INNOFLOUR CALIBRATOR REAGENT SET
Device Facts
| Record ID | K970509 |
|---|---|
| Device Name | INNOFLOUR CALIBRATOR REAGENT SET |
| Applicant | Oxis Intl., Inc. |
| Product Code | DLJ · Clinical Toxicology |
| Decision Date | May 12, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3200 |
| Device Class | Class 2 |
Indications for Use
The INNOFLUOR™ Topiramate Assay System is a fluorescence polarization immunoassay intended for the quantitative determination of total topiramate in serum or heparinized plasma for therapeutic drug monitoring. The assay system is for use on the Abbott TDx® or the TDxFLx® analyzer.
Device Story
Fluorescence polarization immunoassay (FPIA) system; measures total topiramate concentration in serum or heparinized plasma; used on Abbott TDx® or TDxFLx® analyzers; operated by laboratory personnel; provides quantitative results for therapeutic drug monitoring; assists clinicians in managing topiramate dosage; benefits patients by ensuring therapeutic levels and minimizing toxicity.
Clinical Evidence
Bench testing; comparison of 117 patient samples using INNOFLUOR™ Topiramate Assay System on Abbott TDx® analyzer versus Gas Chromatography (GC). Linear regression analysis: y = 0.985x - 0.147; correlation coefficient r = 0.9934.
Technological Characteristics
Fluorescence polarization immunoassay (FPIA) technology; reagents, calibrators, and controls for use on Abbott TDx®/TDxFLx® analyzers; quantitative measurement of topiramate in serum/plasma.
Indications for Use
Indicated for quantitative determination of total topiramate in serum or heparinized plasma for therapeutic drug monitoring in patients receiving topiramate therapy.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- INNOFLUOR™ Phenobarbital Assay System (INNOFLUOR™ Phenobarbital Reagent Set and INNOFLUOR™ Phenobarbital Calibrator Set)
- Abbott Phenobarbital II Assay
Reference Devices
- Topiramate Gas Chromatography
Related Devices
- K970517 — INNOFLUOR TOPIRAMATE CONTROL SET · Oxis Intl., Inc. · May 12, 1997
- K970510 — INNOFLOUR TOPIRAMATE REAGENT SET · Oxis Intl., Inc. · May 12, 1997
- K070645 — SERADYN QMS TOPIRAMATE · Seradyn, Inc. · May 17, 2007