ELECSYS CALCHECK PSA

K970458 · Boehringer Mannheim Corp. · JJX · Mar 6, 1997 · Clinical Chemistry

Device Facts

Record IDK970458
Device NameELECSYS CALCHECK PSA
ApplicantBoehringer Mannheim Corp.
Product CodeJJX · Clinical Chemistry
Decision DateMar 6, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Boehringer Mannheim Elecsys CalCheck PSA is used to verify the calibration assignment for the Boehringer Mannheim Elecsys PSA assay.

Device Story

Elecsys CalCheck PSA consists of three-level lyophilized, human-based materials; used to verify calibration assignment for Elecsys PSA assay. Device requires reconstitution; assayed in triplicate; results compared to target values. Used in clinical laboratory settings by laboratory personnel. Output provides verification of assay calibration accuracy, ensuring reliable patient PSA test results.

Clinical Evidence

No clinical data. Performance characteristics based on value assignment and stability testing.

Technological Characteristics

Three-level lyophilized, human-based material set. Matrix: Bovine serum albumin. Analyte: PSA. Intended for use with Elecsys PSA assay.

Indications for Use

Indicated for use as a calibration verification material for the Elecsys PSA assay to verify calibration assignment.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K 970458 MAR - 6 1997 # 510(k) Summary | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | --- | --- | | 1. Submitter name, address, contact | Boehringer Mannheim Corporation 2400 Bisso Lane Concord, CA 94524-4117 (510) 827 - 8215 Fax number: (510) 687-1850 Contact Person: Patricia M. Klimley Date Prepared: January 29, 1997 | | 2. Device Name | Proprietary name: Elecsys® CalCheck™ PSA Common name: Calibration Verification Material Classification name: Single (specified) analyte controls (assayed + unassayed) | | 3. Predicate device | The Elecsys® CalCheck™ PSA is substantially equivalent to currently marketed, Elecsys® CalCheck™ TSH (K963147). | | 4. Device Description | The Elecsys® CalCheck™ PSA is a three level single analyte set of lyophilized, human based materials. After reconstitution, they are assayed in triplicate and the results compared to the target values. | Continued on next page page 23 {1} page 24 # 510(k) Summary, Continued ## 5. Intended use The Boehringer Mannheim Elecsys CalCheck PSA is used to verify the calibration assignment for the Boehringer Mannheim Elecsys PSA assay. ## 6. Comparison to predicate device The Elecsys® CalCheck™ PSA is substantially equivalent to the currently marketed Elecsys® CalCheck™ TSH (K963147). The following table compares the Elecsys® CalCheck™ PSA with the predicate device, Elecsys® CalCheck™ TSH. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate devices is provided in attachment 6. ### Similarities: - Configuration: 3 CalCheck levels: low, mid, and high - Intended use: To verify calibration ### Differences: | Feature | CalCheck PSA | CalCheck TSH | | --- | --- | --- | | Matrix type | Bovine serum albumin | Horse serum | | Analyte | PSA | TSH | Continued on next page {2} page 25 # 510(k) Summary, Continued 6. Comparison to predicate device, (cont.) Performance Characteristics: - Value assignment and stability
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...