Factor deficient plasma, Factor - VIII is a human plasma immunodepleted of the specific factor and intended for use in the quantitative determination of the specific factor levels in patients suspected of congenital or acquired deficiency of this specific coagulation protein and is performed by clotting assay.
Device Story
Factor deficient coagulation plasma - VIII is human plasma immunodepleted of Factor VIII antigen. Used in clinical laboratories by technicians to perform clotting assays for quantitative determination of Factor VIII levels. Device aids in diagnosing congenital or acquired coagulation protein deficiencies. Product supplied frozen or lyophilized; utilizes donor human plasma, enhanced bovine factor V, and buffers. Performance verified via assay comparison to known samples; safety ensured by testing negative for HIV 1/2, HBsAG, HCV, and HIV-1ag.
Clinical Evidence
Bench testing only. Performance verified by comparing assay results to known samples. Safety validated by testing negative for HIV 1/2, HBsAG, HCV, and HIV-1ag using FDA-approved methods. Deficiency of Factor VIII confirmed at less than 1%.
Technological Characteristics
Human plasma immunodepleted of Factor VIII; contains enhanced bovine factor V and various buffers. Supplied in frozen or lyophilized form. Compatible with various coagulation instruments and reagents per manufacturer instructions.
Indications for Use
Indicated for quantitative determination of Factor VIII levels in patients suspected of congenital or acquired Factor VIII deficiency.
K964459 — UNIVERSAL REAGENT FACTOR X DEFICIENT PLASMA · Universal Reagents, Inc. · Dec 18, 1996
K961929 — FACTOR DEFICIENT PLASMA · Universal Reagents, Inc. · Oct 10, 1996
K964270 — UNIVERSAL REAGENT FACTOR V DEFICIENT PLASMA · Universal Reagents, Inc. · Dec 18, 1996
K971225 — CRYO CHECK FACTOR VIII DEFICIENT PLASMA · Precision Biologicals, Inc. · Jul 17, 1997
K972286 — FACTOR DEFICIENT PLASMAS - FACTOR XI (11) · Universal Reagents, Inc. · Jul 14, 1997
Submission Summary (Full Text)
{0}
UNIVERSAL REAGENTS, INC.
2858 North Pennsylvania Street
Indianapolis, Indiana 46205
PHONE: (317) 926-0015
FAX: (317) 926-0014
K970593
APR 25 1997
# Non-Confidential Summary of Safety and Effectiveness
February 17, 1997
page 1 of 2
Universal Reagents, Inc.
2858 N. Pennsylvania St.
Indianapolis, IN 46205
Tel - (317) 926-0006
Fax - (317) 926-0014
**Official contact:** Jorge Miller, Director, Coagulation Products
**Proprietary or Trade Name:** Factor deficient coagulation plasma - VIII
**Common/Usual Name:** Qualitative and Quantitative Factor Deficiency Test - VIII
**Classification Name:** Qualitative and Quantitative Factor Deficiency Test
**Intended device:** Factor deficient coagulation plasma - VIII
**Predicate devices:** OTC Factor VIII - K823453
**Device description:** Factor deficient plasma to be free of antigen of Factor VIII utilized in *in vitro* diagnostic use.
**Intended use:**
Indicated use - Factor deficient plasma, Factor - VIII is a human plasma immunodepleted of the specific factor and intended for use in the quantitative determination of the specific factor levels in patients suspected of congenital or acquired deficiency of this specific coagulation protein and is performed by clotting assay.
**Environment of use:** Clinical laboratories
**Comparison to predicate devices:**
| Attribute | Intended product | OTC |
| --- | --- | --- |
| Use | | |
| Indicated for use in determination of coagulation of plasma | Yes | Yes |
| In vitro diagnostic use | Yes | Yes |
| Used as a quantitative assay | Yes | Yes |
| Design | | |
| Factor VIII deficient plasma offered | Yes | Yes |
page 3 of 36
{1}
page 4 of 36
# Non-Confidential Summary of Safety and Effectiveness (continued)
February 17, 1997
page 2 of 2
## Comparison to predicate devices: (continued)
| Attribute | Intended products | OTC |
| --- | --- | --- |
| Packaging either - Frozen or Dry / lyophilized | Yes | Yes |
| Can be used with different instruments and reagents per manufacturer instructions | Yes | Yes |
| Materials | | |
| Donor human plasma | Yes | Yes |
| Enhanced bovine factor V | Yes | Yes |
| Various buffers | Yes | Yes |
| Performance Testing | | |
| Compare assay to known sample | Yes | Yes |
| Negative by FDA approved test for HIV 1/2 and HBsAG | Yes | Yes |
| Negative by FDA approved test for HCV and HIV-1ag | Yes | not known |
| Deficiency of relevant factor less than 1% | Yes | not known |
| Negative for HIV and HBsAG | Yes | Yes |
| Negative for HCV, HIV-1ag | Yes | not known |
| No inhibitor present | Yes | not known |
## Differences
The only difference is that the intended product is claimed to be negative for HCV and HIV-1ag by an FDA approved test.
Any other differences that do exist would not have a significant effect on the safety or effectiveness of the device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.