MOBILE OPERATING TABLE 1132.01

K970807 · Stierlen-Maquet AG · GDC · Mar 14, 1997 · General, Plastic Surgery

Device Facts

Record IDK970807
Device NameMOBILE OPERATING TABLE 1132.01
ApplicantStierlen-Maquet AG
Product CodeGDC · General, Plastic Surgery
Decision DateMar 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4960
Device ClassClass 1
Attributes3rd-Party Reviewed

Indications for Use

The device is for use during diagnostic examinations or surgical procedures to support and position a patient.

Device Story

Mobile operating table for patient support and positioning during surgery or diagnostics. Electrohydraulically operated system; controlled via hand control box, IR-transmitter, foot switch, or manual override. Powered by rechargeable batteries. Features include Trendelenburg/reverse Trendelenburg, lateral tilt, leg plate, back plate, and height adjustment; head plate adjustment is manual. Table top consists of radiotranslucent, electrically conductive, foam-upholstered sections. Used in operating rooms by surgical staff. Benefits include precise patient positioning and radiotranslucency for intraoperative imaging.

Clinical Evidence

Bench testing only. Compliance with IEC 601-1, EN 60601-1 (UL 2601), IEC 601-1-2, and prEN 50115. Radiotranslucency verified to meet 21 CFR 1020.30 (Aluminium Equivalent < 2 mm).

Technological Characteristics

Electrohydraulical system; battery-powered with built-in charger. Radiotranslucent, electrically conductive foam-upholstered sections. Complies with IEC 601-1, EN 60601-1, IEC 601-1-2, prEN 50115, DIN EN ISO 29001, and DIN EN ISO 46001. Radiotranslucency < 2 mm Aluminium Equivalent per 21 CFR 1020.30.

Indications for Use

Indicated for patients requiring support and positioning during diagnostic examinations or surgical procedures.

Regulatory Classification

Identification

Operating tables and accessories and operating chairs and accessories are AC-powered or air-powered devices, usually with movable components, intended for use during diagnostic examinations or surgical procedures to support and position a patient.

Predicate Devices

Submission Summary (Full Text)

{1} MAR 14 1997 K970807 # VII. Summary of Safety and Effectiveness Information for 510(k) Submission Mobile Operating Table 1132.01 ## General Information | Proprietary Name: | 1132.01 | | --- | --- | | Common Name: | Mobile Operating Table | | Classification Name: | Operating room table, AC - powered | | Classification: | Class I | | Classification Number: | FQO | | Intended use: | The device is for use during diagnostic examinations or surgical procedures to support and position a patient. | ## Legally marketed device | Proprietary Name: | Betastar 1131.02 | | --- | --- | | Common Name: | Mobile Operating Table | | Classification Name: | - | | Classification: | Class I | | Classification Number: | - | | Date of submission: | October 31, 1996 | | --- | --- | | Establishment Name and Address: | Stierlen-MAQUET AG Kehlerstr. 31 76437 Rastatt / GERMANY Contact Person: Reinhard Pfeuffer, OE Telephone: -49/7222/932-219 FAX: -49/7222/932-634 | Establishment Registration Number: 8010652 Safety and Effectiveness Information supporting the Substantial Equivalence Determination ## Performance Standards The device complies with - IEC 601-1 respectively EN 60601-1 (UL 2601) - Medical Electrical Equipment Part I, - General Requirements for Safety - IEC 601-1-2, prEN 50115 - Medical Electrical Equipment Part II, - Particular Requirements for the Safety of Operating Tables The design, manufacturing and quality control of the device comply with: - DIN EN ISO 29001 - DIN EN ISO 46001 510(k) Notification - Mobile Operating Table 1132.01 - Stierlen MAQUET AG 30 {2} # Technological Characteristics and the intended use of the device The operating table has been designed to fulfil the special requirements for patient positioning during surgical treatment. The side rails are provided for fitting additional parts. Before positioning the patient the device has to be covered with sheets. It is an electrohydraulical operated system which can be controlled by means of a hand control box, IR-transmitter, foot switch or with the override. Power is supplied by batteries, which can be recharged by a built-in battery charger. The operating table has the following functions: | Trendelenburg / reverse Trendelenburg positioning | electrohydraulically | | --- | --- | | Lateral tilt movement (right / left) | electrohydraulically | | Leg plate movement (up/down) | electrohydraulically | | Back plate movement (up / down) | electrohydraulically | | Head plate movement (up / down) | manual | | Hight adjustment | electrohydraulically | The table top is subdivided as follows: Head plate, removable back plate section, back plate, seat plate, leg plates Each foam upholstered patient supporting section is radiotranslucent and electrically conductive. The radiotranslucention correspond to 21 CFR 1020.30 with an Aluminium Equivalent less than 2 mm. The development of the design considers the latest issues of applicable regulations. The following table shows in detail which regulations were applied. | Applied Standards | 1132.01 | 1131.02 | | --- | --- | --- | | IEC 601-1 respectively EN 60601-1 (UL 2601) | yes | yes | | IEC 601-1-2 | yes | yes | | prEN 50115 | yes | no | | Design, manufacturing and quality control of the device comply with: DIN EN ISO 29001, DIN EN ISO 46001 | yes | no | 510(k) Notification - Mobile Operating Table 1132.01 - Stierlen MAQUET AG
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...