K970873 · American Medical Devices, Inc. · HMX · May 20, 1997 · Ophthalmic
Device Facts
Record ID
K970873
Device Name
ENDOSOFT SOFT TIP NEEDLE (20 GA. AND 19 GA.)
Applicant
American Medical Devices, Inc.
Product Code
HMX · Ophthalmic
Decision Date
May 20, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4350
Device Class
Class 1
Indications for Use
For irrigation/aspiration during eye surgery
Device Story
Endosoft Soft Tip Needle is a surgical instrument used during eye surgery. The device consists of a stainless steel needle with a plastic luer lock hub and a soft silastic tubing tip. It connects to a conventional extrusion handpiece via the male luer lock. The device is used by ophthalmic surgeons to perform irrigation and aspiration of the eye. The soft tip design is intended to facilitate these procedures during surgery.
Clinical Evidence
No clinical data; bench testing only. Sterilization validated using AAMI Overkill Method; non-pyrogenic status confirmed via LAL method.
Technological Characteristics
Stainless steel needle, silastic tubing, plastic luer lock hub. ETO sterilization (AAMI Overkill Method). Packaging: Tyvek pouch with polymylar sheath. SAL 10^-6. Non-pyrogenic (LAL method).
Indications for Use
Indicated for irrigation and aspiration during eye surgery.
Regulatory Classification
Identification
A manual ophthalmic surgical instrument is a nonpowered, handheld device intended to aid or perform ophthalmic surgical procedures. This generic type of device includes the manual corneal burr, ophthalmic caliper, ophthalmic cannula, eyelid clamp, ophthalmic muscle clamp, iris retractor clip, orbital compressor, ophthalmic curette, cystotome, orbital depressor, lachrymal dilator, erisophake, expressor, ophthalmic forcep, ophthalmic hook, sphere introducer, ophthalmic knife, ophthalmic suturing needle, lachrymal probe, trabeculotomy probe, cornea-sclera punch, ophthalmic retractor, ophthalmic ring (Flieringa), lachrymal sac rongeur, ophthalmic scissors, enucleating snare, ophthalmic spatula, ophthalmic specula, ophthalmic spoon, ophthalmic spud, trabeculotome or ophthalmic manual trephine.
Reference Devices
Visitec
Grieshaber
Submission Summary (Full Text)
{0}
K970873
AMERICAN MEDICAL DEVICES, Inc.
1100 Northside Drive Atlanta, Georgia 30318
MAY 20 1997
February 26, 1997
# Premarket Notification [510(k)] Summary
**Submitter:** American Medical Devices, Inc
1100 Northside Drive
Atlanta, GA 30318
Phone: (404) 815-5233
Fax: (404) 815-5235
**Official Correspondent:** Frank J. Tighe
**Trade Name:** The American Medical Devices, Inc., 19 and 20 Ga. Endosoft™ Soft Tip Needles
**Common Name:** Irrigation/Aspiration Needle
**Registration Number:** We have registered but have not received our application back as of this date.
**Class:** Class 1
**Class Name:** We were unable to find the device listed in the classification regulations, 21 CFR Parts 862-892 [807.87 (c)].
**Panel:** Ophthalmic
**Product Code:** HMX
Page 1 Summary
Toll-Free Direct Facsimile Internet
800-793-1473 404-815-5233 404-815-5235 ftighe@americanmedicaldevices.com
{1}
K970873
Device Description: The American Medical Devices, Inc 19 and 20 ga. Endosoft™ Soft Tip Needles are for irrigation/aspiration during eye surgery. The device consists of a stainless steel needle with a plastic luer lock on one end and a protruding piece of soft silastic tubing at the tip. This device connects to any conventional extrusion handpiece with a male luer lock adapter.
Statement of indications for use. - For irrigation/aspiration during eye surgery.
## Substantial Equivalence Comparison
| | American Medical Devices, Inc. | Visitec | Grieshaber |
| --- | --- | --- | --- |
| Packaging | X | X | X |
| Tyvek to Poly | | | |
| For irrigation/aspiration | X | X | X |
| Of the eye during surgery | | | |
| Materials: Stainless Steel Needle, Silastic Tubing, Plastic Hub | X | X | X |
| Sterilization ETO | X | X | X |
## Sterility
The Device will be ETO Sterilized.
The method used to validate the sterilization cycle is AAMI Overkill Method
Packaging Material: Tyvek Pouch with a Ploymylar Sheath.
The SAL is 10 to the $-6$.
The maximum levels of residues of ethylene oxide: 25 parts per million; ethylene chlorohydrin: 25 parts per million and ethylene glycol: 250 parts per million.
This device is non-pyrogenic and the LAL Method is used to make that determination.
Pyrogens: We control the manufacturing environment to lessen the likelihood of pyrogen causing bacteria. In addition the LAL Method is used to determine that each lot is non-pyrogenic.
Page 2 Summary
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 20 1997
Mr. Frank J. Tighe
Official Correspondent
American Medical Devices, Inc.
1100 Northside Drive
Atlanta, GA 30318
Re: K970873
Trade Name: Endosoft™ Soft Tip Needle
(19 GA. and 20 GA.)
Regulatory Class: I
Product Code: 86 HMX
Dated: February 26, 1997
Received: March 10, 1997
Dear Mr. Tighe:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Mr. Frank J. Tighe
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,

A. Ralph Rosenthal, M.D.
Director
Division of Ophthalmic Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
AMD
AMERICAN
MEDICAL
DEVICES, Inc.
1100 Northside Drive Atlanta, Georgia 30318
(510(k) Number (if known): K970873
Device Name: Endosoft™ Soft Tip Needle (19 GA. and 20 GA.)
Indications For Use:
For irrigation/aspiration during eye surgery
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K970873
Prescription Use ☑ OR
(Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1-2-96)
| Toll-Free | Direct | Facsimile | Internet |
| --- | --- | --- | --- |
| 800-793-1473 | 404-815-5233 | 404-815-5235 | ftighe@americanmedicaldevices.com |