K970423 · Innovasive Devices, Inc. · MBI · Apr 30, 1997 · Orthopedic
Device Facts
Record ID
K970423
Device Name
INNOVASIVE ROCLET BONE TUNNEL AUGMENTATION DEVICE
Applicant
Innovasive Devices, Inc.
Product Code
MBI · Orthopedic
Decision Date
Apr 30, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The intended use of the ROClet is to protect the suture during its use in transosseous bone tunneling in the repair of rotator cuff tears of the shoulder.
Device Story
ROClet is a bone fixation accessory for rotator cuff repair; consists of molded acetal plastic component (2.8mm x 6.4mm) with flange and barbs; includes nylon suture threader assembly, insertion instrument, and awl. Surgeon uses awl to size bone tunnel; loads ROClet onto inserter; passes suture through threader assembly; withdraws threader; presses ROClet into bone tunnel. Device protects suture from bone abrasion and prevents suture migration through bone; suture is tied off to secure device until bone-to-tissue healing occurs. Used in clinical/OR setting by orthopedic surgeons. Benefits include reduced suture damage and improved load distribution during rotator cuff repair.
Clinical Evidence
Bench testing only. Bone model testing performed to compare ultimate strength of suture with and without ROClet in bone tunnel; cyclic loading characteristics evaluated.
Technological Characteristics
Material: Acetal plastic. Dimensions: 2.8mm diameter, 6.4mm length. Form factor: Molded component with flange and barbs. Includes nylon suture threader assembly. Non-sterile reusable insertion instrument and awl; sterile single-use implant. No software or electronic components.
Indications for Use
Indicated for patients undergoing rotator cuff repair surgery requiring transosseous bone tunneling.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
K970423
APR 30 1997
# 510(K) SUMMARY
## 1. SUBMITTER:
Innovasive Devices, Inc.
734 Forest St.
Marlborough, MA 01752
Telephone: 508-460-8229
Contact: Eric Bannon, Vice President of Quality Assurance and Regulatory Affairs
Date Prepared: February 4, 1997
## 2. DEVICE:
Innovasive 2.8mm x 6.4mm ROClet
Classification Name: Single/multiple component bone fixation appliances and accessories.
Trade Name: Innovasive Devices ROClet
## 3. PREDICATE DEVICE:
The predicate device used to determine substantial equivalence for the Innovasive Devices ROClet was the 2mm GORE-TEX soft tissue patch, produced by W.L. Gore, Flagstaff, AZ.
## 4. DEVICE DESCRIPTIONS:
The ROClet product consists of the implantable ROClet, an insertion instrument and an awl. The ROClet will be offered as a sterile single use device. The insertion instrument and awl will be offered as a non-sterile reusable instrument.
The ROClet consists of a molded component 2.8mm in diameter and 6.4mm long with a flange at one end and barbs along its length. The flange is intended to rest of the humeral head bone surface. The barbs are used to keep the device in place only until the suture is tied during the repair procedure. The ROClet is fitted with a nylon suture threader assembly placed in the ID of the device and is used to pass the sutures used during the tunnel procedure down the center of the ROClet until it is eventually tied during the repair.
The ROClet and suture threader assembly will be in a holding cartridge when received by the surgeon. The device is loaded onto the insertion instrument and removed from the cartridge. The sutures exiting the bone tunnel already established to repair the rotator cuff is passed through the suture threader assembly. The suture threader assembly with the suture is then withdrawn from the back of the ROClet and discarded. This results in the suture exiting the bone tunnel now passing through the center of the ROClet device and exiting the back
{1}
of the device at its flanged end. If needed, the awl can be used to size the bone tunnel. This ensures the device works with all methods used to create a bone tunnel.
The ROClet, still on the insertion tool is then manually pressed into the bone tunnel. The ribbed end of the ROClet, now in the tunnel, aids in keeping the device in place while the end of the suture exiting the device is tied off and the repair completed. The tied suture ensures the device remains in place in the bone tunnel hole until the bone to tissue interface heals. One ROClet is used for each bone tunnel.
## 5. INTENDED USE:
The intended use of the ROClet is to protect the suture during its use in transosseous bone tunneling in the repair of rotator cuff tears of the shoulder.
## 6. COMPARISON OF CHARACTERISTICS:
The Innovative Devices ROClet is comprised of acetal plastic. This device is used to protect suture when used to complete a repair of the rotator cuff through the use of transosseous bone tunnels. The ROClet is placed into the bone tunnel and the suture passed through its center. This protects the suture from damage from the surrounding hole and prevents migration of the suture through the bone. An inserter is used to place the device and an awl is used to ensure each bone tunnel is appropriately sized to receive the device.
The 2mm GORE-TEX soft tissue patch is a square of material constructed of Gore-tex which is used as a pladget. The material is used to augment the strength of the bone and surrounding tissue by more evenly distributing the load from the sutures over a wider area.
## 7. PERFORMANCE DATA:
The following performance data was provided in support of the substantial equivalence determination:
1. Bone model testing: Comparison of the ultimate strength of suture with and without the ROClet placed in the bone tunnel. In addition, the cyclic loading with the device was characterized.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.