K970991 · Promedic, Inc. · BTM · Jun 16, 1997 · Anesthesiology
Device Facts
Record ID
K970991
Device Name
DATEX ENGSTOM MANUAL RESUSCITATOR
Applicant
Promedic, Inc.
Product Code
BTM · Anesthesiology
Decision Date
Jun 16, 1997
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 868.5915
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
To ventilate the apnoeic patient and to augment ventilation and / or oxygen delivery to the spontaneously breathing patient. For use as an adjunct to artificial respiration and cardiopulmonary resuscitation. Environment of use: Hospital - Emergency Medical Services, Operating Room, during patient transport, Intensive Care Units, where patient assisted ventilation may be administered or required.
Device Story
Manual resuscitator designed for emergency and clinical settings; used by trained medical personnel to provide assisted ventilation. Device consists of four primary components: silicone bag, rebreathing valve, reservoir valve, and oxygen reservoir. Operator manually compresses silicone bag to deliver air/oxygen to patient via face mask or endotracheal tube. May incorporate PEEP valve to maintain positive end-expiratory pressure. Available in three sizes. Device is reusable; components can be disassembled for cleaning via autoclave or EtO sterilization. Provides immediate respiratory support for apnoeic or spontaneously breathing patients in hospitals, ORs, ICUs, and during transport.
Clinical Evidence
Bench testing only. Device meets all specifications and requirements of ASTM 920 / ISO 8382 for manual resuscitators.
Technological Characteristics
Manual, non-powered resuscitator. Materials: silicone (bag, duckbill valve), polysulfone (patient connector, housing), aluminum (pop-off valve), stainless steel (PEEP valve spring), PVC (reservoir bag, face mask). Patient fittings: 15/22 mm per ASTM 1054 / ISO 5356. Sterilization: autoclave, EtO. No software or electronic components.
Indications for Use
Indicated for adult, child, and infant patients requiring assisted ventilation, including apnoeic patients or spontaneously breathing patients needing augmented ventilation/oxygen delivery. Used as an adjunct to artificial respiration and CPR by qualified medical personnel.
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}
Datex Engstrom
K970991
Non-Confidential Summary of Safety and Effectiveness
page 1 of 3
March 12, 1997
JUN 16
Datex Engstrom AB
Box 20109
S-16102 Bromma SWEDEN
Tel - 011-46-8-629-3100
Fax - 011-46-8-298418
Official Contact: Arne Salo, President
Proprietary or Trade Name: Datex Engstrom reusable manual resuscitator
Common/Usual Name: Manual resuscitator
Classification Name: Ventilator, Emergency, Manual (Resuscitator)
Device: Datex Engstrom Reusable Manual Resuscitator
Predicate Devices:
Hudson RCI - Durable Manual resuscitator - K895589
Hudson RCI - PEEP valves - K902062
## Device Description:
The Datex Engstrom manual resuscitator is designed for use as an adjunct to artificial respiration and CPR. It can be used to ventilate the apnoeic patient and to augment ventilation and / or oxygen delivery to the spontaneously breathing patient.
Indicated Use -- Provide assisted ventilation
Environment of Use -- Emergency medical services, hospital, patient transport,
Operating Room (OR), and where assisted ventilation may be required.
Patient population -- Adult, child and infant
## Comparison to Predicate Devices:
Datex-Engstrom AB
Postadress/Postal address
Box 20109
Gatuadress/Visiting address
Karlskrona-ager 18
Telefon/Telephone
Nat 08-629 31 00
Telefax
Nat 08-29 84 18
Reg Nr/VAT No
536044-63-4
{1}
# Non-Confidential Summary of Safety and Effectiveness (continued)
page 2 of 3
March 12, 1997
| Attribute | Datex-Engstrom | Hudson RCI |
| --- | --- | --- |
| **Use** | | |
| Intended for use an adjunct to artificial respiration and CPR | Yes | Yes |
| Can be used to ventilate apnoeic patients | Yes | Yes |
| Augment ventilation and / or oxygen delivery to spontaneously breathing patients | Yes | Yes |
| Environment of use - Hospital, OR, anesthesia, PACU, ICU, transport, EMS, where assisted ventilation may be required or needed | Yes | Yes |
| Intended for use by qualified medical and emergency personnel trained in pulmonary ventilation and advanced cardiac life support techniques | Yes | Yes |
| Indicated for cleaning and reuse | Yes | Yes |
| Indicated population - adult, child, infant | Yes | Yes |
| **Design / Theory of Operation** | | |
| Connects to a Face mask or endotracheal tube | Yes | Yes |
| Has four major components | | |
| Rebreathing valve, Silicone bag, Reservoir valve, Oxygen reservoir | Yes | Yes |
| May incorporate a PEEP valve | Yes | Yes |
| Available is three (3) sizes | Yes | Yes |
| Has available oxygen reservoir | Yes | Yes |
page 5 of 76
{2}
Page 6 of 76
# Non-Confidential Summary of Safety and Effectiveness
(continued)
page 3 of 3
March 12, 1997
| Attribute | Datex-Engstrom | Hudson RCI |
| --- | --- | --- |
| **Design / Theory of Operation** | | |
| Fits to standard 15 / 22 mm patient end fittings | Yes | Yes |
| Offered with standard reusable face mask - various styles | Yes | Yes |
| May be taken apart and cleaned | Yes | Yes |
| Cleaning methods - autoclave, EtO | Yes | Yes |
| **Materials** | | |
| Bag and duckbill valve - silicone | Yes | Yes |
| Patient connector, housing parts made of Polysulfone | Yes | Yes |
| Pop-off valve - aluminum | Yes | Yes |
| PEEP valve spring stainless steel | Yes | Yes |
| Face mask - silicone or PVC | Yes | Yes |
| Reservoir bag - PVC | Yes | Yes |
| **Performance Standards / Specifications** | | |
| Patient fittings - 15 / 22 mm ASTM 1054 / ISO 5356 | Yes | Yes |
| Meets all specifications, testing and requirements of ASTM 920 / ISO 8382 | Yes | Yes |
| **Differences between Other Legally Marketed Predicate Devices** | | |
| There is no differences between the intended device and the predicate devices which would be significant to patient safety or effectiveness. | | |
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 16 1997
Mr. Paul Dryden
Datex Engstrom AB
c/o ProMedic, Inc.
6329 W. Waterview Court
McCordsville, Indiana 46055-9501
Re: K970991
Datex-Engstrom Medical Reusable Manual Resuscitator
Regulatory Class: II (two)
Product Code: 73 BTM
Dated: March 13, 1997
Received: March 19, 1997
Dear Mr. Dryden:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device
{4}
Page 2 - Mr. Paul Dryden
can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
JUN 16 1991
# SECTION 3 INDICATIONS FOR USE
Page 1 of 1
Pursuant to the Notice of February 6, 1996 regarding listing of Indications for Use on a separate sheet, the following is per that request.
510(k) Number: K970991 (To be assigned)
Device Name: Reusable Manual Resuscitator
Intended Use: To ventilate the apnoeic patient and to augment ventilation and / or oxygen delivery to the spontaneously breathing patient. For use as an adjunct to artificial respiration and cardiopulmonary resuscitation.
Environment of use: Hospital - Emergency Medical Services, Operating Room, during patient transport, Intensive Care Units, where patient assisted ventilation may be administered or required.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Christy Foreman
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970991
Prescription Use ☑ (Per CFR 801.109)
or
Over-the-counter use ☐
Page 10 of 76
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.