LIFESCAN SURESTEP PRO GLUCOSE CONTROL/LINEARITY SOLUTIONS
Device Facts
| Record ID | K971014 |
|---|---|
| Device Name | LIFESCAN SURESTEP PRO GLUCOSE CONTROL/LINEARITY SOLUTIONS |
| Applicant | Lifescan, Inc. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Apr 23, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
SureStep®Pro Glucose Control Solutions are for use with SureStep® Brand Blood Glucose Meters and SureStep®Pro Test Strips as a quality control material for verification of System performance. The SureStep®Pro Linearity Test Kit is for use with SureStep® Brand Glucose Meters and SureStep®Pro Test Strips for verification of System linearity.
Device Story
SureStep®Pro Glucose Control and Linearity Test Solutions; aqueous glucose-based solutions; used with SureStep® blood glucose meters and test strips. Control solutions verify system performance by comparing meter results against expected ranges printed on test strip bottles. Linearity test solutions verify system linearity across measurement range. Used in clinical or home settings by patients or healthcare providers. Output provides qualitative/quantitative verification of system accuracy; failure to fall within expected ranges indicates system malfunction. Benefits patient by ensuring reliable glucose monitoring results.
Clinical Evidence
Bench testing only. Performance verified by comparing meter results using control/linearity solutions against established expected ranges to confirm system accuracy and linearity.
Technological Characteristics
Aqueous glucose-based solutions; formulated to provide specific glucose concentrations for system verification. Compatible with SureStep® brand meters and test strips. No electronic components, software, or connectivity.
Indications for Use
Indicated for use by healthcare professionals or patients as a quality control material to verify performance and linearity of LifeScan SureStep® Brand Blood Glucose Meters and SureStep®Pro Test Strips.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- LifeScan SureStep® Glucose (Normal) Control Solution
- Streck Laboratories Sugar-Chex® Linearity kit
Related Devices
- K973676 — PRESTIGE AQUEOUS GLUCOSE CONTROL LOW, PRESTIGE AQUEOUS GLUCOSE CONTROL HIGH · Sorrento Biochemical, Inc. · Dec 16, 1997
- K070506 — SMS MULTI GLUCOSE CONTROL, MODEL 2120049 · Medical Specialties Supply Co., Inc. · Apr 18, 2007
- K092645 — FDI GLUCOSE CONTROL LEVELS 1, 2 AND 3 · Fujirebio Diagnostics,Inc. · Sep 5, 2009
- K093724 — FORA CONTROL SOLUTION, TAIDOC CONTROL SOLUTION · Taidoc Technology Corporation · Jul 29, 2010