EVT ANGIOSCALE ANGIOGRAPHIC CATHETER
Device Facts
| Record ID | K970469 |
|---|---|
| Device Name | EVT ANGIOSCALE ANGIOGRAPHIC CATHETER |
| Applicant | Guidant Cardiac and Vascular Surgery |
| Product Code | DQO · Cardiovascular |
| Decision Date | May 8, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1200 |
| Device Class | Class 2 |
Intended Use
The EVT AngioScale angiography catheters provides angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to selected sites in the vascular system.
Device Story
Diagnostic intravascular catheter; PEBAX pigtail design; 6-7 French; 0.038” guidewire compatible. Features 11-20 radiopaque platinum marker bands along tubing. Used by physicians in clinical settings for angiographic visualization. Marker bands function as scaled ruler during imaging to provide linear measurements of vasculature. Contrast media injected via proximal polycarbonate luer adapter; dispersed through distal ports. Disposable; single-procedure use; provided sterile.
Clinical Evidence
Bench testing only.
Technological Characteristics
PEBAX catheter body; radiopaque platinum marker bands; polycarbonate luer adapter. 6-7 French diameter; 0.038” guidewire compatibility. Sterile, single-use, disposable. No software or electronic components.
Indications for Use
Indicated for angiographic visualization and linear measurement of the vasculature in patients requiring contrast media delivery to selected vascular sites.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
- Cordis Standard and High Flow Femoral-Ventricular Pigtail Catheters (K862006)
- Cordis 5.2 French Super Torque Plus Cardiology Catheter (K914007)
- Cordis Quanticor Marker Catheter
- Cordis Ducor Angiographic Catheter
Related Devices
- K180286 — Polyethylene Catheter · Cook Incorporated · Oct 26, 2018
- K991673 — CORDIS TEMPO ANGIOGRAPHY CATHETER · Cordis Corp. · May 26, 1999
- K122937 — SLIP-CATH BEACON TIP CATHETER MODEL SCBR (PRODUCT PREFIX), SHUTTLE SELECT SLIP-CATH CATHETER MODEL SCBR (PRODUCT PREFIX) · Cook, Inc. · Dec 14, 2012
- K961499 — LAGUNA SOFTIP DIAGNOSTIC CATHETERS · Boston Scientific Scimed, Inc. · Oct 17, 1996
- K113819 — MONGOOSE ANGIOGRAPHIC CATHETER · Pediacath, Inc. · Jul 11, 2012