contamination. System computes ratios of reagent-positive events to CD45-positive events for percentages; for absolute counts, computes ratio of reagent-positive events to bead events. Used in clinical laboratories by trained personnel to assist in diagnosing/monitoring patients with tumors or viral infections. Provides quantitative data to clinicians for immune status assessment.
Clinical Evidence
Multi-site clinical studies (Cleveland Clinic, Johns Hopkins, etc.) evaluated accuracy, linearity, reproducibility, and stability. Accuracy demonstrated equivalence to predicate and predicate plus automated differential cell counter. Linearity confirmed across 200–16,700 lymphocytes/μL. Within-specimen reproducibility acceptable. Stability studies established 6-hour post-draw processing window. No significant differences found using isotype controls.
Technological Characteristics
Three-color direct immunofluorescence reagent (CD3 FITC, CD16+CD56 PE, CD45 PerCP). Uses flow cytometry for detection. Employs fluorescence gating on CD45+ populations. Includes TRUCOUNT tubes with internal bead standards for absolute quantification. Compatible with various flow cytometers.
Indications for Use
Indicated for use with flow cytometers on erythrocyte-lysed whole blood to identify and enumerate percentages and absolute counts of CD3+ and CD16+CD56+ lymphocytes in normal individuals and patients with tumors or viral infections. May be used with or without isotype control.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
NOV 25 1997
K971110
# Attachment F
## Summary of Safety and Effectiveness
### Submitter Information (21 CFR 807.92(a)(1))
Submitter: Becton Dickinson Immunocytometry Systems
2350 Qume Drive
San Jose, CA 95131-1807
Contact: Anna Longwell, Esq.
Director, Regulatory Affairs - Corporate
(408) 954-2254
Summary date: March 21, 1997
### Name of Device and Classification (21 CFR 807.92(a)(2))
Name: Becton Dickinson TriTEST™ reagent CD3 FITC/CD16+CD56 PE/CD45 PerCP; TRUCOUNT™ Absolute Count Tubes
Classification: Class II
### Predicate Device (21 CFR 807.92(a)(3))
The BDIS TriTEST™ CD3 FITC/CD16+CD56 PE/CD45 PerCP reagent with TRUCOUNT Absolute Count Tubes when used to enumerate percentages of T lymphocytes and Natural Killer lymphocytes (NK cells), is substantially equivalent to Simultest™ IMK-Lymphocyte Tube F (cleared to market under 510(k) K913192). When used to enumerate absolute counts of T and NK cells, it is substantially equivalent to Simultest IMK-Lymphocyte Tube F plus automated differential cell counter (ADCC).
### Description of the Device (21 CFR 807.92(a)(4))
The BDIS TriTEST CD3 fluorescein isothiocyanate (FITC)/CD16+CD56 phycoerythrin (PE)/CD45 peridinin chlorophyll protein (PerCP) reagent is a three-color, direct immunofluorescence reagent for identifying and enumerating percentages of T lymphocytes (CD3+) and Natural Killer lymphocytes (CD3- and CD16+ and/or CD56+) in erythrocyte-lysed whole blood (LWB). When used with TRUCOUNT Absolute Count Tubes, the product will yield absolute counts in cells/μL. The Becton Dickinson TriTEST/TRUCOUNT system for immunophenotyping consists of a flow cytometer (either from BDIS or from another manufacturer), conjugated monoclonal reagent (TriTEST CD3 FITC/CD16+CD56 PE/CD45 PerCP) and TRUCOUNT Absolute Count Tubes.
The process to obtain lymphocyte subset percentages includes: 1) obtaining a whole blood sample, 2) cell-surface antigen staining with three-color monoclonal antibody reagents, 3) erythrocyte lysis, and 4) flow cytometric acquisition and analysis of list mode data. Analysis involves computing the ratio of reagent-positive events to the CD45 positive events, and expressing the ratio as a percentage.
To obtain absolute counts, the TriTEST reagent and whole blood are added directly to an Absolute Count Tube prior to lysis. The remaining process, until analysis, is identical to that for percentages. Analysis for absolute counts requires that an additional region, the bead region, be identified and the events in this region counted. The proportion of reagent positive
323
{1}
# Summary of Safety and Effectiveness
events to bead events (P) is computed. The absolute count is P x (beads/pellet)/(volume of blood sample).
When monoclonal antibody reagents are added to human whole blood, the fluorochrome-labeled antibodies bind specifically to antigens on the surface of leukocytes, thus identifying lymphocyte populations. The patient blood sample is added to the counting bead pellet and is treated with fluorochrome-labeled antibodies and the erythrocytes are lysed with FACS® Lysing Solution. The flow cytometer is set up so that cell populations for most samples occupy approximately the same region of fluorescence space. The sample is then introduced into the flow cytometer and the stained cells and beads fluoresce when excited by a laser beam.
The three-color reagent permits identification of lymphocyte subsets using fluorescence gating instead of scatter gating. This three-color reagent allows direct gating on the CD45-positive population using a combination of fluorescence and side scatter parameters. By gating on the CD45-positive population, a maximum number of lymphocytes may be captured in the gate and non-lymphocyte contamination may be minimized.
## Intended Use (21 CFR 807.92(a)(5))
For in vitro diagnostic use to identify and enumerate percentages and absolute counts of T and NK lymphocytes in blood.
## Indications for Use
- For use with any flow cytometer with specified detection ranges
- For use with erythrocyte lysed whole blood
- For use with or without an isotype control
- For in vitro diagnostic use
- To identify and enumerate percentages and absolute counts of CD3+ and CD16+CD56+ lymphocytes in normal individuals and patients with certain tumors and viral infections
## Comparison to Predicate Device (21 CFR 807.92(a)(6))
The three-color reagent (for percentages) is substantially equivalent to Simultest IMK-Lymphocyte Tube F in that they share the same intended uses and use essentially the same monoclonal antibody/flow cytometric methodology. The TriTEST and Simultest products differ in the steps used to determine analysis gates to identify the lymphocyte population.
The TriTEST/TRUCOUNT system (for absolute counts) is substantially equivalent to Simultest IMK-Lymphocyte Tube F plus ADCC in that they share the same intended uses. Compared to the predicate, the TriTEST/TRUCOUNT system requires less sample handling, a simpler sample preparation scheme and fewer sample tubes to achieve the same end result. Results demonstrate that the products yield essentially equivalent performance characteristics.
## Performance Data (21 CFR 807.92(b)(2))
Performance of the product was established by testing at Cleveland Clinic, Johns Hopkins Hospital, Institute of Tropical Medicine, University of North Carolina, and at Becton Dickinson Immunocytometry Systems laboratories in San Jose, California.
CD3/CD16+CD56/CD45; TRUCOUNT 510(k) Notification
Attachment F: Summary of Safety and Effectiveness
{2}
# Summary of Safety and Effectiveness
Several studies were performed:
- **Accuracy** was determined for both percentages and absolute counts. Accuracy data demonstrated the TriTEST product's equivalence to Simultest and TriTEST/TRUCOUNT product's equivalence to Simultest/ADCC.
- Use of an *Isotype Control* was studied. Data was analyzed for percent T lymphocytes and percent NK lymphocytes first using a control to set lymphocyte gate and quadrant markers and then using only the stained sample to set a lymphocyte gate and quadrant markers. Data indicated that the reagent may be used with or without an isotype control.
- **Reference range** studies were performed. Many variables, such as sex, age and geographical location may influence the reference range. Each site must determine its own reference range.
- A *stability* study was conducted to assess the time effect relating to age of blood (time-from-draw) and the time effect relating to the age of the stain (time-from-sample preparation), as well as the combined effect of both. Stability was determined for both percentages and absolute counts. Results indicate that samples should be stained and analyzed within 6 hours of draw.
- **Within-specimen reproducibility** was performed at BDIS; 10 replicates from 3 specimens were assessed. Within-specimen reproducibility was also performed at 3 clinical sites; 3 aliquots from each donor were assessed. Results demonstrated acceptable within-sample reproducibility.
- **Linearity** was determined using blood samples from 3 normal donors diluted to 5 concentrations, ranging from 16,700 to 200 lymphocytes/μL and from 31,000 to 2500 WBC/μL. Results indicate a linear response over this range.
- **Cross reactivity** of these clones is reported in the literature. Conjugation and product formulation have not changed their specificity.
- Results from a *cross platform reproducibility* study indicated that the TriTEST reagent CD3/CD16+CD56/CD45, with or without TRUCOUNT tubes, can be used on flow cytometers not made by Becton Dickinson.
## Performance Data - Conclusions (21 CFR 807.92(b)(3))
The results of the clinical studies demonstrate that the device is as safe and effective as the predicate device.
CD3/CD16+CD56/CD45; TRUCOUNT 510(k) Notification
Attachment F: Summary of Safety and Effectiveness
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Cindy Morrow
Sr. Regulatory Specialist
BECTON DICKINSON IMMUNOCYTOMETRY SYSTEMS
2350 Qume Drive
San Jose, CA 95131-1807
Re: K971110
Trade Name: Becton Dickinson TriTEST Reagent CD3 FITC/CD16+CD56 PE/CD45 PerCP; TruCOUNT Absolute Count Tubes
Regulatory Class: II
Product Code: GKZ, 81
Dated: March 24, 1997
Received: March 26, 1997
Dear Ms. Morrow:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number (if known): K971110
Device Name:
## Indications For Use:
- For use with any flow cytometer with specified detection ranges
- For use with erythrocyte lysed whole blood
- For use with or without an isotype control
- For *in vitro* diagnostic use
- To identify and enumerate percentages and absolute counts of CD3+ and CD16+CD56+ lymphocytes in normal individuals and patients with certain tumors and viral infections
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)