← Product Code [GEI](/productcode/GEI) · K970987

# ELECTROSURGICAL ELECTRODE (K970987)

_Fusion Medical Technologies, Inc. · GEI · May 5, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K970987

## Device Facts

- **Applicant:** Fusion Medical Technologies, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** May 5, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Intended for use by practitioners in a manner identical to similar electrosurgical electrodes of this type presently in commercial distribution. It is intended to provide electrocautery over a large surface.

## Device Story

Fusion Ball Tip Electrode; reusable monopolar electrosurgical accessory. Input: electrical energy from standard electrosurgical generator via electrosurgical pencil/chuck. Operation: delivers high-frequency current to tissue via 99% pure silver ball tip for electrocautery. Used by clinicians in surgical/clinical settings. Output: thermal tissue effect (coagulation/cautery). Benefits: provides large surface area electrocautery; reusable design via autoclave sterilization.

## Clinical Evidence

Bench testing only; no clinical data provided. Equivalence based on design, material biocompatibility, and intended use similarity to predicate.

## Technological Characteristics

Monopolar electrosurgical electrode. Materials: Stainless steel shank, Polyphenylsulfone insulation, 99% pure silver tip. Dimensions: 3/32 inch (2.4 mm) shaft. Energy: Electrosurgical (RF). Sterilization: Gamma irradiation (initial), autoclave (reusable). Biocompatibility: ISO 10993-1 compliant.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Electrosurgical Electrode by Premier Dental Products ([K923332](/device/K923332.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

046
K970987

510(k) Summary
Fusion Medical Technologies, Inc.
Ball Tip Electrode
MAY - 5 1997

1. Sponsor: Fusion Medical Technologies, Inc.
1615 Plymouth Street
Mountain View, CA 94043

Contact Person: Debera Brown
Vice President, Regulatory and Clinical Affairs

2. Device Name:
Trade Name: To Be Determined
Common/Usual Name: Electrosurgical Electrode
Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories

3. Identification of the Predicate or Legally Marketed Devices to Which Equivalence is being Claimed:
Electrosurgical Electrode by Premier Dental Products

4. Device Description:
The Fusion Ball Tip Electrode is a reusable electrode for use in monopolar electrosurgical accessories.

5. Intended Use:
Intended for use by practitioners in a manner identical to similar electrosurgical electrodes of this type presently in commercial distribution. It is intended to provide electrocautery over a large surface.

{1}

047

6. A Statement of How the Technological Characteristics of the Device Compare to Those of the Predicate or Legally Marketed Device.

|  Parameter | Fusion Ball Tip Electrode | Premier Dental Electrode K923332  |
| --- | --- | --- |
|  Indications for Use | Intended for use by practitioners in a manner identical to similar electrosurgical electrodes of this type presently in commercial distribution. It is intended to provide electrocautery over a large surface. | Intended for use by practitioners in a manner identical to similar electrosurgical electrodes of this type presently in commercial distribution. It is intended for cutting and coagulation of tissue.  |
|  Use Profile | Provided Sterile; reusable after cleaning and autoclave sterilization | Provided non-sterile; first use and reuse after autoclave or ETO sterilization  |
|  Compatibility | Standard 3/32 inch (2.4 mm) shaft fits most electrosurgery chucks and pencils | 44 different configurations provided, all standard 3/32 shaft to fit most electrosurgery chucks and pencils  |
|  Material Composition | Stainless Steel shank, Polyphenylsulfone for insulation, 99% Pure Silver as electrode tip | Stainless Steel shank, Korvex 4X Teflon TFE heat shrinkable tubing, Type 304 stainless steel electrode tip  |
|  Sterilization | Provided sterile by gamma irradiation for first use | Provided non-sterile, must be sterilized by user before first use  |
|  Biocompatibility | Materials used in other medical devices; ISO 10993-1 | Materials used in other devices  |

---

**Source:** [https://fda.innolitics.com/device/K970987](https://fda.innolitics.com/device/K970987)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
